FDA PMA FDA Class 3 Approved 🇺🇸 United States

Implant, Cochlear

PMA: P840024 · Supplement: S060 · Decision Mar 26, 1996
Classifications
1
FEI Numbers
22
Registration Numbers
22

Basic Information

Device Name
Implant, Cochlear
Trade Name
NUCLEUS 22 CHANNEL COCHLEAR IMPLANT SYSTEM
PMA Number
P840024
Supplement Number
S060
Device Class
FDA Class 3
Product Code
MCM
Generic Name
Implant, cochlear
Medical Specialty
Unknown
Advisory Committee
Ear, Nose, Throat
Decision
Approved
Decision Code
APPR
Decision Date
March 26, 1996
Date Received
November 24, 1995
Supplement Type
Normal 180 Day Track
Supplement Reason
Change Design/Components/Specifications/Material
Expedited Review
N

Advisory Committee Statement

APPROVAL TO MAKE THE FOLLOWING CHANGES TO THE SPECTRA 22 SPEECH PROCESSOR:1)THE PRINTED CIRCUIT BOARD WILL BE REDESIGNED FROM A MULTI-LAYER BOARD CONFIGURATION TO A SINGLE-LAYER BOARD CONFIGURATION. 2)AN AUDIO MIXING FUNCTION, CALLED THE AUTOSWITCHING AUDIO CIRCUIT, WILL BE INCLUDED. 3) THE CLOCK OSCILLATOR FREQUENCY WILL BE CHANGED FROM A NOMINAL 5 MHZ TO 5.011 MHZ. 4)COMPONENTS WILL BE ADDED TO GIVE FURTHER PROTECTION FROM ELECTROSTATIC DISCHARGE (ESD). 5)THE EAR-LEVEL MICROPHONE POWER SUPPLY WILL BE CHANGED TO A COSTANT CURRENT SOURCE. 6)THE EXTERNAL INPUT SOCKET WILL BE REPLACED WITH AMORE ROBUST TYPE. 7)THE PATIENT SWITCH POSITIONS WILL BE CHANGED TO PUT THE "NORMAL" SENSITIVITY SELECTION FIRST. 8)A LONGER-LIFE BATTERY WILL BE PROVIDED. 9)A RESISTOR WILL BE INSERTED IN SERIES TO PROTECT THE BACK-UP BATTERY FROM EXCESSIVE CURRENT. 10)THERE WILL BE A VERSION OF THE SPECTRA 22 SPEECH PROCESSOR WHICH PROVIDES INCREASED TRANSMITTER POWER. 11)SOME MANUFACTURING TEST PARAMETERS WILL BE SLIGHTLY MODIFIED. 12)FOUR TESTS WILL BE ADDED DURING THE MANUFACTURING PROCESS. 13)THE USER INSTRUCTION MANUAL WILL BE REVISED

Classifications

This FDA Pre-Market Approval entry is associated with 1 FDA classification via its product code.

Product Code Device Name
MCM Implant, Cochlear