FDA PMA FDA Class 3 Approved 🇺🇸 United States

Normothermic Machine Perfusion System For The Preservation Of Donor Livers Prior To Transplantation

PMA: P200031 · Decision Sep 28, 2021
Classifications
1
FEI Numbers
6
Registration Numbers
6

Basic Information

Device Name
Normothermic Machine Perfusion System For The Preservation Of Donor Livers Prior To Transplantation
Trade Name
Organ Care System (OCS™) Liver
PMA Number
P200031
Device Class
FDA Class 3
Product Code
QQK
Generic Name
Normothermic machine perfusion system for the preservation of donor livers prior to transplantation
Medical Specialty
Unknown
Advisory Committee
Gastroenterology, Urology
Decision
Approved
Decision Code
APPR
Decision Date
September 28, 2021
Date Received
June 30, 2020
Expedited Review
N

Advisory Committee Statement

Approval for Organ Care System (OCS) Liver for the indication of: The TransMedics® Organ Care System (OCS) Liver is a portable extracorporeal liver perfusion and monitoring system indicated for preservation and monitoring of hemodynamics and metabolic function which allows for ex-vivo assessment of liver allografts from donors after brain death (DBD) or liver allografts from donors after circulatory death (DCD) <=55 years old and with <=30 mins of warm ischemic time, macrosteatosis <=15%, in a near-physiologic, normothermic and functioning state intended for a potential transplant recipient.

Classifications

This FDA Pre-Market Approval entry is associated with 1 FDA classification via its product code.

Product Code Device Name
QQK Normothermic Machine Perfusion System For The Preservation Of Donor Livers Prior To Transplantation