FDA PMA
FDA Class 3
Approved
🇺🇸 United States
Next Generation Sequencing Oncology Panel, Somatic Or Germline Variant Detection System
PMA: P200010
·
Supplement: S022
·
Decision Feb 7, 2025
Classifications
1
FEI Numbers
12
Registration Numbers
12
Basic Information
- Device Name
- Next Generation Sequencing Oncology Panel, Somatic Or Germline Variant Detection System
- Trade Name
- Guardant360 CDx
- PMA Number
- P200010
- Supplement Number
- S022
- Device Class
- FDA Class 3
- Product Code
- PQP
- Generic Name
- Next generation sequencing oncology panel, somatic or germline variant detection system
- Medical Specialty
- Unknown
- Advisory Committee
- Pathology
- Decision
- Approved
- Decision Code
- APPR
- Decision Date
- February 7, 2025
- Date Received
- November 12, 2024
- Supplement Type
- Real-Time Process
- Supplement Reason
- Labeling Change - Indications/instructions/shelf life/tradename
- Expedited Review
- N
Advisory Committee Statement
Approval to update the Guardant360 CDx (P200010) workflow to allow patient samples with cell free DNA (cfDNA) input less than 5ng to be processed by the Guardant360 CDx workflow and be evaluated based on post-sequencing quality control metric.
Classifications
This FDA Pre-Market Approval entry is associated with 1 FDA classification via its product code.
| Product Code | Device Name | Device Class | Medical Specialty |
|---|---|---|---|
| PQP | Next Generation Sequencing Oncology Panel, Somatic Or Germline Variant Detection System | FDA class 3 | Unknown |