FDA PMA FDA Class 3 30-Day Notice Accepted 🇺🇸 United States

Cochlear Implant With Combined Electrical Stimulation And Acoustic Amplification

PMA: P130016 · Supplement: S029 · Decision Jul 6, 2017
Classifications
1
FEI Numbers
11
Registration Numbers
11

Basic Information

Device Name
Cochlear Implant With Combined Electrical Stimulation And Acoustic Amplification
Trade Name
Nucleus 24 Hybrid System
PMA Number
P130016
Supplement Number
S029
Device Class
FDA Class 3
Product Code
PGQ
Generic Name
Cochlear implant with combined electrical stimulation and acoustic amplification
Medical Specialty
Unknown
Advisory Committee
Ear, Nose, Throat
Decision
30-Day Notice Accepted
Decision Code
OK30
Decision Date
July 6, 2017
Date Received
June 7, 2017
Supplement Type
30-Day Notice
Supplement Reason
Process Change - Manufacturer/Sterilizer/Packager/Supplier
Expedited Review
N

Advisory Committee Statement

External supplier, Jabil Circuit in China, to be added as an alternate supplier of the CP800 Series Tamper resistant Battery Cover, CP800 Series Battery Holder and Nucleus Battery Holder components. A change was also requested to the solder material used in the battery holder printed circuit board assembly (PCBA).

Classifications

This FDA Pre-Market Approval entry is associated with 1 FDA classification via its product code.

Product Code Device Name
PGQ Cochlear Implant With Combined Electrical Stimulation And Acoustic Amplification