FDA PMA
FDA Class 3
30-Day Notice Accepted
🇺🇸 United States
Cochlear Implant With Combined Electrical Stimulation And Acoustic Amplification
PMA: P130016
·
Supplement: S029
·
Decision Jul 6, 2017
Classifications
1
FEI Numbers
11
Registration Numbers
11
Basic Information
- Device Name
- Cochlear Implant With Combined Electrical Stimulation And Acoustic Amplification
- Trade Name
- Nucleus 24 Hybrid System
- PMA Number
- P130016
- Supplement Number
- S029
- Device Class
- FDA Class 3
- Product Code
- PGQ
- Generic Name
- Cochlear implant with combined electrical stimulation and acoustic amplification
- Medical Specialty
- Unknown
- Advisory Committee
- Ear, Nose, Throat
- Decision
- 30-Day Notice Accepted
- Decision Code
- OK30
- Decision Date
- July 6, 2017
- Date Received
- June 7, 2017
- Supplement Type
- 30-Day Notice
- Supplement Reason
- Process Change - Manufacturer/Sterilizer/Packager/Supplier
- Expedited Review
- N
Advisory Committee Statement
External supplier, Jabil Circuit in China, to be added as an alternate supplier of the CP800 Series Tamper resistant Battery Cover, CP800 Series Battery Holder and Nucleus Battery Holder components. A change was also requested to the solder material used in the battery holder printed circuit board assembly (PCBA).
Classifications
This FDA Pre-Market Approval entry is associated with 1 FDA classification via its product code.
| Product Code | Device Name | Device Class | Medical Specialty |
|---|---|---|---|
| PGQ | Cochlear Implant With Combined Electrical Stimulation And Acoustic Amplification | FDA class 3 | Unknown |