FDA PMA FDA Class 3 30-Day Notice Accepted 🇺🇸 United States

Automated Insulin Dosing , Threshold Suspend

PMA: P120010 · Supplement: S084 · Decision Apr 5, 2016
Classifications
1
FEI Numbers
18
Registration Numbers
18

Basic Information

Device Name
Automated Insulin Dosing , Threshold Suspend
Trade Name
MINIMED 530G SYSTEM
PMA Number
P120010
Supplement Number
S084
Device Class
FDA Class 3
Product Code
OZO
Generic Name
Automated insulin dosing , threshold suspend
Medical Specialty
Unknown
Advisory Committee
Clinical Chemistry
Decision
30-Day Notice Accepted
Decision Code
OK30
Decision Date
April 5, 2016
Date Received
March 7, 2016
Supplement Type
30-Day Notice
Supplement Reason
Process Change - Manufacturer/Sterilizer/Packager/Supplier
Expedited Review
N

Advisory Committee Statement

Relocation of a contract manufacturing facility for printed circuit boards and board stack assemblies used in all Medtronic MiniMed Paradigm Real-Time Revel insulin pumps, Paradigm Real-Time insulin pumps, and MiniMed 530G insulin pumps. The firm is also requesting changes in the manufacturing process sequence as well as updates to mechanical drawings to add vendor name and pump models. The Paradigm Real-Time insulin pumps are a component of the Paradigm Real-Time System, Paradigm Real-Time Revel System, and the Paradigm Real-Time Revel System with Enlite Sensor. The MiniMed 530G insulin pumps are a component of the MiniMed 530G System.

Classifications

This FDA Pre-Market Approval entry is associated with 1 FDA classification via its product code.

Product Code Device Name
OZO Automated Insulin Dosing , Threshold Suspend