FDA PMA
FDA Class 3
Approved
🇺🇸 United States
Aortic Valve, Prosthesis, Percutaneously Delivered
PMA: P100041
·
Supplement: S027
·
Decision May 22, 2013
Classifications
1
FEI Numbers
24
Registration Numbers
24
Basic Information
- Device Name
- Aortic Valve, Prosthesis, Percutaneously Delivered
- Trade Name
- RETROFLEX 3 DELIVERY SYSTEM, RETROFLEX BALLON CATHETER, EDWARDS TRANSFEMORAL BALLON CATHETER, CRIMPER
- PMA Number
- P100041
- Supplement Number
- S027
- Device Class
- FDA Class 3
- Product Code
- NPT
- Generic Name
- Aortic valve, prosthesis, percutaneously delivered
- Medical Specialty
- Unknown
- Advisory Committee
- Cardiovascular
- Decision
- Approved
- Decision Code
- APPR
- Decision Date
- May 22, 2013
- Date Received
- April 23, 2013
- Supplement Type
- Special (Immediate Track)
- Supplement Reason
- Process Change - Manufacturer/Sterilizer/Packager/Supplier
- Expedited Review
- N
Advisory Committee Statement
APPROVAL FOR ADDITION OF POLYETHYLENE FOAM TO THE BASE OF THE CRIMPER DURING PACKAGING, A CHANGE TO THE CRIMPER PACKAGE RELEASE TEMPERATURE SET POINT, AND ADDITION OF POUCH INTEGRITY INSPECTIONS FOR ALL SAPIEN ACCESSORIES.
Classifications
This FDA Pre-Market Approval entry is associated with 1 FDA classification via its product code.
| Product Code | Device Name | Device Class | Medical Specialty |
|---|---|---|---|
| NPT | Aortic Valve, Prosthesis, Percutaneously Delivered | FDA class 3 | Unknown |