FDA PMA FDA Class 3 30-Day Notice Accepted 🇺🇸 United States

System, Endovascular Graft, Aortic Aneurysm Treatment

PMA: P100021 · Supplement: S052 · Decision Jun 16, 2016
Classifications
1
FEI Numbers
34
Registration Numbers
34

Basic Information

Device Name
System, Endovascular Graft, Aortic Aneurysm Treatment
Trade Name
MEDTRONIC VASCULAR ENDURANT STENT GRAFT SYSTEM
PMA Number
P100021
Supplement Number
S052
Device Class
FDA Class 3
Product Code
MIH
Generic Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Medical Specialty
Unknown
Advisory Committee
Cardiovascular
Decision
30-Day Notice Accepted
Decision Code
OK30
Decision Date
June 16, 2016
Date Received
May 19, 2016
Supplement Type
30-Day Notice
Supplement Reason
Process Change - Manufacturer/Sterilizer/Packager/Supplier
Expedited Review
N

Advisory Committee Statement

Addition of an inspection step to the rear handle assembly process; 2) narrowing of the thread radius dimension tolerances on the handle lock; and 3) addition of detailed cleaning instructions at supplier for the moulding equipment.

Classifications

This FDA Pre-Market Approval entry is associated with 1 FDA classification via its product code.

Product Code Device Name
MIH System, Endovascular Graft, Aortic Aneurysm Treatment