Test, Hepatitis B (B Core, Be Antigen, Be Antibody, B Core Igm)
Basic Information
- Device Name
- Test, Hepatitis B (B Core, Be Antigen, Be Antibody, B Core Igm)
- Trade Name
- ELECSYS ANTI-HBS TEST SYSTEM
- PMA Number
- P010054
- Supplement Number
- S008
- Device Class
- FDA Class 3
- Product Code
- LOM
- Generic Name
- Test, hepatitis b (b core, be antigen, be antibody, b core igm)
- Medical Specialty
- Unknown
- Advisory Committee
- Microbiology
- Decision
- Approved
- Decision Code
- APPR
- Decision Date
- October 30, 2008
- Date Received
- April 30, 2008
- Supplement Type
- Normal 180 Day Track
- Supplement Reason
- Change Design/Components/Specifications/Material
- Expedited Review
- N
Advisory Committee Statement
APPROVAL TO EXTEND APPLICATION OF THE ELECSYS HBSAG AND ELECSYS ANTI-HBS TEST SYSTEMS ONTO THE UPDATED VERSION OF THE MODULAR ANALYTICS E170 ANALYZER KNOWN AS THE COBAS E 601 ANALYZER. WHILE THE MODULAR ANALYTICS MODULAR SYSTEM INTEGRATES THE E170 MODULE WITH CLINICAL CHEMISTRY MODULES P OR D, THE COBAS E 601 IMMUNOASSAY ANALYZER IS A COMPONENT OF THE C6000 MODULAR SYSTEM WHICH INTEGRATES IMMUNOLOGY TESTING WITH THE COBAS C 501 CLINICAL CHEMISTRY ANALYSER. THERE HAVE BEEN NO CHANGES TO THE COMPOSITION OF ANY OF TE TEST SYSTEM REAGENTS (IMMUNOASSAY, CONTROLS OR CONFIRMATORY TEST), THE APPLICATION PARAMETERS, THE ANALYZER¿S ANALYTICAL MEASURING COMPONENT OR HOW THE RESULTS ARE CALCULATED. THE INDICATIONS FOR USE OF THE TWO TEST SYSTEMS REMAIN THE SAME.
Classifications
This FDA Pre-Market Approval entry is associated with 1 FDA classification via its product code.
| Product Code | Device Name | Device Class | Medical Specialty |
|---|---|---|---|
| LOM | Test, Hepatitis B (B Core, Be Antigen, Be Antibody, B Core Igm) | FDA class 3 | Unknown |