FDA Enforcement
Class II
Ongoing
WALLFLEX ESOPHAGEAL FC 18/23-25MMX15CM- Agile Esophageal Partially Covered and Fully Covered OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN): M00516720
Recall: Z-3277-2024
·
Reported October 2, 2024
Enforcement
- Recall Number
- Z-3277-2024
- Event ID
- 95124
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 2, 2024
- Initiation Date
- August 1, 2024
- Classification Date
- September 26, 2024
- Address
- 100 Boston Scientific Way, Marlborough, MA, 01752-1234, United States
Description
WALLFLEX ESOPHAGEAL FC 18/23-25MMX15CM- Agile Esophageal Partially Covered and Fully Covered OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN): M00516720
Reason
WallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
Code Info
GTIN: 08714729778059 Lot Numbers: 32770266, 32780362, 32865975, 32906417, 32975732, 33017656, 33128876, 33167135
Distribution
Worldwide distribution - US Nationwide and the countries of APAC, Canada, EMEA, LATAM.