Bausch + Lomb Stellaris Elite Vision Enhancement System, 25 ga. Combined Mid-Field Elite Pack, REF SE5525MV (containing the Stellaris Elite 25 GA Vit Cutter, Model BL5628), packaged in sterile trays, 6 trays/shipper.
Enforcement
- Recall Number
- Z-3276-2018
- Event ID
- 80837
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Bausch & Lomb Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 3, 2018
- Initiation Date
- August 8, 2018
- Classification Date
- September 27, 2018
- Termination Date
- August 28, 2019
- Address
- 3365 Tree Court Industrial Blvd, Saint Louis, MO, 63122-6615, United States
Description
Bausch + Lomb Stellaris Elite Vision Enhancement System, 25 ga. Combined Mid-Field Elite Pack, REF SE5525MV (containing the Stellaris Elite 25 GA Vit Cutter, Model BL5628), packaged in sterile trays, 6 trays/shipper.
The back cap separates from the body of the vitrectomy cutter.
Lot numbers W0787, exp. 5/11/2019; W0888, exp. 6/4/2019; W1003, exp. 6/23/2019; and W1129, exp 7/6/2019, UDI 00757770059430.
Distribution was made to AL, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MA, ME, MO, MT, NC, NJ, NY, OH, OK, OR, PA, TN, TX, UT, and WV. There was government distribution but no military distribution. Foreign distribution was made to Russia, Romania, Poland, The Netherlands, Jordan, Italy, Israel, Switzerland, Hungary, Greece, Guadeloupe, Great Britain, France, Spain, Singapore, Malaysia, and Thailand.
66 cases