HipLOC Hip Fracture Fixation System-Hiploc System 135¿ Plate 38 mm Barrel 4 Hole Plate Item Number: 2110-135-004 Product Usage: The HipLOC Compression Hip Screw may be used primarily for fixation of bone fractures. Specific indications include open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures, intertrochanteric fractures and subtrochanteric fractures
Enforcement
- Recall Number
- Z-3263-2018
- Event ID
- 80921
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomet UK Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- October 3, 2018
- Initiation Date
- August 9, 2018
- Classification Date
- September 27, 2018
- Termination Date
- May 12, 2020
- Address
- Waterton Industrial Estate, Bridgend, N/A, United Kingdom
Description
HipLOC Hip Fracture Fixation System-Hiploc System 135¿ Plate 38 mm Barrel 4 Hole Plate Item Number: 2110-135-004 Product Usage: The HipLOC Compression Hip Screw may be used primarily for fixation of bone fractures. Specific indications include open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures, intertrochanteric fractures and subtrochanteric fractures
HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel
Lot Number: 3222067 3355792 3609499 3609508 3619245 3623355 3623356 3700289 3700610 3829536 3843022 3847671 3847672 3879198 3879199 3879200 3879202 3889526 3897103 3897106 3897115 3902248 3918363 3918364
US Nationwide in the states of AZ, MO, NJ and NM
44 units