FDA Enforcement Class II Terminated

VISIONIST X4 CRT-P (Cardiac Resynchronization Therapy Pacemaker)

Recall: Z-3259-2018 · Reported October 3, 2018

Enforcement

Recall Number
Z-3259-2018
Event ID
81001
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Boston Scientific Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 3, 2018
Initiation Date
September 10, 2018
Classification Date
September 27, 2018
Termination Date
November 4, 2019
Address
4100 Hamline Ave N Bldg 3, Saint Paul, MN, 55112-5700, United States

Description

VISIONIST X4 CRT-P (Cardiac Resynchronization Therapy Pacemaker)

Reason

Boston Scientific has identified a subset of active pacemaker and CRT-Ps systems with an elevated potential for early pacemaker replacement due to hydrogen-induced accelerated battery depletion.

Code Info

VISIONIST X4 (CRT-P) model U228

Distribution

AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, W A, WI, WV, and WY