FDA Enforcement Class II Ongoing

LINK Universal Handle, with quick coupling, Stainless Steel, straight, Item Number 130-394/01

Recall: Z-3245-2024 · Reported October 2, 2024

Enforcement

Recall Number
Z-3245-2024
Event ID
95344
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Waldemar Link GmbH & Co. KG (Corp. Hq.)
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 2, 2024
Initiation Date
September 11, 2024
Classification Date
September 24, 2024
Address
Barkhausenweg 10, Hamburg, N/A, Germany

Description

LINK Universal Handle, with quick coupling, Stainless Steel, straight, Item Number 130-394/01

Reason

Some lots of the LINK Universal Handle, with quick coupling, Stainless Steel, straight may suffer an immediate material failure intraoperatively when the instrument is used for the first time. The instrument can break in the middle.

Code Info

UDI-DI: 04026575215539; Lot Codes: C204156; C219063; C219064; C233033; C233034; C332135

Distribution

US Nationwide distribution in the states of FL, KS, MI, NJ, VA.

Quantity

32 devices