FDA Enforcement Class II Ongoing

Nasal Sinus Pack, MTSN26B; Medical convenience kit

Recall: Z-3242-2024 · Reported October 2, 2024

Enforcement

Recall Number
Z-3242-2024
Event ID
95261
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
American Contract Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 2, 2024
Initiation Date
August 7, 2024
Classification Date
September 24, 2024
Address
2610 Ne Industrial Dr Ste 220, Kansas City, MO, 64117-2648, United States

Description

Nasal Sinus Pack, MTSN26B; Medical convenience kit

Reason

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Code Info

UDI-DI: 00191072200242; Lot: 2405093; Exp: 05/09/2025

Distribution

US Nationwide distribution in the states of MO, MN, MA, OH, NE.

Quantity