FDA Enforcement Class II Terminated

Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002

Recall: Z-3240-2018 · Reported October 3, 2018

Enforcement

Recall Number
Z-3240-2018
Event ID
80863
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker GmbH
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 3, 2018
Initiation Date
July 26, 2018
Classification Date
September 22, 2018
Termination Date
May 26, 2020
Address
Bohnackerweg 1, Selzach, N/A, Switzerland

Description

Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002

Reason

When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan.

Code Info

Version #: HLRF 1.0-2/002, HLRF 1.1-2/002

Distribution

Nationwide

Quantity

233