ACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.0 Diagnostic Ultrasound System, REF: 11711779, with software
Enforcement
- Recall Number
- Z-3217-2024
- Event ID
- 95254
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 2, 2024
- Initiation Date
- August 15, 2024
- Classification Date
- September 24, 2024
- Address
- 22010 Se 51st St, Issaquah, WA, 98029-7298, United States
Description
ACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.0 Diagnostic Ultrasound System, REF: 11711779, with software
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
11711750, UDI-DI: 04056869995809, Software VA10A 11711779, UDI-DI: 04056869995830, Software VA10A
US: MS, WA, TX, NJ, GA, PN, NY, AL, MN, NE, IN, CT, CA, IL, TN, WI, ND, NM, AR, NC, PR, RI, WY, MD, FL, HI, AZ, SC, MO, OK, VA, ME, KS, MI, LA, IA, OH, SD, NH, KY, AL, OR, MA, GU. OUS: Afghanistan, Albania, Algeria, Andorra, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Cameroon, Canada, Chile, China, Colombia, Congo, the Democratic Republic of the, Costa Rica, C¿te dIvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guinea, French Polynesia, Georgia, Germany, Greece, Guatemala, Guernsey, Guyana, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein LI B, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Malta, Mexico, Moldova, Rep. of, Mongolia, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Reunion, Romania, Ruanda, Russian Federation, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe
321