FDA Enforcement Class II Terminated

Medtronic Clip Gun Magazines are packaged in individually sealed pouches and distributed in boxes containing 5 individually pouched Clip Gun Magazine. REF CM8902, Medtronic Clip Gun Magazine, Rx only, CE 0344, STERILE EO, Rx only, Single Use

Recall: Z-3212-2018 · Reported October 3, 2018

Enforcement

Recall Number
Z-3212-2018
Event ID
80657
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Medtronic Neurosurgery
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 3, 2018
Initiation Date
May 3, 2018
Classification Date
September 21, 2018
Termination Date
August 30, 2022
Address
125 Cremona Dr, Goleta, CA, 93117-3083, United States

Description

Medtronic Clip Gun Magazines are packaged in individually sealed pouches and distributed in boxes containing 5 individually pouched Clip Gun Magazine. REF CM8902, Medtronic Clip Gun Magazine, Rx only, CE 0344, STERILE EO, Rx only, Single Use

Reason

The packaging sterile barrier does not comply with the applicable packaging standard for terminally sterilized medical devices. While the Instructions for Use (IFU) state Do not use if package has been previously opened or damaged , it is possible that the seal breach may go undetected by the user's naked eye.

Code Info

Product Number/CFN: CM8902 UDI-Device Identifier (GTIN/UPN): 20643169647200 Lot Numbers: 54984; 57188; 59558; 66312; and 67988

Distribution

Domestic: AL, AZ, AR, CA, CO, DC, FL, GA, ID, IL, IN, IA, KS, LA, MD, MI, MN, MS, MO, NB, NV, NJ, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY OUS: Australia, Germany, Kenya, Russian Federation, South Africa

Quantity

2616