Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems
Recall: Z-3211-2018
·
Reported September 26, 2018
Enforcement
- Recall Number
- Z-3211-2018
- Event ID
- 80933
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Accuray Incorporated
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- September 26, 2018
- Initiation Date
- August 31, 2018
- Classification Date
- September 20, 2018
- Termination Date
- December 16, 2020
- Address
- 1240 Deming Way, N/A, Madison, WI, 53717-1954, United States
Description
Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems
Reason
the Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System
Code Info
All units with the Ashland Brand distributed between August 1, 2017 and July 31, 2018
Distribution
Worldwide Distribution: United States (nationwide), and countries of: United Kingdom, Germany, France, Finland, Netherlands, Portugal, Hong Kong, Poland, Australia, United Arab Emirates, Thailand, and Japan.
Quantity
259 units