Endoscopic Cleaning Brush, single use, sold in kits and labeled as the following: a. Item Number DYK1002DCB, BRUSH, CLEANING, DOUBLE ENDED, W/ HANDLE - 300 EA/CS, 50EA/BX, 6BX/CS; b. Item Number DYK1002DBDE, BRUSH, CLEANING, DBL, DIF ENDS - 500 EA/CS, 50 EA/BX, 10 BX/CS; c. Item Number DYK1002DBSE, BRUSH, CLEANING, DBL, SAME END - 500 EA/CS, 50 EA/BX, 10 BX/CS; d. Item Number DYK1002SBC, BRUSH, CLEANING, SINGLE END - 500 EA/CS, 50 EA/BX, 10 BX/CS
Enforcement
- Recall Number
- Z-3210-2018
- Event ID
- 80758
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medline Industries Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 26, 2018
- Initiation Date
- June 15, 2018
- Classification Date
- September 20, 2018
- Termination Date
- May 18, 2020
- Address
- Three Lakes Drive, N/A, Northfield, IL, 60093, United States
Description
Endoscopic Cleaning Brush, single use, sold in kits and labeled as the following: a. Item Number DYK1002DCB, BRUSH, CLEANING, DOUBLE ENDED, W/ HANDLE - 300 EA/CS, 50EA/BX, 6BX/CS; b. Item Number DYK1002DBDE, BRUSH, CLEANING, DBL, DIF ENDS - 500 EA/CS, 50 EA/BX, 10 BX/CS; c. Item Number DYK1002DBSE, BRUSH, CLEANING, DBL, SAME END - 500 EA/CS, 50 EA/BX, 10 BX/CS; d. Item Number DYK1002SBC, BRUSH, CLEANING, SINGLE END - 500 EA/CS, 50 EA/BX, 10 BX/CS
There is potential risk of brush detachment during cleaning. If brush detachment is not identified, there is the potential for the brush to be left within the scope.
Endoscopic cleaning brush finished good lot numbers: GM10, GJ25, HA05, HA20. Kits: a. Item Number DYK1002DCB, kit Number (Kit Lot Number): DYK1001DCB (18BA0735), DYK1002144B (18BA0644, 18DA0124), DYK1008972E (18BA2199, 18DA1418, 18EAA023, 18EAA653), DYK1012268E1 (18BA1667, 18CA0872, 18CA1289, 18DA1732), DYK1012414E1 (18CA0793, 18CA1644, 18DA0413), DYK1017922E (18CA1300), DYK1018542E (18CA0622, 18DA1159, 18EAB507), DYK1044978E (18EBE085, 18NB1289), DYK1067817E1 (18AA1969, 18EAA004, 18EAB365), DYK1072097E (18DA0437, 18EAA790), DYK1097615E (18CA0687, 18EAA084), DYK1177017E (18BA2352, 18DA1481), DYK1188722E (18EAA016, 18EAB529), DYK1253578E (18AA2092, 18DA0331), DYK1253578F (18DA1595), DYK1266757E (18BA0203, 18DA0062, 18DF0085), DYKE1003A (18DA1488, 18FAA025), DYKE1011B (18AA1276), DYKE1012C (18DA0090, 18EAA521), DYKE1013 (18CB4388, 18EBL745), DYKE1014B (17KA1195, 18AA0984, 18CA0691, 18CA1969, 18DA0344, 18DA1167), DYKE1022A (18EBE830), DYKE1029C (18BA0279, 18CA0842), DYKE1032C (18BB1877, 18BB6935, 18EBH874, 18EBS890, 18OB9988), DYKE1033B (18MB9361), DYKE1033C (18CB1145, 18DB0018, 18EBH875), DYKE1037 (18CA1072, 18DA0377), DYKE1043C (18DA0411, 18EAA198), DYKE1048 (18BA0550), DYKE1050 (18BA0830, 18CA1744, 18EAA936), DYKE1056C (18DA0952, 18EBF206, 18EBF513, 18EBN896, 18PB3439), DYKE1058A (18BB6939, 18CB5355, 18PB4193), DYKE1059A (18BB582118EBF514), DYKE1062A (18CA1504), DYKE1062B (18EAB341), DYKE1063 (18BB4018, 18EBL836, 18OB2537, 18PB6224), DYKE1064 (18BB2161, 18DB8246, 18EBL833), DYKE1069A (18CA0564, 18CA1222), DYKE1069B (18DA0510), DYKE1072 (18BA0524, 18DA0412, 18EAA070), DYKE1077 (18BA1594, 18DA0096, 18EAB672), DYKE1096A (18BA1239, 18DA0345, 18FAA190), DYKE1099 (18BA1707), DYKE1108C (17JA2124, 17KA1225), DYKE1108D (18AA2499, 18BA1296, 18CA0336), DYKE1112 (18CA0139), DYKE1120 (18BA2330, 18DA0523, 18EAB011), DYKE1123 (18CA0181, 18DA0779, 18EAB549), DYKE1126A (18BA0589, 18CA1597), DYKE1138A (18BA1559, 18CA2056), DYKE1143A (18BA0473), DYKE1152 (18DA1187, 18EAA714), DYKE1158A (18BA2208, 18EAA253, 18EAB695, 18FAA253), DYKE1159A (18FAA227), DYKE1161A (18BA1435, 18EAB349), DYKE1162G (17JA2795), DYKE1162I (18AA1616, 18AA2211, 18BA0354), DYKE1162J (18BA1753, 18DA1469), DYKE1164 (18EBD023, 18OB2534), DYKE1172C (18EAA129), DYKE1176 (18CA1092, 18DA1719), DYKE1184B (17IA1038, 17JA1424, 17LA1730, 18BA1937, 18DA1028), DYKE1185 (18CA1513), DYKE1203A (18DA0680, 18FAA307), DYKE1207 (18BA1174, 18CA0313, 18EAA076, 18EAA397), DYKE1224C (18DA0018, 18EAA709), DYKE1232 (18CA0337, 18EAA730), DYKE1233 (18AA1392, 18BA0937, 18DA0390, 18DB7237, 18EBG997, 18EBT332), DYKE1234 (18FAA404), DYKE1245 (18BA0307, 18BA0908), DYKE1249 (18BA1440, 18DA1612, 18FAA258), DYKE1250 (18BA2009, 18EAA141), DYKE1258 (18BA1480, 18EAA325), DYKE1265 (18DA0068, 18DA0206, 18EAB322, 18EAB747), DYKE1274A (18CA0987, 18EAB342), DYKE1293 (18BA0492, 18CA1224, 18DA1377, 18EAB042), DYKE1295 (18BA0770, 18CA1963, 18DA1479), DYKE1303 (18MB9887, 18OB2532), DYKE1304 (18BB5870, 18DB7236), DYKE1308 (18CA0292, 18DA0841), DYKE1313B (18CA1559, 18EAB036), DYKE1314A (18CA1327), DYKE1318 (18BA0888, 18DA1352, 18FAA368), DYKE1332A (18EAA061), DYKE1333 (18BA0763, 18EAA432), DYKE1334B (18BA0597, 18EAA447), DYKE1345 (18CA1207, 18DA1486), DYKE1350A (18BB2129, 18DB7371, 18EBO982), DYKE1357B (18CA0267), DYKE1357C (18DA1208, 18FAA100), DYKE1364A (17LA1129, 18BA0029, 18DB7239, 18EBD050), DYKE1366 (18BA0796, 18CA1707, 18EAB410), DYKE1369C (17JA2209, 17LA2210, 18BA0882, 18CA1422), DYKE1372 (18NB4423), DYKE1375A (18BA1778, 18CA2032), DYKE1385 (18BA0562, 18CA1193, 18EAA304), DYKE1386 (18CA0393, 18DA0589, 18FAA222), DYKE1391 (18EAB593), DYKE1393A (18BA0591, 18CA1406, 18DA0833, 18EAA789), DYKE1395A (18CA0886, 18DA0717, 18FAA293), DYKE1399 (18DA0204, 18EAA422, 18FAA144), DYKE1402 (18BB5817, 18DB7373, 18EBS803), DYKE1406 (18CA0328, 18DA0091), DYKE1409 (18DA0216, 18EAB769), DYKE1410 (18BA0708, 18CA0837, 18DA1564, 18EAA453), DYKE1415 (18CA1723, 18DA1887, 18EAB265), DYKE1416 (18DA1072, 18EAA886), DYKE1422A (18BA1978, 18DA1468), DYKE1423A (18CA1662, 18DA1174, 18EAA686), DYKE1431B (18BA0503, 18DA1586, 18EAA301, 18EAB901), DYKE1438A (18AA1176), DYKE1438B (18CA0630, 18EAB038), DYKE1441B (18DA1320, 18DA1745, 18EAB829), DYKE1442A (18DB7372, 18NB2540), DYKE1448 (18BA1463, 18CA1544), DYKE1455 (18CA0109, 18EAA273, 18EAB548), DYKE1456 (18BA0145, 18CA0268, 18DA0642, 18EAA706), DYKE1457 (18BA0626, 18EAA078), DYKE1460 (18AA2138, 18CA1928, 18EAB528), DYKE1466 (18BA1634, 18CA1187, 18DA1776), DYKE1468 (18BA1335, 18EAB838), DYKE1469A (18BA0868, 18DA1474), DYKE1475A (17LA0444, 18AA2019, 18BA1162, 18CA1545), DYKE1480 (18BA0028, 18BA0211, 18BB6704, 18EBQ945, 18OB9926, 18PB5260), DYKE1481 (18BA1471, 18CA1402, 18DA1569, 18EAB051), DYKE1487A (18AA2145, 18BA1736), DYKE1487B (18DA0599), DYKE1488 (18DB7238, 18EBE095), DYKE1491 (18DA0948), DYKE1494 (18BA1450, 18DA1778), DYKE1498 (18CA0824), DYKE1503B (17LA0480, 18AA0783, 18BA0551, 18CA0587, 18DA0778, 18EAA371, 18FAA192), DYKE1506 (18FAA157), DYKE1507 (18DA0591, 18EAB736, 18FAA039), DYKE1516A (18AA1029, 18BA0055, 18BA1034, 18EAB049), DYKE1519 (17IA0093, 18BA1540, 18CA1650, 18EAA320), DYKE1521 (17IA2305, 17JA2046), DYKE1521A (17LA1124, 17LA2004, 18AA2110, 18DA0535), DYKE1526 (17JA0857, 17KA1032, 18AA2090), DYKE1526A (18BA1336, 18CA1705, 18DA1783, 18EAA523, 18FAA247), DYKE1527 (17IA1501, 17KA1034, 18AA0950, 18CA1957, 18DA1731, 18EAB046), DYKE1535 (17JA1057, 17LA0481, 18AA2114, 18DA1728), DYKE1536 (17JA0776, 17KA1944, 18AA1963, 18BA1970, 18DA0211, 18EAA468), DYKE1537 (17JA1066), DYKE1538 (17LA0810, 18BA0475, 18BA2349, 18DA1031, 18EAB415), DYKE1543 (17KA0386, 17LA1373, 18AA2292), DYKE1554 (18AA0653, 18CA1845), DYKE1555 (17LA1505), DYKE1559 (18AA1924, 18BA0444), DYKE1561 (18BA1796, 18DA0372, 18EAB826), DYKE1564 (18CA1183, 18EAA246), DYKE1567 (18CA0739), DYKE1573 (18EAA049, 18FAA288), DYKE1574 (18DA1130, 18DA1594), DYKE1579 (18EAB289, 18EAB583), DYKE1586 (18EAB222), DYKE1588 (18FAA342), DYKE18CMPLY (18CA1467), DYKENDOBASIC (18CA0995, 18DA1811, 18FAA369), DYKENDOBPLUS (18BA1853, 18DA1190, 18FAA207), DYKENDOCLEAN (18DA0382, 18EAB587), DYKENDOCMPLY (18BA1056, 18CA0124, 18DA0129, 18EAB667), DYKENDODBCMP (18BA0916, 18DA0383, 18EAA260, 18EAB752), DYKENDOLWCMP (18DA1812, 18FAA202), DYKENDOUPCMP (18BA1854, 18DA0718, 18EAA508), DYKM1404 (18CA1636, 18DA0654), DYKM1555 (17JA1068, 17LA1482, 18AA1177, 18AA2029, 18AA2612, 18CA0565, 18DA0069, 18EAA433), DYSAMP200458 (18DA1963), DYSAMP200460 (18FAA265), P880995 (18EKD315), P881009 (18EAA498), P971311 (18BA0897, 18CA1642, 18DA0859, 18EAA859), PHS347468000 (18DK2120, 18FKB069); b. Item Number DYK1002DBDE, kit Number (Kit Lot Number): DYK1001DBDE (17IA0598, 17LA0129), DYK1002DBDE (18BA0041), DYK1276244E1 (17JA0713), DYKE1122C (17JA0877, 17JA1514, 17JA2093, 17LA0460, 17LA1984, 18BA2299, 18CA0795, 18CA1927, 18EAA705), DYKE1446A (17JA2156, 17LA1648, 18BA0062, 18CA1226, 18DA1726), DYKE1447A (17IA1474, 17LA1178, 18BA0942), DYKE1516 (17HA1857, 17JA2263), DYKE1539 (17JA1281, 17KA1644, 18AA2157), DYKE1540 (17KA1547), DYKE1540A (18BA0740, 18CA1706, 18DA1020, 18EAB360), DYKE1552 (17LA1291, 18CA0448, 18CA0843), DYKE1580 (18EAA623, 18FAA191); c. Item Number DYK1002DBSE, kit Number (Kit Lot Number): DYK1091457E1 (17LA2255, 18CA0974, 18EAA258), DYK1245641E (17LA1486, 18CA1307), DYKE1000A (17LA0061, 17LA1462, 18CA1093, 18EAA814), DYKE1071 (17LA1174, 18AA0941, 18AA1383), DYKE1110 (18AA1490, 18CA0971, 18DA1557), DYKE1131 (18CA0972, 18EAA321), DYKE1135 (18AA0928, 18BA2275), DYKE1136B (18AA0807, 18AA2529, 18CA0970), DYKE1142 (18AA1125, 18DA0283, 18EAA075), DYKE1144 (18AA0905, 18BA0411, 18CA1924), DYKE1156 (18AA1587, 18CA0976), DYKE1157 (18BA1683, 18CA0975, 18DA1464), DYKE1193 (18AA2159, 18DA0924), DYKE1200B (18AA0869, 18BA2046, 18DA1022), DYKE1201C (17LA2024), DYKE1222B (18AA1458, 18DA0221), DYKE1307A (17JA0617, 17LA1157, 18AA0914, 18CA0967, 18CA1964, 18DA1473), DYKE1392 (17LA1416, 18AA0963, 18DA0376), DYKE1424A (18AA2437, 18CA0985, 18DA1013), DYKE1462A (18BA1818, 18DA1309, 18EAA071), DYKE1505A (17LA2183, 18DA1017), DYKE1533 (17JA0051, 18CA1355), DYKE1581 (18EAA459); d. Item Number DYK1002SBC, kit Number (Kit Lot Number): DYKE1098A (17KA0949, 17LA0422, 18BA0284, 18CA1966, 18DA1296, 18EAA950), DYKE1105 (17JA2744, 17KA1969, 18BB5818, 18EBU560), DYKE1294 (18AA0754), DYKE1363 (17LA0251), DYKE1493 (17KA2116), DYKE1560 (18BA1585, 18DA0003)
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, VA, VT, WA, WI, WV, WY
183,606 devices total