FDA Enforcement Class II Ongoing

Evolution Upgrade 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: (1) 782117 (2) 782143

Recall: Z-3193-2024 · Reported October 2, 2024

Enforcement

Recall Number
Z-3193-2024
Event ID
95289
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips North America Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 2, 2024
Initiation Date
September 3, 2024
Classification Date
September 20, 2024
Address
222 Jacobs St, Cambridge, MA, 02141-2289, United States

Description

Evolution Upgrade 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: (1) 782117 (2) 782143

Reason

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

Code Info

782117 (01)00884838099739(21) 782143 (01)00884838108660(21) Serial Numbers: 85675 85687 85695

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.

Quantity

3 units