FDA Enforcement Class II Terminated

PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.

Recall: Z-3186-2018 · Reported September 26, 2018

Enforcement

Recall Number
Z-3186-2018
Event ID
80724
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Polymedco, Inc.,/Polymedco Cancer Diagnostic Products, Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 26, 2018
Initiation Date
April 30, 2018
Classification Date
September 17, 2018
Termination Date
April 17, 2019
Address
510 Furnace Dock Rd, N/A, Cortlandt Manor, NY, 10567-6220, United States

Description

PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.

Reason

A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.

Code Info

Lot : A735 Exp: 10/31/2018 UDI: (01)04987595301636(17)181031(10)A735 and Lot: O736 Exp: 01/03/2019 UDI: (01)04987595301636(17)190131(10)O736

Distribution

US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.

Quantity

272 KITS in total