FDA Enforcement
Class II
Terminated
ECT Internal Fracture Fixation Drill Bits 1.1 mm Diameter, Item Number/EDI 00231801106, Nonsterile. orthopedic surgical instrument.
Recall: Z-3161-2017
·
Reported October 4, 2017
Enforcement
- Recall Number
- Z-3161-2017
- Event ID
- 77417
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer Biomet, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- October 4, 2017
- Initiation Date
- May 22, 2017
- Classification Date
- September 22, 2017
- Termination Date
- March 29, 2019
- Address
- 1800 W Center St, N/A, Warsaw, IN, 46580-2304, United States
Description
ECT Internal Fracture Fixation Drill Bits 1.1 mm Diameter, Item Number/EDI 00231801106, Nonsterile. orthopedic surgical instrument.
Reason
Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.
Code Info
Lot Numbers: 63562129, 63578148, 63377521, 63398755, 63415317, 63431736, 63438195, 63438272, 63444661, 63462322, 63474385, 63485128, 63497202, 63505360, 63510813, 63524432 & 63530407.
Distribution
Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.
Quantity
439 units