VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria

Recall: Z-3134-2017 · Reported September 27, 2017

Enforcement

Recall Number
Z-3134-2017
Event ID
77935
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Biomerieux Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 27, 2017
Initiation Date
August 25, 2017
Classification Date
September 18, 2017
Termination Date
November 9, 2020
Address
595 Anglum Rd, N/A, Hazelwood, MO, 63042-2320, United States

Description

VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria

Reason

Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC¿ 700327(TM) strain is positive.

Code Info

Lot Number 2420240403, Exp. 24-JUL-18, 2420303403, Exp 24-SEP-2018

Distribution

US and Angola, Argentina, Austria, Belarus, Belgium, Burkina Faso, Cameroon, Canada, China, Congo, Croatia, Egypt, France, Germany, Greece, Guam, Hungary, Israel, Italy, Latvia, Luxembourg, Macedonia, Mali, Mexico, Morocco, Netherlands, Nicaragua, Oman, Panama, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Swaziland, Switzerland, Tunisia, Turkey, Turkmenistan, United Arab Emirates, United Kingdom

Quantity

11,703 cartons