FDA Enforcement Class II Terminated

Firebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A Decorticating Planer is used to decorticate bone surrounding the spherical head of an implanted bone screw in order to provide clearance to allow for proper seating of the modular body, as well as to provide full range of modular body angulation. It is placed axially over the spherical head of the bone screw and rotated clockwise and counterclockwise to achieve the needed decortication

Recall: Z-3127-2017 · Reported September 20, 2017

Enforcement

Recall Number
Z-3127-2017
Event ID
77882
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Orthofix, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 20, 2017
Initiation Date
August 2, 2017
Classification Date
September 12, 2017
Termination Date
May 4, 2018
Address
3451 Plano Pkwy, N/A, Lewisville, TX, 75056-9453, United States

Description

Firebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A Decorticating Planer is used to decorticate bone surrounding the spherical head of an implanted bone screw in order to provide clearance to allow for proper seating of the modular body, as well as to provide full range of modular body angulation. It is placed axially over the spherical head of the bone screw and rotated clockwise and counterclockwise to achieve the needed decortication

Reason

The plastic core of the Decorticating Planer was found to be missing, which could result in scoring of the bone screw neck if used at or beyond a 23 degree angle.

Code Info

Lot #203013-TD26

Distribution

Distribution was made to distributors located in CA, FL, and RI. There was no foreign/military/government distribution.

Quantity

6 planers