FDA Enforcement
Class II
Terminated
Affixus¿ Hip Fracture Nail Right 130¿ 9 mm x 440 mm, Item Number: 814509440
Recall: Z-3118-2018
·
Reported September 19, 2018
Enforcement
- Recall Number
- Z-3118-2018
- Event ID
- 80755
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer Biomet, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 19, 2018
- Initiation Date
- July 17, 2018
- Classification Date
- September 13, 2018
- Termination Date
- June 2, 2020
- Address
- 56 E Bell Dr, N/A, Warsaw, IN, 46582-6989, United States
Description
Affixus¿ Hip Fracture Nail Right 130¿ 9 mm x 440 mm, Item Number: 814509440
Reason
The set screw is not able to advance or reverse from the original position in the nail.
Code Info
054150 054170 130740 130750 188310 519480 472970 503970
Distribution
The products were distributed to the following US states: AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WY, and Puerto Rico. The products were distributed to the following foreign countries: Chile, Ecuador, Japan, Netherlands.
Quantity
5947 total