FDA Enforcement
Class II
Terminated
Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United Kingdom, and Sekisui Diagnostics P.E.I. Inc., Canada. Used in laboratory testing.
Recall: Z-3118-2017
·
Reported September 20, 2017
Enforcement
- Recall Number
- Z-3118-2017
- Event ID
- 77896
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- September 20, 2017
- Initiation Date
- July 20, 2017
- Classification Date
- September 12, 2017
- Termination Date
- January 4, 2018
- Address
- 250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States
Description
Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United Kingdom, and Sekisui Diagnostics P.E.I. Inc., Canada. Used in laboratory testing.
Reason
Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification.
Code Info
All lots
Distribution
Distribution US nationwide.
Quantity
320 units