FDA Enforcement Class II Ongoing

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) BAPTIST FLOYD BRONCHOSCOPY, Pack Number DYKE1456D; b) ENT MINOR, Pack Number DYNJ908099B; c) HEAD & NECK, Pack Number DYNJ908101B; d) NASAL FESS, Pack Number DYNJ908105B; e) NASAL-LF, Pack Number DYNJ908429; f) MASTOPEXY-CANNOVA, Pack Number DYNJ909948

Recall: Z-3109-2024 · Reported September 18, 2024

Enforcement

Recall Number
Z-3109-2024
Event ID
95128
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
MEDLINE INDUSTRIES, LP - Northfield
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 18, 2024
Initiation Date
June 27, 2024
Classification Date
September 11, 2024
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) BAPTIST FLOYD BRONCHOSCOPY, Pack Number DYKE1456D; b) ENT MINOR, Pack Number DYNJ908099B; c) HEAD & NECK, Pack Number DYNJ908101B; d) NASAL FESS, Pack Number DYNJ908105B; e) NASAL-LF, Pack Number DYNJ908429; f) MASTOPEXY-CANNOVA, Pack Number DYNJ909948

Reason

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

Code Info

a) DYKE1456D: UDI/DI case 40195327639939, UDI/DI each 10195327639938, Lot Numbers: 24CBK289, 24BBK989; b) DYNJ908099B: UDI/DI case 40195327507085, UDI/DI each 10195327507084, Lot Numbers: 24DBI371, 24BBA188; c) DYNJ908101B: UDI/DI case 40195327507092, UDI/DI each 10195327507091, Lot Numbers: 24CBE408, 24BBA260, 24ABB112; d) DYNJ908105B: UDI/DI case 40195327507122, UDI/DI each 10195327507121, Lot Numbers: 24ABV362; e) DYNJ908429: UDI/DI case 40195327078820, UDI/DI each 10195327078829, Lot Numbers: 24CBB310, 23LBV007; f) DYNJ909948: UDI/DI case 40195327512911, UDI/DI each 10195327512910, Lot Numbers: 24DBD609

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Panama.

Quantity

407 units