Merge Unity software, formerly known as DR Systems Unity PACS software. The firm name on the label is Merge Healthcare, Hartland, WI. Merge Unity is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems

Recall: Z-3105-2017 · Reported September 13, 2017

Enforcement

Recall Number
Z-3105-2017
Event ID
77811
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Merge Healthcare, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 13, 2017
Initiation Date
August 14, 2017
Classification Date
September 5, 2017
Termination Date
May 8, 2019
Address
900 Walnut Ridge Dr, N/A, Hartland, WI, 53029-8347, United States

Description

Merge Unity software, formerly known as DR Systems Unity PACS software. The firm name on the label is Merge Healthcare, Hartland, WI. Merge Unity is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems

Reason

The software is not identifying the patient as having atypical hyperplasia, resulting in an incorrect Gail Risk calculation.

Code Info

Software versions 10.0, 10.0.1, 10.0.2, 10.0.3, 10.0.4, 10.0.5, 11.0, 11.0.1, 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1, 11.1.1.1, 11.1.1 Patch 2, 11.1.1 Patch 3, 11.1.2, 11.1.2 Patch 1, 11.1.2 Patch 2, 11.1.2 Patch 3, and 11.1.2 Patch 4. Only customers with one of the above listed software versions AND a license for Mammo Tracking are impacted by this recall.

Distribution

Nationwide Distribution

Quantity

98 sites have the potentially affected software