FDA Enforcement
Class II
Terminated
MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58
Recall: Z-3102-2017
·
Reported September 6, 2017
Enforcement
- Recall Number
- Z-3102-2017
- Event ID
- 77930
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Encore Medical, Lp
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- September 6, 2017
- Initiation Date
- August 4, 2017
- Classification Date
- August 30, 2017
- Termination Date
- June 28, 2023
- Address
- 9800 Metric Blvd, N/A, Austin, TX, 78758-5445, United States
Description
MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58
Reason
It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.
Code Info
Catalog number, Lot Number S-200956, 535U1000 S-200957, 536U1000 S-200958, 537U1000
Distribution
OH, VA
Quantity
5 units