Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese patients. Procedures that can be performed with the AXIOM Artis-zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis
Enforcement
- Recall Number
- Z-3099-2017
- Event ID
- 77908
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 6, 2017
- Initiation Date
- February 18, 2016
- Classification Date
- August 29, 2017
- Termination Date
- November 17, 2017
- Address
- 40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States
Description
Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese patients. Procedures that can be performed with the AXIOM Artis-zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Artis zeego.Q - Serial number: 117108 117001 117023 117021 117029 117006 117112 117034 117020 117018 117030 117026 117100 117111 117024 117113 117040 117041 117114 117014 117009
Nationwide Distribution
986 units in USA