smiths medical portex BLUselect, tracheostomy tube, labeled as follows: 1) BLUSELECT 6.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/060CZ; 2) BLUSELECT 7.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/070CZ; 3) BLUSELECT 7.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/075CZ; 4) BLUSELECT 8.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/080CZ; 5) BLUSELECT 8.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/085CZ; 6) BLUSELECT 9.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/090CZ; 7) BLUSELECT 10.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/100CZ; 8) BLUSELECT 6.0, CUFFED, FEN, 10/BX, Item Number 101/802/060CZ; 9) BLUSELECT 7.0, CUFFED, FEN, 10/BX, Item Number 101/802/070CZ; 10) BLUSELECT 7.5, CUFFED, FEN, 10/BX, Item Number 101/802/075CZ; 11) BLUSELECT 8.0, CUFFED, FEN, 10/BX, Item Number 101/802/080CZ; 12) BLUSELECT 8.5, CUFFED, FEN, 10/BX, Item Number 101/802/085CZ; 13) BLUSELECT 9.0, CUFFED, FEN, 10/BX, Item Number 101/802/090CZ; 14) BLUSELECT 10.0, CUFFED, FEN, 10/BX, Item Number 101/802/100CZ; 15) BLUSELECT 6.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/060CZ; 16) BLUSELECT 7.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/070CZ; 17) BLUSELECT 7.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/075; 18) BLUSELECT 7.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/075CZ; 19) BLUSELECT 8.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/080; 20) BLUSELECT 8.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/080CZ; 21) BLUSELECT 8.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/085CZ; 22) BLUSELECT 9.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/090CZ; 23) BLUSELECT 10.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/100CZ; 24) BLUSELECT 6.0, CUFFED, FEN, 1/EA, Item Number 101/812/060CZ; 25) BLUSELECT 7.0, CUFFED, FEN, 1/EA, Item Number 101/812/070CZ; 26) BLUSELECT 7.5, CUFFED, FEN, 1/EA, Item Number 101/812/075CZ; 27) BLUSELECT 8.0, CUFFED, FEN, 1/EA, Item Number 101/812/080CZ; 28) BLUSELECT 8.5, CUFFED, FEN, 1/EA, Item Number 101/812/085CZ; 29) BLUSELECT 9.0, CUFFED, FEN, 1/EA, Item Number 101/812/090CZ; 30) BLUSELECT 10.0, CUFFED, FEN, 1/EA, Item Number 101/812/100CZ; 31) BLUSELECT 6.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/060; 32) BLUSELECT 7.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/070; 33) BLUSELECT 7.5, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/075; 34) BLUSELECT 8.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/080; 35) BLUSELECT 8.5, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/085; 36) BLUSELECT 9.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/090; 37) BLUSELECT 10.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/100; 38) BLUSELECT 6.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/060; 39) BLUSELECT 7.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/070; 40) BLUSELECT 8.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/080; 41) BLUSELECT 9.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/090; 42) BLUSELECT 10.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/100; 43) BLUSELECT 6.0, SUCTIONAID, CUFFED, NON-FEN,10/BX, Item Number 101/860/060CZ; 44) BLUSELECT 7.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/070CZ; 45) BLUSELECT 7.5, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/075CZ; 46) BLUSELECT 8.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/080CZ; 47) BLUSELECT 8.5, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/085CZ; 48) BLUSELECT 9.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/090CZ; 49) BLUSELECT 10.0, SUCTIONAID, CUFFED, NON-FEN,10/BX, Item Number 101/860/100CZ; 50) BLUSELECT 6.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/060CZ; 51) BLUSELECT 7.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/070CZ; 52) BLUSELECT 7.5, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/075CZ; 53) BLUSELECT 8.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/0
Enforcement
- Recall Number
- Z-3086-2024
- Event ID
- 95016
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Smiths Medical ASD Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 18, 2024
- Initiation Date
- June 7, 2024
- Classification Date
- September 11, 2024
- Address
- 6000 Nathan Ln N, Minneapolis, MN, 55442-1690, United States
Description
smiths medical portex BLUselect, tracheostomy tube, labeled as follows: 1) BLUSELECT 6.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/060CZ; 2) BLUSELECT 7.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/070CZ; 3) BLUSELECT 7.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/075CZ; 4) BLUSELECT 8.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/080CZ; 5) BLUSELECT 8.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/085CZ; 6) BLUSELECT 9.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/090CZ; 7) BLUSELECT 10.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/100CZ; 8) BLUSELECT 6.0, CUFFED, FEN, 10/BX, Item Number 101/802/060CZ; 9) BLUSELECT 7.0, CUFFED, FEN, 10/BX, Item Number 101/802/070CZ; 10) BLUSELECT 7.5, CUFFED, FEN, 10/BX, Item Number 101/802/075CZ; 11) BLUSELECT 8.0, CUFFED, FEN, 10/BX, Item Number 101/802/080CZ; 12) BLUSELECT 8.5, CUFFED, FEN, 10/BX, Item Number 101/802/085CZ; 13) BLUSELECT 9.0, CUFFED, FEN, 10/BX, Item Number 101/802/090CZ; 14) BLUSELECT 10.0, CUFFED, FEN, 10/BX, Item Number 101/802/100CZ; 15) BLUSELECT 6.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/060CZ; 16) BLUSELECT 7.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/070CZ; 17) BLUSELECT 7.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/075; 18) BLUSELECT 7.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/075CZ; 19) BLUSELECT 8.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/080; 20) BLUSELECT 8.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/080CZ; 21) BLUSELECT 8.5, CUFFED, NON-FEN, 1/EA, Item Number 101/810/085CZ; 22) BLUSELECT 9.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/090CZ; 23) BLUSELECT 10.0, CUFFED, NON-FEN, 1/EA, Item Number 101/810/100CZ; 24) BLUSELECT 6.0, CUFFED, FEN, 1/EA, Item Number 101/812/060CZ; 25) BLUSELECT 7.0, CUFFED, FEN, 1/EA, Item Number 101/812/070CZ; 26) BLUSELECT 7.5, CUFFED, FEN, 1/EA, Item Number 101/812/075CZ; 27) BLUSELECT 8.0, CUFFED, FEN, 1/EA, Item Number 101/812/080CZ; 28) BLUSELECT 8.5, CUFFED, FEN, 1/EA, Item Number 101/812/085CZ; 29) BLUSELECT 9.0, CUFFED, FEN, 1/EA, Item Number 101/812/090CZ; 30) BLUSELECT 10.0, CUFFED, FEN, 1/EA, Item Number 101/812/100CZ; 31) BLUSELECT 6.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/060; 32) BLUSELECT 7.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/070; 33) BLUSELECT 7.5, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/075; 34) BLUSELECT 8.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/080; 35) BLUSELECT 8.5, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/085; 36) BLUSELECT 9.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/090; 37) BLUSELECT 10.0, CUFFED, NON-FEN, W/WEDGE, 1/EA, Item Number 101/815/100; 38) BLUSELECT 6.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/060; 39) BLUSELECT 7.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/070; 40) BLUSELECT 8.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/080; 41) BLUSELECT 9.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/090; 42) BLUSELECT 10.0, CUFFED, FEN, W/WEDGE AND DECAN CAP, 1/EA, Item Number 101/817/100; 43) BLUSELECT 6.0, SUCTIONAID, CUFFED, NON-FEN,10/BX, Item Number 101/860/060CZ; 44) BLUSELECT 7.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/070CZ; 45) BLUSELECT 7.5, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/075CZ; 46) BLUSELECT 8.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/080CZ; 47) BLUSELECT 8.5, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/085CZ; 48) BLUSELECT 9.0, SUCTIONAID, CUFFED, NON-FEN, 10/BX, Item Number 101/860/090CZ; 49) BLUSELECT 10.0, SUCTIONAID, CUFFED, NON-FEN,10/BX, Item Number 101/860/100CZ; 50) BLUSELECT 6.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/060CZ; 51) BLUSELECT 7.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/070CZ; 52) BLUSELECT 7.5, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/075CZ; 53) BLUSELECT 8.0, SUCTIONAID, CUFFED, NON-FEN, 1/EA, Item Number 101/870/0
There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.
101/800/060CZ, Lot Number 3895097; 101/800/060CZ, Lot Number 3925835; 101/800/060CZ, Lot Number 4025600; 101/800/060CZ, Lot Number 4106716; 101/800/060CZ, Lot Number 4230149; 101/800/060CZ, Lot Number 4252500; 101/800/060CZ, Lot Number 4382002; 101/800/070CZ, Lot Number 3895101; 101/800/070CZ, Lot Number 3925960; 101/800/070CZ, Lot Number 3933088; 101/800/070CZ, Lot Number 4010050; 101/800/070CZ, Lot Number 4022790; 101/800/070CZ, Lot Number 4034826; 101/800/070CZ, Lot Number 4037975; 101/800/070CZ, Lot Number 4133612; 101/800/070CZ, Lot Number 4152440; 101/800/070CZ, Lot Number 4156971; 101/800/070CZ, Lot Number 4217804; 101/800/070CZ, Lot Number 4238237; 101/800/070CZ, Lot Number 4249341; 101/800/070CZ, Lot Number 4305944; 101/800/070CZ, Lot Number 4305947; 101/800/070CZ, Lot Number 4368641; 101/800/075CZ, Lot Number 3895100; 101/800/075CZ, Lot Number 3925748; 101/800/075CZ, Lot Number 3933086; 101/800/075CZ, Lot Number 4025599; 101/800/075CZ, Lot Number 4034825; 101/800/075CZ, Lot Number 4037974; 101/800/075CZ, Lot Number 4073768; 101/800/075CZ, Lot Number 4117927; 101/800/075CZ, Lot Number 4134827; 101/800/075CZ, Lot Number 4152439; 101/800/075CZ, Lot Number 4156970; 101/800/075CZ, Lot Number 4238234; 101/800/075CZ, Lot Number 4249234; 101/800/075CZ, Lot Number 4348898; 101/800/075CZ, Lot Number 4348899; 101/800/075CZ, Lot Number 4365892; 101/800/080CZ, Lot Number 3895103; 101/800/080CZ, Lot Number 3909670; 101/800/080CZ, Lot Number 3933044; 101/800/080CZ, Lot Number 4037971; 101/800/080CZ, Lot Number 4052621; 101/800/080CZ, Lot Number 4138225; 101/800/080CZ, Lot Number 4156969; 101/800/080CZ, Lot Number 4224695; 101/800/080CZ, Lot Number 4230078; 101/800/080CZ, Lot Number 4232686; 101/800/080CZ, Lot Number 4245469; 101/800/080CZ, Lot Number 4247407; 101/800/080CZ, Lot Number 4270931; 101/800/080CZ, Lot Number 4291190; 101/800/080CZ, Lot Number 4305933; 101/800/080CZ, Lot Number 4305934; 101/800/080CZ, Lot Number 4305935; 101/800/080CZ, Lot Number 4305936; 101/800/080CZ, Lot Number 4305938; 101/800/085CZ, Lot Number 3922027; 101/800/085CZ, Lot Number 3951221; 101/800/085CZ, Lot Number 4034824; 101/800/085CZ, Lot Number 4256383; 101/800/085CZ, Lot Number 4348861; 101/800/090CZ, Lot Number 3895099; 101/800/090CZ, Lot Number 3951220; 101/800/090CZ, Lot Number 3968406; 101/800/090CZ, Lot Number 4039797; 101/800/090CZ, Lot Number 4197025; 101/800/090CZ, Lot Number 4217803; 101/800/090CZ, Lot Number 4238236; 101/800/090CZ, Lot Number 4261648; 101/800/090CZ, Lot Number 4273231; 101/800/090CZ, Lot Number 4306047; 101/800/090CZ, Lot Number 4310439; 101/800/090CZ, Lot Number 4376612; 101/800/090CZ, Lot Number 4387369; 101/800/100CZ, Lot Number 3895110; 101/800/100CZ, Lot Number 3951219; 101/800/100CZ, Lot Number 4000129; 101/800/100CZ, Lot Number 4230076; 101/800/100CZ, Lot Number 4306058; 101/800/100CZ, Lot Number 4372161; 101/800/100CZ, Lot Number 4386600; 101/800/100CZ, Lot Number 4390744; 101/800/100CZ, Lot Number 4413606; 101/802/060CZ, Lot Number 3906886; 101/802/060CZ, Lot Number 3925839; 101/802/060CZ, Lot Number 4028452; 101/802/060CZ, Lot Number 4382046; 101/802/070CZ, Lot Number 3895109; 101/802/070CZ, Lot Number 3925740; 101/802/070CZ, Lot Number 3972531; 101/802/075CZ, Lot Number 3895104; 101/802/075CZ, Lot Number 3951183; 101/802/075CZ, Lot Number 3968265; 101/802/075CZ, Lot Number 4388729; 101/802/080CZ, Lot Number 3895105; 101/802/080CZ, Lot Number 3951185; 101/802/080CZ, Lot Number 4268672; 101/802/080CZ, Lot Number 4310441; 101/802/080CZ, Lot Number 4423828; 101/802/080CZ, Lot Number 4436040; 101/802/080CZ, Lot Number 4444762; 101/802/080CZ, Lot Number 4456953; 101/802/085CZ, Lot Number 3922029; 101/802/085CZ, Lot Number 3925759; 101/802/085CZ, Lot Number 4028454; 101/802/090CZ, Lot Number 3895106; 101/802/090CZ, Lot Number 3925752; 101/802/090CZ, Lot Number 3954847; 101/802/100CZ, Lot Number 3910328; 101/802/100CZ, Lot Number 3951214; 101/802/100CZ, Lot Number 3968282; 101/810/060CZ, Lot Number 3887495; 101/810/060CZ, Lot Number 3933190; 101/810/060CZ, Lot Number 3948313; 101/810/060CZ, Lot Number 3958315; 101/810/060CZ, Lot Number 4106722; 101/810/060CZ, Lot Number 4268673; 101/810/070CZ, Lot Number 3887478; 101/810/070CZ, Lot Number 4106813; 101/810/070CZ, Lot Number 4243097; 101/810/070CZ, Lot Number 4254571; 101/810/070CZ, Lot Number 4285856; 101/810/075, Lot Number 4271670; 101/810/075CZ, Lot Number 3887504; 101/810/075CZ, Lot Number 3925795; 101/810/075CZ, Lot Number 4010055; 101/810/075CZ, Lot Number 4073763; 101/810/075CZ, Lot Number 4177908; 101/810/075CZ, Lot Number 4222083; 101/810/075CZ, Lot Number 4249233; 101/810/075CZ, Lot Number 4281153; 101/810/075CZ, Lot Number 4381994; 101/810/075CZ, Lot Number 4412129; 101/810/080, Lot Number 3900816; 101/810/080CZ, Lot Number 3887481; 101/810/080CZ, Lot Number 4173265; 101/810/080CZ, Lot Number 4207825; 101/810/080CZ, Lot Number 4217114; 101/810/080CZ, Lot Number 4230143; 101/810/080CZ, Lot Number 4285773; 101/810/080CZ, Lot Number 4306016; 101/810/080CZ, Lot Number 4312272; 101/810/080CZ, Lot Number 4381980; 101/810/080CZ, Lot Number 4387365; 101/810/080CZ, Lot Number 4403461; 101/810/080CZ, Lot Number 4415817; 101/810/080CZ, Lot Number 4423832; 101/810/085CZ, Lot Number 3922028; 101/810/085CZ, Lot Number 3951211; 101/810/085CZ, Lot Number 3989280; 101/810/085CZ, Lot Number 4067915; 101/810/085CZ, Lot Number 4098973; 101/810/085CZ, Lot Number 4152435; 101/810/085CZ, Lot Number 4178022; 101/810/085CZ, Lot Number 4281079; 101/810/085CZ, Lot Number 4306011; 101/810/090CZ, Lot Number 3887482; 101/810/090CZ, Lot Number 3993623; 101/810/090CZ, Lot Number 4106736; 101/810/090CZ, Lot Number 4173266; 101/810/090CZ, Lot Number 4243117; 101/810/090CZ, Lot Number 4306027; 101/810/090CZ, Lot Number 4327669; 101/810/090CZ, Lot Number 4403367; 101/810/090CZ, Lot Number 4426600; 101/810/090CZ, Lot Number 4458561; 101/810/100CZ, Lot Number 3887505; 101/810/100CZ, Lot Number 3909242; 101/810/100CZ, Lot Number 4058669; 101/810/100CZ, Lot Number 4175497; 101/810/100CZ, Lot Number 4217875; 101/810/100CZ, Lot Number 4249333; 101/810/100CZ, Lot Number 4291208; 101/810/100CZ, Lot Number 4310443; 101/810/100CZ, Lot Number 4312263; 101/810/100CZ, Lot Number 4387368; 101/810/100CZ, Lot Number 4397557; 101/812/060CZ, Lot Number 3887479; 101/812/060CZ, Lot Number 3925779; 101/812/060CZ, Lot Number 3959080; 101/812/060CZ, Lot Number 3976752; 101/812/060CZ, Lot Number 4106752; 101/812/060CZ, Lot Number 4294444; 101/812/070CZ, Lot Number 3887499; 101/812/070CZ, Lot Number 3925781; 101/812/070CZ, Lot Number 3948188; 101/812/070CZ, Lot Number 3968279; 101/812/070CZ, Lot Number 4106754; 101/812/070CZ, Lot Number 4242910; 101/812/075CZ, Lot Number 3887552; 101/812/075CZ, Lot Number 3968280; 101/812/075CZ, Lot Number 4217831; 101/812/075CZ, Lot Number 4268677; 101/812/075CZ, Lot Number 4381979; 101/812/075CZ, Lot Number 4413594; 101/812/080CZ, Lot Number 3887488; 101/812/080CZ, Lot Number 3948295; 101/812/080CZ, Lot Number 3989255; 101/812/080CZ, Lot Number 4073770; 101/812/080CZ, Lot Number 4106756; 101/812/080CZ, Lot Number 4227246; 101/812/080CZ, Lot Number 4234757; 101/812/080CZ, Lot Number 4385065; 101/812/085CZ, Lot Number 3922030; 101/812/085CZ, Lot Number 3925793; 101/812/085CZ, Lot Number 4028504; 101/812/085CZ, Lot Number 4191164; 101/812/085CZ, Lot Number 4222085; 101/812/085CZ, Lot Number 4261649; 101/812/085CZ, Lot Number 4382043; 101/812/090CZ, Lot Number 3887489; 101/812/090CZ, Lot Number 3972311; 101/812/090CZ, Lot Number 4075795; 101/812/090CZ, Lot Number 4106757; 101/812/090CZ, Lot Number 4170983; 101/812/090CZ, Lot Number 4217120; 101/812/090CZ, Lot Number 4243118; 101/812/100CZ, Lot Number 3887507; 101/812/100CZ, Lot Number 3948187; 101/812/100CZ, Lot Number 4064797; 101/812/100CZ, Lot Number 4245460; 101/812/100CZ, Lot Number 4259283; 101/812/100CZ, Lot Number 4381995; 101/812/100CZ, Lot Number 4388732; 101/812/100CZ, Lot Number 4413618; 101/812/100CZ, Lot Number 4423838; 101/812/100CZ, Lot Number 4443290; 101/815/060, Lot Number 3933722; 101/815/060, Lot Number 3940999; 101/815/060, Lot Number 3949223; 101/815/060, Lot Number 4014730; 101/815/060, Lot Number 4021067; 101/815/060, Lot Number 4035539; 101/815/060, Lot Number 4091934; 101/815/060, Lot Number 4118283; 101/815/060, Lot Number 4121883; 101/815/060, Lot Number 4159150; 101/815/060, Lot Number 4164109; 101/815/060, Lot Number 4166710; 101/815/060, Lot Number 4174793; 101/815/060, Lot Number 4177293; 101/815/060, Lot Number 4177295; 101/815/060, Lot Number 4187097; 101/815/060, Lot Number 4269270; 101/815/060, Lot Number 4369411; 101/815/060, Lot Number 4369412; 101/815/060, Lot Number 4380967; 101/815/070, Lot Number 3905154; 101/815/070, Lot Number 3949269; 101/815/070, Lot Number 3965244; 101/815/070, Lot Number 4032511; 101/815/070, Lot Number 4039241; 101/815/070, Lot Number 4091935; 101/815/070, Lot Number 4115649; 101/815/070, Lot Number 4121851; 101/815/070, Lot Number 4131921; 101/815/070, Lot Number 4187074; 101/815/070, Lot Number 4187085; 101/815/070, Lot Number 4196637; 101/815/070, Lot Number 4214909; 101/815/070, Lot Number 4219471; 101/815/070, Lot Number 4220399; 101/815/070, Lot Number 4223153; 101/815/070, Lot Number 4260338; 101/815/070, Lot Number 4277368; 101/815/070, Lot Number 4292624; 101/815/070, Lot Number 4294949; 101/815/070, Lot Number 4306727; 101/815/070, Lot Number 4306851; 101/815/070, Lot Number 4306852; 101/815/070, Lot Number 4332676; 101/815/070, Lot Number 4332678; 101/815/070, Lot Number 4332684; 101/815/070, Lot Number 4336640; 101/815/070, Lot Number 4369413; 101/815/075, Lot Number 3956297; 101/815/075, Lot Number 4071378; 101/815/075, Lot Number 4126686; 101/815/075, Lot Number 4174782; 101/815/075, Lot Number 4294945; 101/815/075, Lot Number 4332682; 101/815/075, Lot Number 4362485; 101/815/075, Lot Number 4414625; 101/815/075, Lot Number 4428012; 101/815/080, Lot Number 3905015; 101/815/080, Lot Number 3940109; 101/815/080, Lot Number 3942887; 101/815/080, Lot Number 3965245; 101/815/080, Lot Number 4021055; 101/815/080, Lot Number 4029161; 101/815/080, Lot Number 4058347; 101/815/080, Lot Number 4100474; 101/815/080, Lot Number 4118284; 101/815/080, Lot Number 4169838; 101/815/080, Lot Number 4187081; 101/815/080, Lot Number 4190342; 101/815/080, Lot Number 4192320; 101/815/080, Lot Number 4202303; 101/815/080, Lot Number 4204265; 101/815/080, Lot Number 4204266; 101/815/080, Lot Number 4206466; 101/815/080, Lot Number 4209172; 101/815/080, Lot Number 4274832; 101/815/080, Lot Number 4294948; 101/815/080, Lot Number 4336646; 101/815/080, Lot Number 4336649; 101/815/080, Lot Number 4336668; 101/815/085, Lot Number 3956296; 101/815/085, Lot Number 4071392; 101/815/085, Lot Number 4090264; 101/815/085, Lot Number 4209173; 101/815/085, Lot Number 4332683; 101/815/085, Lot Number 4369414; 101/815/090, Lot Number 3933809; 101/815/090, Lot Number 3949146; 101/815/090, Lot Number 4011017; 101/815/090, Lot Number 4021088; 101/815/090, Lot Number 4026491; 101/815/090, Lot Number 4076707; 101/815/090, Lot Number 4159151; 101/815/090, Lot Number 4169840; 101/815/090, Lot Number 4169841; 101/815/090, Lot Number 4178241; 101/815/090, Lot Number 4185327; 101/815/090, Lot Number 4200170; 101/815/090, Lot Number 4200171; 101/815/090, Lot Number 4204267; 101/815/090, Lot Number 4212316; 101/815/090, Lot Number 4216557; 101/815/090, Lot Number 4260339; 101/815/090, Lot Number 4279525; 101/815/090, Lot Number 4294947; 101/815/090, Lot Number 4334212; 101/815/090, Lot Number 4336647; 101/815/090, Lot Number 4336650; 101/815/100, Lot Number 3933697; 101/815/100, Lot Number 4021081; 101/815/100, Lot Number 4029154; 101/815/100, Lot Number 4112822; 101/815/100, Lot Number 4148895; 101/815/100, Lot Number 4161638; 101/815/100, Lot Number 4169842; 101/815/100, Lot Number 4187096; 101/815/100, Lot Number 4192321; 101/815/100, Lot Number 4206174; 101/815/100, Lot Number 4284378; 101/817/060, Lot Number 3905152; 101/817/060, Lot Number 4106973; 101/817/060, Lot Number 4174794; 101/817/070, Lot Number 3956189; 101/817/070, Lot Number 3986585; 101/817/070, Lot Number 4107000; 101/817/070, Lot Number 4118288; 101/817/070, Lot Number 4130527; 101/817/070, Lot Number 4187078; 101/817/070, Lot Number 4208001; 101/817/080, Lot Number 3933831; 101/817/080, Lot Number 3962715; 101/817/080, Lot Number 4062550; 101/817/080, Lot Number 4068195; 101/817/080, Lot Number 4230801; 101/817/080, Lot Number 4294937; 101/817/090, Lot Number 3941008; 101/817/090, Lot Number 4071380; 101/817/090, Lot Number 4076692; 101/817/090, Lot Number 4106651; 101/817/100, Lot Number 3933799; 101/817/100, Lot Number 4187080; 101/860/060CZ, Lot Number 3895092; 101/860/060CZ, Lot Number 3951193; 101/860/060CZ, Lot Number 4010035; 101/860/060CZ, Lot Number 4058593; 101/860/060CZ, Lot Number 4117932; 101/860/060CZ, Lot Number 4134876; 101/860/060CZ, Lot Number 4156952; 101/860/060CZ, Lot Number 4177861; 101/860/060CZ, Lot Number 4242923; 101/860/060CZ, Lot Number 4376593; 101/860/070CZ, Lot Number 3887509; 101/860/070CZ, Lot Number 3895108; 101/860/070CZ, Lot Number 3951303; 101/860/070CZ, Lot Number 3959094; 101/860/070CZ, Lot Number 3993578; 101/860/070CZ, Lot Number 4010067; 101/860/070CZ, Lot Number 4017181; 101/860/070CZ, Lot Number 4022712; 101/860/070CZ, Lot Number 4025593; 101/860/070CZ, Lot Number 4037923; 101/860/070CZ, Lot Number 4039786; 101/860/070CZ, Lot Number 4052498; 101/860/070CZ, Lot Number 4053320; 101/860/070CZ, Lot Number 4058579; 101/860/070CZ, Lot Number 4061184; 101/860/070CZ, Lot Number 4062929; 101/860/070CZ, Lot Number 4067465; 101/860/070CZ, Lot Number 4067814; 101/860/070CZ, Lot Number 4074986; 101/860/070CZ, Lot Number 4138104; 101/860/070CZ, Lot Number 4163119; 101/860/070CZ, Lot Number 4165742; 101/860/070CZ, Lot Number 4170834; 101/860/070CZ, Lot Number 4183961; 101/860/070CZ, Lot Number 4186645; 101/860/070CZ, Lot Number 4186963; 101/860/070CZ, Lot Number 4198020; 101/860/070CZ, Lot Number 4202488; 101/860/070CZ, Lot Number 4205530; 101/860/070CZ, Lot Number 4221990; 101/860/070CZ, Lot Number 4232682; 101/860/070CZ, Lot Number 4234708; 101/860/070CZ, Lot Number 4234709; 101/860/070CZ, Lot Number 4249306; 101/860/070CZ, Lot Number 4252463; 101/860/070CZ, Lot Number 4252464; 101/860/070CZ, Lot Number 4285771; 101/860/070CZ, Lot Number 4294329; 101/860/070CZ, Lot Number 4303383; 101/860/070CZ, Lot Number 4306006; 101/860/070CZ, Lot Number 4306007; 101/860/070CZ, Lot Number 4306008; 101/860/070CZ, Lot Number 4368707; 101/860/070CZ, Lot Number 4370061; 101/860/075CZ, Lot Number 3895094; 101/860/075CZ, Lot Number 3909675; 101/860/075CZ, Lot Number 3951196; 101/860/075CZ, Lot Number 3951197; 101/860/075CZ, Lot Number 3954753; 101/860/075CZ, Lot Number 4002904; 101/860/075CZ, Lot Number 4005269; 101/860/075CZ, Lot Number 4010074; 101/860/075CZ, Lot Number 4017180; 101/860/075CZ, Lot Number 4020216; 101/860/075CZ, Lot Number 4022713; 101/860/075CZ, Lot Number 4034797; 101/860/075CZ, Lot Number 4037922; 101/860/075CZ, Lot Number 4052614; 101/860/075CZ, Lot Number 4054784; 101/860/075CZ, Lot Number 4054785; 101/860/075CZ, Lot Number 4054787; 101/860/075CZ, Lot Number 4062434; 101/860/075CZ, Lot Number 4062931; 101/860/075CZ, Lot Number 4065368; 101/860/075CZ, Lot Number 4065369; 101/860/075CZ, Lot Number 4067466; 101/860/075CZ, Lot Number 4117933; 101/860/075CZ, Lot Number 4122876; 101/860/075CZ, Lot Number 4127325; 101/860/075CZ, Lot Number 4129462; 101/860/075CZ, Lot Number 4133614; 101/860/075CZ, Lot Number 4134878; 101/860/075CZ, Lot Number 4138103; 101/860/075CZ, Lot Number 4139223; 101/860/075CZ, Lot Number 4139224; 101/860/075CZ, Lot Number 4165615; 101/860/075CZ, Lot Number 4170859; 101/860/075CZ, Lot Number 4183962; 101/860/075CZ, Lot Number 4183963; 101/860/075CZ, Lot Number 4186583; 101/860/075CZ, Lot Number 4186965; 101/860/075CZ, Lot Number 4189388; 101/860/075CZ, Lot Number 4197067; 101/860/075CZ, Lot Number 4199269; 101/860/075CZ, Lot Number 4199270; 101/860/075CZ, Lot Number 4212066; 101/860/075CZ, Lot Number 4214633; 101/860/075CZ, Lot Number 4214634; 101/860/075CZ, Lot Number 4217813; 101/860/075CZ, Lot Number 4221051; 101/860/075CZ, Lot Number 4224578; 101/860/075CZ, Lot Number 4225227; 101/860/075CZ, Lot Number 4242918; 101/860/075CZ, Lot Number 4243105; 101/860/075CZ, Lot Number 4245286; 101/860/075CZ, Lot Number 4249223; 101/860/075CZ, Lot Number 4252465; 101/860/075CZ, Lot Number 4254572; 101/860/075CZ, Lot Number 4259236; 101/860/075CZ, Lot Number 4270939; 101/860/075CZ, Lot Number 4281157; 101/860/075CZ, Lot Number 4285842; 101/860/075CZ, Lot Number 4285843; 101/860/075CZ, Lot Number 4285844; 101/860/075CZ, Lot Number 4294330; 101/860/075CZ, Lot Number 4306004; 101/860/075CZ, Lot Number 4306005; 101/860/075CZ, Lot Number 4312287; 101/860/075CZ, Lot Number 4312288; 101/860/075CZ, Lot Number 4317594; 101/860/075CZ, Lot Number 4336058; 101/860/075CZ, Lot Number 4368643; 101/860/075CZ, Lot Number 4382006; 101/860/075CZ, Lot Number 4382007; 101/860/075CZ, Lot Number 4386514; 101/860/075CZ, Lot Number 4387362; 101/860/075CZ, Lot Number 4408613; 101/860/080CZ, Lot Number 3887546; 101/860/080CZ, Lot Number 3895095; 101/860/080CZ, Lot Number 3903503; 101/860/080CZ, Lot Number 3909276; 101/860/080CZ, Lot Number 3951327; 101/860/080CZ, Lot Number 3968390; 101/860/080CZ, Lot Number 3977665; 101/860/080CZ, Lot Number 3980383; 101/860/080CZ, Lot Number 3993592; 101/860/080CZ, Lot Number 4002998; 101/860/080CZ, Lot Number 4009812; 101/860/080CZ, Lot Number 4009826; 101/860/080CZ, Lot Number 4017177; 101/860/080CZ, Lot Number 4017243; 101/860/080CZ, Lot Number 4017244; 101/860/080CZ, Lot Number 4017245; 101/860/080CZ, Lot Number 4026070; 101/860/080CZ, Lot Number 4031414; 101/860/080CZ, Lot Number 4031415; 101/860/080CZ, Lot Number 4031416; 101/860/080CZ, Lot Number 4034842; 101/860/080CZ, Lot Number 4037140; 101/860/080CZ, Lot Number 4037141; 101/860/080CZ, Lot Number 4037986; 101/860/080CZ, Lot Number 4041420; 101/860/080CZ, Lot Number 4054804; 101/860/080CZ, Lot Number 4054805; 101/860/080CZ, Lot Number 4057098; 101/860/080CZ, Lot Number 4058607; 101/860/080CZ, Lot Number 4058608; 101/860/080CZ, Lot Number 4061058; 101/860/080CZ, Lot Number 4061061; 101/860/080CZ, Lot Number 4061062; 101/860/080CZ, Lot Number 4067791; 101/860/080CZ, Lot Number 4071181; 101/860/080CZ, Lot Number 4071182; 101/860/080CZ, Lot Number 4079446; 101/860/080CZ, Lot Number 4079447; 101/860/080CZ, Lot Number 4079448; 101/860/080CZ, Lot Number 4133622; 101/860/080CZ, Lot Number 4134843; 101/860/080CZ, Lot Number 4134844; 101/860/080CZ, Lot Number 4134845; 101/860/080CZ, Lot Number 4134846; 101/860/080CZ, Lot Number 4136573; 101/860/080CZ, Lot Number 4136574; 101/860/080CZ, Lot Number 4136575; 101/860/080CZ, Lot Number 4165613; 101/860/080CZ, Lot Number 4165614; 101/860/080CZ, Lot Number 4170920; 101/860/080CZ, Lot Number 4178020; 101/860/080CZ, Lot Number 4179046; 101/860/080CZ, Lot Number 4183972; 101/860/080CZ, Lot Number 4186574; 101/860/080CZ, Lot Number 4186575; 101/860/080CZ, Lot Number 4189343; 101/860/080CZ, Lot Number 4191167; 101/860/080CZ, Lot Number 4191168; 101/860/080CZ, Lot Number 4195032; 101/860/080CZ, Lot Number 4202463; 101/860/080CZ, Lot Number 4205543; 101/860/080CZ, Lot Number 4207746; 101/860/080CZ, Lot Number 4211996; 101/860/080CZ, Lot Number 4211997; 101/860/080CZ, Lot Number 4211998; 101/860/080CZ, Lot Number 4214635; 101/860/080CZ, Lot Number 4217853; 101/860/080CZ, Lot Number 4219600; 101/860/080CZ, Lot Number 4221985; 101/860/080CZ, Lot Number 4231321; 101/860/080CZ, Lot Number 4231322; 101/860/080CZ, Lot Number 4232674; 101/860/080CZ, Lot Number 4234707; 101/860/080CZ, Lot Number 4242937; 101/860/080CZ, Lot Number 4249310; 101/860/080CZ, Lot Number 4252471; 101/860/080CZ, Lot Number 4252472; 101/860/080CZ, Lot Number 4254577; 101/860/080CZ, Lot Number 4254578; 101/860/080CZ, Lot Number 4259268; 101/860/080CZ, Lot Number 4259269; 101/860/080CZ, Lot Number 4281080; 101/860/080CZ, Lot Number 4281081; 101/860/080CZ, Lot Number 4283735; 101/860/080CZ, Lot Number 4285774; 101/860/080CZ, Lot Number 4285775; 101/860/080CZ, Lot Number 4305958; 101/860/080CZ, Lot Number 4305959; 101/860/080CZ, Lot Number 4312278; 101/860/080CZ, Lot Number 4312279; 101/860/080CZ, Lot Number 4312280; 101/860/080CZ, Lot Number 4319745; 101/860/080CZ, Lot Number 4319746; 101/860/080CZ, Lot Number 4357891; 101/860/080CZ, Lot Number 4365949; 101/860/080CZ, Lot Number 4376685; 101/860/080CZ, Lot Number 4376686; 101/860/080CZ, Lot Number 4376687; 101/860/080CZ, Lot Number 4384987; 101/860/080CZ, Lot Number 4384988; 101/860/080CZ, Lot Number 4386575; 101/860/080CZ, Lot Number 4388728; 101/860/080CZ, Lot Number 4392808; 101/860/080CZ, Lot Number 4392809; 101/860/080CZ, Lot Number 4405273; 101/860/080CZ, Lot Number 4405274; 101/860/085CZ, Lot Number 3895107; 101/860/085CZ, Lot Number 3951199; 101/860/085CZ, Lot Number 3964928; 101/860/085CZ, Lot Number 4010054; 101/860/085CZ, Lot Number 4020273; 101/860/085CZ, Lot Number 4031338; 101/860/085CZ, Lot Number 4054788; 101/860/085CZ, Lot Number 4079454; 101/860/085CZ, Lot Number 4129463; 101/860/085CZ, Lot Number 4138224; 101/860/085CZ, Lot Number 4165743; 101/860/085CZ, Lot Number 4177836; 101/860/085CZ, Lot Number 4197019; 101/860/085CZ, Lot Number 4205591; 101/860/085CZ, Lot Number 4214079; 101/860/085CZ, Lot Number 4238235; 101/860/085CZ, Lot Number 4249237; 101/860/085CZ, Lot Number 4256364; 101/860/085CZ, Lot Number 4270940; 101/860/085CZ, Lot Number 4310532; 101/860/085CZ, Lot Number 4368644; 101/860/085CZ, Lot Number 4382005; 101/860/090CZ, Lot Number 3895091; 101/860/090CZ, Lot Number 3914422; 101/860/090CZ, Lot Number 3951202; 101/860/090CZ, Lot Number 3954853; 101/860/090CZ, Lot Number 3989279; 101/860/090CZ, Lot Number 4009810; 101/860/090CZ, Lot Number 4017242; 101/860/090CZ, Lot Number 4030604; 101/860/090CZ, Lot Number 4037924; 101/860/090CZ, Lot Number 4058605; 101/860/090CZ, Lot Number 4071183; 101/860/090CZ, Lot Number 4100371; 101/860/090CZ, Lot Number 4103825; 101/860/090CZ, Lot Number 4117879; 101/860/090CZ, Lot Number 4124564; 101/860/090CZ, Lot Number 4134881; 101/860/090CZ, Lot Number 4138175; 101/860/090CZ, Lot Number 4183964; 101/860/090CZ, Lot Number 4183965; 101/860/090CZ, Lot Number 4191506; 101/860/090CZ, Lot Number 4199354; 101/860/090CZ, Lot Number 4209002; 101/860/090CZ, Lot Number 4232675; 101/860/090CZ, Lot Number 4239848; 101/860/090CZ, Lot Number 4249235; 101/860/090CZ, Lot Number 4252466; 101/860/090CZ, Lot Number 4276630; 101/860/090CZ, Lot Number 4285839; 101/860/090CZ, Lot Number 4306063; 101/860/090CZ, Lot Number 4306064; 101/860/090CZ, Lot Number 4310533; 101/860/090CZ, Lot Number 4382009; 101/860/090CZ, Lot Number 4385029; 101/860/100CZ, Lot Number 3895093; 101/860/100CZ, Lot Number 3925786; 101/860/100CZ, Lot Number 3951305; 101/860/100CZ, Lot Number 3968281; 101/860/100CZ, Lot Number 4010004; 101/860/100CZ, Lot Number 4025595; 101/860/100CZ, Lot Number 4054789; 101/860/100CZ, Lot Number 4199355; 101/860/100CZ, Lot Number 4217881; 101/860/100CZ, Lot Number 4285765; 101/860/100CZ, Lot Number 4412135; 101/870/060CZ, Lot Number 3887510; 101/870/060CZ, Lot Number 3925734; 101/870/060CZ, Lot Number 3951306; 101/870/060CZ, Lot Number 4031337; 101/870/060CZ, Lot Number 4106745; 101/870/060CZ, Lot Number 4217882; 101/870/070CZ, Lot Number 3889311; 101/870/070CZ, Lot Number 3951204; 101/870/070CZ, Lot Number 3968404; 101/870/070CZ, Lot Number 4017237; 101/870/070CZ, Lot Number 4022711; 101/870/070CZ, Lot Number 4034792; 101/870/070CZ, Lot Number 4058662; 101/870/070CZ, Lot Number 4073762; 101/870/070CZ, Lot Number 4134828; 101/870/070CZ, Lot Number 4138214; 101/870/070CZ, Lot Number 4156953; 101/870/070CZ, Lot Number 4165744; 101/870/070CZ, Lot Number 4178001; 101/870/070CZ, Lot Number 4217826; 101/870/070CZ, Lot Number 4249291; 101/870/070CZ, Lot Number 4256336; 101/870/070CZ, Lot Number 4283719; 101/870/070CZ, Lot Number 4291171; 101/870/075CZ, Lot Number 3906881; 101/870/075CZ, Lot Number 3951207; 101/870/075CZ, Lot Number 3968405; 101/870/075CZ, Lot Number 4054813; 101/870/075CZ, Lot Number 4058606; 101/870/075CZ, Lot Number 4061022; 101/870/075CZ, Lot Number 4134829; 101/870/075CZ, Lot Number 4138176; 101/870/075CZ, Lot Number 4183954; 101/870/075CZ, Lot Number 4212004; 101/870/075CZ, Lot Number 4238185; 101/870/075CZ, Lot Number 4249236; 101/870/075CZ, Lot Number 4270941; 101/870/075CZ, Lot Number 4276631; 101/870/075CZ, Lot Number 4382010; 101/870/080CZ, Lot Number 3887515; 101/870/080CZ, Lot Number 3951288; 101/870/080CZ, Lot Number 3980351; 101/870/080CZ, Lot Number 3989180; 101/870/080CZ, Lot Number 4017250; 101/870/080CZ, Lot Number 4051928; 101/870/080CZ, Lot Number 4054796; 101/870/080CZ, Lot Number 4070522; 101/870/080CZ, Lot Number 4117931; 101/870/080CZ, Lot Number 4134866; 101/870/080CZ, Lot Number 4138056; 101/870/080CZ, Lot Number 4156930; 101/870/080CZ, Lot Number 4163106; 101/870/080CZ, Lot Number 4208997; 101/870/080CZ, Lot Number 4238168; 101/870/080CZ, Lot Number 4254564; 101/870/080CZ, Lot Number 4303380; 101/870/080CZ, Lot Number 4310523; 101/870/080CZ, Lot Number 4312285; 101/870/080CZ, Lot Number 4317581; 101/870/080CZ, Lot Number 4319800; 101/870/080CZ, Lot Number 4368642; 101/870/080CZ, Lot Number 4388736; 101/870/080CZ, Lot Number 4390747; 101/870/085CZ, Lot Number 3887484; 101/870/085CZ, Lot Number 3925788; 101/870/085CZ, Lot Number 4002986; 101/870/085CZ, Lot Number 4054842; 101/870/085CZ, Lot Number 4085217; 101/870/085CZ, Lot Number 4110784; 101/870/085CZ, Lot Number 4121517; 101/870/085CZ, Lot Number 4156983; 101/870/085CZ, Lot Number 4183981; 101/870/085CZ, Lot Number 4212070; 101/870/085CZ, Lot Number 4245285; 101/870/085CZ, Lot Number 4249238; 101/870/085CZ, Lot Number 4376585; 101/870/090CZ, Lot Number 3892136; 101/870/090CZ, Lot Number 3907683; 101/870/090CZ, Lot Number 3951191; 101/870/090CZ, Lot Number 3962340; 101/870/090CZ, Lot Number 4000188; 101/870/090CZ, Lot Number 4061185; 101/870/090CZ, Lot Number 4075797; 101/870/090CZ, Lot Number 4106832; 101/870/090CZ, Lot Number 4129451; 101/870/090CZ, Lot Number 4138105; 101/870/090CZ, Lot Number 4155508; 101/870/090CZ, Lot Number 4178002; 101/870/090CZ, Lot Number 4183974; 101/870/090CZ, Lot Number 4242948; 101/870/090CZ, Lot Number 4249307; 101/870/090CZ, Lot Number 4249334; 101/870/090CZ, Lot Number 4276656; 101/870/090CZ, Lot Number 4285763; 101/870/090CZ, Lot Number 4285764; 101/870/090CZ, Lot Number 4303381; 101/870/090CZ, Lot Number 4387367; 101/870/100CZ, Lot Number 3887465; 101/870/100CZ, Lot Number 3925771; 101/870/100CZ, Lot Number 3996000; 101/870/100CZ, Lot Number 4075798; 101/870/100CZ, Lot Number 4087868; 101/870/100CZ, Lot Number 4106718; 101/870/100CZ, Lot Number 4209008; 101/870/100CZ, Lot Number 4217806; 101/870/100CZ, Lot Number 4245461; 101/870/100CZ, Lot Number 4256343; 101/870/100CZ, Lot Number 4306010; 101/875/060, Lot Number 3973580; 101/875/060, Lot Number 3994206; 101/875/060, Lot Number 4077390; 101/875/060, Lot Number 4077409; 101/875/060, Lot Number 4097405; 101/875/060, Lot Number 4106652; 101/875/060, Lot Number 4121852; 101/875/060, Lot Number 4174792; 101/875/060, Lot Number 4185328; 101/875/060, Lot Number 4228918; 101/875/060, Lot Number 4294942; 101/875/060, Lot Number 4306861; 101/875/070, Lot Number 3965241; 101/875/070, Lot Number 3997476; 101/875/070, Lot Number 4040743; 101/875/070, Lot Number 4040744; 101/875/070, Lot Number 4040748; 101/875/070, Lot Number 4083147; 101/875/070, Lot Number 4090267; 101/875/070, Lot Number 4106653; 101/875/070, Lot Number 4106657; 101/875/070, Lot Number 4130528; 101/875/070, Lot Number 4219470; 101/875/070, Lot Number 4244044; 101/875/070, Lot Number 4244045; 101/875/075, Lot Number 3956207; 101/875/075, Lot Number 3965238; 101/875/075, Lot Number 4106984; 101/875/075, Lot Number 4230803; 101/875/075, Lot Number 4294936; 101/875/075, Lot Number 4318743; 101/875/075, Lot Number 4380983; 101/875/080, Lot Number 3905060; 101/875/080, Lot Number 4038638; 101/875/080, Lot Number 4071393; 101/875/080, Lot Number 4083162; 101/875/080, Lot Number 4088887; 101/875/080, Lot Number 4091522; 101/875/080, Lot Number 4106655; 101/875/080, Lot Number 4124771; 101/875/080, Lot Number 4187079; 101/875/080, Lot Number 4187084; 101/875/080, Lot Number 4202304; 101/875/080, Lot Number 4220404; 101/875/080, Lot Number 4279527; 101/875/080, Lot Number 4294941; 101/875/080, Lot Number 4375952; 101/875/085, Lot Number 3956339; 101/875/085, Lot Number 4106981; 101/875/085, Lot Number 4119879; 101/875/085, Lot Number 4187077; 101/875/085, Lot Number 4202297; 101/875/085, Lot Number 4216554; 101/875/090, Lot Number 4034431; 101/875/090, Lot Number 4067217; 101/875/090, Lot Number 4068186; 101/875/090, Lot Number 4106650; 101/875/090, Lot Number 4106999; 101/875/090, Lot Number 4115631; 101/875/090, Lot Number 4118290; 101/875/090, Lot Number 4187086; 101/875/090, Lot Number 4206467; 101/875/090, Lot Number 4216550; 101/875/090, Lot Number 4220407; 101/875/100, Lot Number 3905062
Worldwide
799732 units