FDA Enforcement Class II Terminated

LX3 Floor Stand for LuxOR Ophthalmic Microscope; Endure LuxOR Q-VUE Ophthalmic Microscope. LX3 Microscope Floor Stands are not stand-alone medical devices, but are used in combination with the LuxOR Ophthalmic Microscope and the LuxOR Ophthalmic Microscope with Q-VUE 3-D Assistant Visualization. These microscopes are used for low magnification visualization during cataract, retina, and cornea surgeries

Recall: Z-3081-2017 · Reported September 6, 2017

Enforcement

Recall Number
Z-3081-2017
Event ID
77480
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Alcon Research, Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 6, 2017
Initiation Date
May 25, 2017
Classification Date
August 28, 2017
Termination Date
February 4, 2022
Address
6201 South Fwy, N/A, Fort Worth, TX, 76134-2099, United States

Description

LX3 Floor Stand for LuxOR Ophthalmic Microscope; Endure LuxOR Q-VUE Ophthalmic Microscope. LX3 Microscope Floor Stands are not stand-alone medical devices, but are used in combination with the LuxOR Ophthalmic Microscope and the LuxOR Ophthalmic Microscope with Q-VUE 3-D Assistant Visualization. These microscopes are used for low magnification visualization during cataract, retina, and cornea surgeries

Reason

Multiple shipments or forceful impact to a caster on a LuxOr/LX3 microscope may stress the microscope base, which could contribute to a caster breaking or detaching from the base.

Code Info

Domestic Serial Numbers: 1402548501X, 1402548502X, 1402548503X, 1402548504X, 1402548505X, 1402548506X, 1402577401X, 1402577402X, 1402577403X, 1402577404X, 1402577405X, 1402577406X, 1402612001X, 1402612002X, 1402612003X, 1402612004X, 1402612005X, 1402612006X, 1402655703X, 1402655704X, 1402655705X, 1402655706X, 1402705203X, 1402705204X, 1402728703X, 1402728704X, 1402728705X, 1402728706X, 1402804805X, 1402804806X, 1402887601X, 1402887602X, 1402887603X, 1402887604X, 1402959002X, 1402959004X, 1402959005X, 1402959006X, 1403012601X, 1403012602X, 1403012603X, 1403012604X, 1403012605X, 1403012606X, 1403045903X, 1403045904X, 1403045905X, 1403045906X, 1403072801X, 1403072802X, 1403072803X, 1403072804X, 1403072805X, 1403072806X, 1403124901X, 1403124902X, 1403124903X, 1403124904X, 1403124905X, 1403124906X, 1403145301X, 1403145302X, 1403145303X, 1403145304X, 1403145305X, 1403145306X, 1403173401X, 1403173402X, 1403173403X, 1403173404X, 1403173405X, 1403173406X, 1403211701X, 1403211702X, 1403211704X, 1403211706X, 1403239601X, 1403239602X, 1403239603X, 1403239604X, 1403239605X, 1403239606X, 1403253702X, 1403253703X, 1403253705X, 1403253706X, 1403272701X, 1403272702X, 1403351301X, 1403351302X, 1403351303X, 1403351304X, 1403351305X, 1403351306X, 1403390506X, 1403390803X, 1403390805X, 1403440402X, 1403440406X, 1403440502X, 1403440503X, 1403440505X, 1403440506X, 1403484502X, 1403484602X, 1403484603X, 1403484604X, 1403539702X, 1403539703X, 1403539704X, 1403539706X, 1403625401X, 1403625402X, 1403625403X, 1403625404X, 1403625405X, 1403625406X, 1403625501X, 1403625502X, 1403625504X, 1403625505X, 1403688701X, 1403688702X, 1403688801X, 1403688802X, 1403688803X, 1403688804X, 1403688805X, 1403688806X, 1403736301X, 1403736302X, 1403736303X, 1403736304X, 1403736305X, 1403736306X, 1403736401X, 1403736404X, 1403791801X, 1403791803X, 1403791805X, 1403791806X, 1403830904X, 1403830905X, 1403830906X, 1403831004X, 1403831005X, 1403831006X, 1403862204X, 1403862205X, 1403862206X, 1403862305X, 1403862306X, 1403886401X, 1403886402X, 1403886403X, 1403886404X, 1403886405X, 1403886406X, 1403886501X, 1403886502X, 1403886503X, 1403886504X, 1403886505X, 1403886506X, 1403986603X, 1403986604X, 1403986605X, 1403986606X, 1403990601X, 1403990602X, 1403990603X, 1403990604X, 1403990701X, 1403990703X, 1403990704X, 1403990705X, 1403990706X, 1404022104X, 1404022105X, 1404022202X, 1404022204X, 1501030401X, 1501030601X, 1501030606X, 1501088001X, 1501088006X, 1501088101X, 1501088105X, 1501088106X, 1501132903X, 1501132905X, 1501132906X, 1501158301X, 1501158302X, 1501158303X, 1501158304X, 1501158401X, 1501158402X, 1501158403X, 1501158404X, 1501158406X, 1501201501X, 1501201506X, 1501274503X, 1501355602X, 1501355603X, 1501355604X, 1501355605X, 1501355606X, 1501355702X, 1501565905X, 1501566001X, 1501566003X, 1501566004X, 150156600SX, 1501566006X, 1501619502X, 1501619503X, 1501619506X, 1501619601X, 1501619604X, 150161960SX, 1501619606X, 1501646804X, 1501646902X, 1501646903X, 1501646906X, 1501672006X, 1501672102X, 1501672105X, 1501672106X, 1501731601X, 1501731602X, 1501731604X, 1501731805X, 1501731806X, 1501790001X, 1501790002X, 1501790003X, 1501790004X, 1501790005X, 1501790006X, 1501790103X, 1501790104X, 1501790105X, 1501790106X, 1501838401X, 1501838402X, 1501838403X, 1501838404X, 1501838405X, 1501838406X, 1501898901X, 1501898902X, 1501898903X, 1501898904X, 1501898905X, 1501899003X, 1501899006X, 1501940101X, 1501940102X, 1501940104X, 1501940105X, 1501940106X, 1501955601X, 1501955602X, 1501955603X, 1501955604X, 1501955605X, 1502051002X, 1502051004X, 1502051006X, 1502134606X, 1502134701X, 1502134703X, 1502134705X, 1502176206X, 1502326701X, 1502326702X, 1502384602X, 1502384603X, 1502384605X, 1502384606X, 1502384705X, 1502384706X, 1502443501X, 1502443503X, 1502443504X, 1502443506X, 1502459501X, 1502459502X, 1502459601X, 1502459603X, 1502459604X, 1502473801X, 1502473802X, 1502473803X, 1502473804X, 1502473805X, 1502473806X, 1502473904X, 1502473905X, 1502499501X, 1502499504X, 1502499505X, 1502499506X, 1502499704X, 1502499706X, 1502522701X, 1502522702X, 1502522801X, 1502522803X, 1502568201X, 1502568202X, 1502568203X, 1502568204X, 1502568301X, 1502568302X, 1502568303X, 1502568304X, 1502568305X, 1502568306X, 1502594901X, 1502594902X, 1502594905X, 1502609201X, 1502609202X, 1502609203X, 1502609301X, 1502609303X, 1502609304X, 1502609305X, 1502609306X, 1502640502X, 1502640504X, 1502640505X, 1502640506X, 1502640602X, 1502640606X, 1502672101X, 1502672106X, 1502672204X, 1502672205X, 1502672206X, 1502695602X, 1502695604X, 1502695704X, 1502695705X, 1502733402X, 1502733403X, 1502733404X, 1502733504X, 1502733506X, 1502771903X, 1502771906X, 1502772002X, 1502772003X, 1502772006X, 1502782102X, 1502782103X, 1502782104X, 1502782105X, 1502782106X, 1502782201X, 1502782202X, 1502782205X, 1502782206X, 1502834802X, 1502834803X, 1502834804X, 1502834805X, 1502834901X, 1502834905X, 1502834906X, 1502857506X, 1502857601X, 1502857602X, 1502857603X, 1502913001X, 1502913002X, 1502913003X, 1502913004X, 1502913005X, 1502913006X, 1502913101X, 1502913103X, 1502913105X, 1502958103X, 1502958104X, 1502958105X, 1502958106X, 1502958301X, 1502958303X, 1502958306X, 1502985702X, 1502985703X, 1502985704X, 1502985706X, 1502985801X, 1502985803X, 1502985804X, 1502985805X, 1503042801X, 1503042802X, 1503090601X, 1503090602X, 1503090605X, 1503143001X, 1503143002X, 1503143004X, 1503143101X, 1503143102X, 1503175402X, 1503175404X, 1503175501X, 1503175506X, 1503218802X, 1503218803X, 1503218804X, 1503218806X, 1503218901X, 1503218905X, 1503261504X, 1503261505X, 1503261604X, 1503261605X, 1503316001X, 1503316002X, 1503316003X, 1503316102X, 1503316106X, 1503372606X, 1503402202X, 1503402302X, 1503402303X, 1503402304X, 1503402305X, 1503402306X, 1503436204X, 1503436205X, 1503436301X, 1503436302X, 1503436303X, 1503436305X, 1503436306X, 1503544602X, 1503545005X, 1503581101X, 1503581102X, 1503581204X, 1503822001X, 1503822002X, 1503822003X, 1503822101X, 1503822105X, 1503822106X, 1504063501X, 1504063503X, 1504063505X, 1504063601X, 1504063602X, 1504063603X, 1504063606X, 1504108101X, 1504108103X, 1504108104X, 1504108105X, 1504108106X, 1504108601X, 1601025103X, 1601093502X, 1601093503X, 1601093504X, 1601093505X, 1601093506X, 1601093704X, 1601135701X, 1601135702X, 1601135703X, 1601135705X, 1601135706X, 1601749301X, 1601749303X, 1601749304X, 1601749305X, 1601749306X, 1601749401X, 1601749404X, 1601749405X, 1601749406X, 1601818403X, 1601818404X, 1601818405X, 1601818406X, 1601818504X, 1601818505X, 1602059202X, 1602059203X, 1602059204X, 1602059206X, 1602059301X, 1602059304X, 1602152403X, 1602152404X, 1602152405X, 1602152406X, 1602152601X, 1602152602X, 1602152605X, 1602152606X, 1602327203X, 1602327204X, 1602327301X, 1602327302X, 1602327303X, 1602327304X, 1602487902X, 1602487905X, 1602647702X, 1602647703X, 1602647706X, 1602647801X, 1602647802X, 1602647803X, 1602750701X, 1602750704X, 1602750705X, 1602750706X, 1602750801X, 1602750802X, 1602750804X, 1602890901X, 1602890902X, 1603141405X, 1603141406X, 1603141504X, 1603141505X, 1603364801X, 1603364802X, 1603364806X, 1603364902X, 1603364903X, 1603364906X, 1603479703X, 1603479704X, 1603479801X, 1603479802X, 1603701601X, 1603762804X, 1603762805X, 1603762806X, 1603831005X, 1603831006X, 1603831101X, 1603831102X, 1603831104X, 1603875403X, 1603875406X, 1603875502X, 1603875504X, 1603875506X, 1604011304X, 1604011403X, 1604011405X, 1604115302X, 1604115304X, 1604115305X, 1604115403X, 1604115404X, 1604115405X, 1604115406X, 1604329801X, 1604329803X, 1604374601X, 1604374602X, 1604374603X, 1604374604X, 1604374605X, 1604374606X, 1701000401X, 1701000402X, 1701000403X, 1701042103X, 1701042105X, 1701042106X, 1701079602X, 1701079603X, 1701079604X; International Serial Numbers: 1503175502X, 1501731606X, 1501898906X, 1501899002X, 1502176303X, 1502695606X, 1602059306X, 1602152401X, 1403484606X, 1403862202X, 1501355706X, 1501565901X, 1503261502X, 1503544603X, 1503544604X, 1403484504X, 1501273405X, 1501201706X, 1502051003X, 1602890805X, 1602890806X, 1503545004X, 1503316006X, 1501088005X, 1501201701X, 1501201703X, 1502326706X, 1501088003X, 1501731603X, 1502672105X, 1502857503X, 1503218902X, 1503261602X, 1601749302X, 1402455303X, 1502594803X, 1502695701X, 1602750703X, 1602750803X, 1602750805X, 1602750806X, 1602890803X, 1403440405X, 1403484505X, 1502384702X, 1502443505X, 1502443805X, 1502459605X, 1502522704X, 1502522805X, 1503042701X, 1501838501X, 1501838504X, 1502176306X, 1502568205X, 1503042803X, 1503090504X, 1503090505X, 1503090603X, 1503143105X, 1503372604X, 1503372701X, 1503545002X, 1603831105X, 1502005104X, 1501672004X, 1502499502X, 1503372705X, 1602890804X, 1403990702X, 1404022205X, 1501646803X, 1502134601X, 1502640605X, 1503042704X, 1503218906X, 1503261501X, 1503261506X, 1503316004X, 1503372703X, 1604011301X, 1501273404X, 1501273406X, 1501672005X, 1502134602X, 1502640603X, 1503090604X, 1503316005X, 1503316101X, 1503436203X, 1501133005X, 1403986805X, 1501355704X, 1402959001X, 1403272703X, 1403440403X, 1403539705X, 1501088004X, 1502985701X, 1404022103X, 1501273403X, 1501274502X, 1501355601X, 1502005106X, 1502499703X, 1502499705X, 1403440401X, 1403688706X, 1501274501X, 1501274504X, 1504063502X, 1403390503X, 1403390802X, 1501672003X, 1403539604X, 1501273402X, 1502913104X, 1502985802X, 1503545001X, 1501619504X, 1501619505X, 1403272704X, 1403986602X, 1403986803X, 1501088103X, 1501201503X, 1501201505X, 1501565902X, 1501565904X, 1501940201X, 1501940203X, 1501940205X, 1501955501X, 1502050905X, 1502176204X, 1502176302X, 1502522806X, 1502594804X, 1502985806X, 1503042706X, 1503090502X, 1503143006X, 1503143104X, 1503175401X, 1503175406X, 1503544606X, 1503822102X, 1503822103X, 1601025001X, 1501132902X, 1403862203X, 1404022101X, 1404022201X, 1404022206X, 1501030603X, 1501030604X, 1501030605X, 1501133004X, 1501201502X, 1501201504X, 1501201702X, 1501201704X, 1501355701X, 1501565906X, 1501619603X, 1501731803X, 1501838503X, 1501838505X, 1501955503X, 1501955505X, 1502005103X, 1502134603X, 1502176304X, 1502384703X, 1502443801X, 1502473901X, 1502609204X, 1502640501X, 1502640503X, 1502672203X, 1603762701X, 1603762702X, 1603762703X, 1603762704X, 1603762705X, 1603762706X, 1603762801X, 1603762802X, 1603831001X, 1501940206X, 1502640604X, 1502672102X, 1503372702X, 1503402206X, 1603141403X, 1603831003X, 1603831004X, 1602059302X, 1502733401X, 1503175403X, 1403253704X, 1603831103X, 1502459503X, 1502459504X, 1502473902X, 1502473903X, 1502499702X, 1502834903X, 1502985705X, 1503175405X, 1503175505X, 1503372602X, 1504063504X, 1504063506X, 1504063604X, 1504063605X, 1601025102X, 1601025104X, 1601025106X, 1601093703X, 1601135601X, 1601135602X, 1601135603X, 1601135605X, 1601749403X, 1601818401X, 1601818402X, 1601818501X, 1601818506X, 1502050903X, 1602152402X, 1501731802X, 1501940202X, 1502050902X, 1403484605X, 1501355703X, 1502672201X, 1503143003X, 1503261601X, 1503316104X, 1501672002X, 1501838502X, 1502771904X, 1502772004X, 1502782203X, 1603875402X, 1602152603X, 1602152604X, 1501273401X, 1501899001X, 1502857606X, 1403736405X, 1501132904X, 1502594805X, 1503042804X, 1503090503X, 1503090606X, 1503544605X, 1602327206X, 1602327305X, 1402887605X, 1402887606X, 1403351005X, 1403688703X, 1501158405X, 1501565903X, 1502005201X, 1502522705X, 1502568206X, 1502672104X, 1503372605X, 1503545003X, 1603701604X, 1603701605X, 1501133001X, 1502733501X, 1502771901X, 1502913106X, 1502958101X, 1603875501X, 1603875503X, 1603875505X, 1502834902X, 1403830901X, 1501646904X, 1501672001X, 1502594904X, 1502857501X, 1503042705X, 1503042805X, 1503143103X, 1503218904X, 1503436201X, 1503436202X, 1604329805X, 1604329806X, 1503402204X, 1502772005X, 1503372601X, 1503402205X, 1503545006X, 1503581103X, 1503581104X, 1503581201X, 1503581203X, 1503822104X, 1502134702X, 1502384604X, 1502384701X, 1502594801X, 1601025002X, 1604115301X, 1601093501X, 1501355705X, 1403791802X, 1403831001X, 1403831002X, 1502005202X, 1502609205X, 1502695603X, 1502695706X, 1502958305X, 1503143005X, 1503218805X, 1603831106X, 1603875401X, 1403390504X, 1403390505X, 1403390801X, 1403736403X, 1502695703X, 1503372704X, 1503581202X, 1503822004X, 1504108102X, 1601025004X, 1602890903X, 1602890905X, 1602890906X, 1603141401X, 1701000404X, 1701042104X, 1701079601X, 1701079606X, 1701134801X, 1701134803X, 1701134806X, 1701211101X, 1701211102X, 1701211103X, 1701211104X, 1701211202X, 1701211206X, 1403791706X, 1403862301X, 1403862303X, 1403862304X, 1403211703X, 1403332301X, 1403351001X, 1403351003X, 1403351004X, 1403484506X, 1403625503X, 1403791701X, 1403791804X, 1403986802X, 1501274506X, 1501619602X, 1501646802X, 1501646805X, 1501646806X, 1501646905X, 1501672103X, 1501790101X, 1501899005X, 1501955502X, 1502005101X, 1502326406X, 1502326703X, 1502443803X, 1502459506X, 1502459606X, 1502499701X, 1502594802X, 1502733503X, 1502958304X, 1503143106X, 1503175503X, 1503261503X, 1503372603X, 1503436304X, 1503544601X, 1601749402X, 1602327202X, 1602327306X, 1602487903X, 1602488001X, 1602488002X, 1602488003X, 1602488004X, 160248800SX, 1602488006X, 1602647701X, 1602647704X, 1602890801X, 1602890802X, 1602890904X, 1603141402X, 1603141404X, 1603141503X, 1603364804X, 1603364901X, 1603364905X, 1603479701X, 1603479702X, 1603479705X, 1603479706X, 1603479803X, 1603479804X, 1603479805X, 1603479806X, 1603701602X, 1603701603X, 1603701606X, 1603875404X, 1603875405X, 1604011302X, 1604011303X, 1701277501X, 1701277502X, 1403539605X, 1403539701X, 1403688704X, 1403688705X, 1403831003X, 1403986801X, 1403484503X, 1501646801X, 1502834806X, 1603141501X, 1501030402X, 1501030405X, 1503042702X, 1403253701X, 1403332306X, 1403539602X, 1501940103X, 1502005206X, 1502134605X, 1502176305X, 1502326704X, 1502050904X, 1502134704X, 1502695601X, 1503175504X, 1603831002X, 1501940204X, 1502134706X, 1502176205X, 1502326404X, 1502443802X, 1502522703X, 1502522706X, 1502609206X, 1402705202X, 1501030403X, 1S01030404X, 1501030406X, 1501088102X, 1501955504X, 1501955606X, 1502005105X, 1502051005X, 1502522804X, 1502594906X, 1503090506X, 1503218801X, 1503261603X, 1601025003X, 1601025005X

Distribution

Worldwide Distribution - US (nationwide and Puerto Rico): Units were also distributed globally to the following countries: Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bosnia, Herzegovina, Bulgaria, Canada, Chile, Columbia, Costa Rica, Croatia, Curacao, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Ecuador, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hungary, India, Italy, Jamaica, Japan, Jordan, Kazakhstan, South Korea, Kuwait, Lebanon, Lithuania, Luxembourg, Malaysia, Mexico, Mongolia, Morocco, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, Ukraine, United Kingdom, and Uzbekistan

Quantity

1,074 units