Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Catheters. Please refer to the applicable catheter Instructions for Use for directions on the use of these accessories.
Enforcement
- Recall Number
- Z-3062-2024
- Event ID
- 95039
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Bard Peripheral Vascular Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 18, 2024
- Initiation Date
- July 10, 2024
- Classification Date
- September 6, 2024
- Address
- 1625 W 3rd St Bldg 1, 850 W. Rio Salado Prkwy., Tempe, AZ, 85281-2438, United States
Description
Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Catheters. Please refer to the applicable catheter Instructions for Use for directions on the use of these accessories.
Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur if the user attempts to insert the proximal end into the patient.
Catalog Number: 5583705 Lot Number / UDI codes: REGZ2564 (01)00801741085451(17)271130(10)REGZ2564 REFV0870 (01)00801741085451(17)260731(10)REFV0870 REFU4050 (01)00801741085451(17)260630(10)REFU4050 REFT3309 (01)00801741085451(17)260531(10)REFT3309 REEX1002 (01)00801741085451(17)250930(10)REEX1002 REEW2740 (01)00801741085451(17)250831(10)REEW2740 REEV0897 (01)00801741085451(17)250731(10)REEV0897 REDY2576 (01)00801741085451(17)241031(10)REDY2576 REEU1584 (01)00801741085451(17)250630(10)REEU1584 REER0990 (01)00801741085451(17)250331(10)REER0990 REEQ0825 (01)00801741085451(17)250228(10)REEQ0825
Worldwide - U.S. Nationwide distribution in the states of OH, TX, and WV. The countries of Belgium, Canada, France, Italy, and United Kingdom.
625 units