Medtronic Curved Suction 90 9733451 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it
Enforcement
- Recall Number
- Z-3053-2017
- Event ID
- 77739
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medtronic Navigation, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 30, 2017
- Initiation Date
- June 29, 2017
- Classification Date
- August 21, 2017
- Termination Date
- August 3, 2018
- Address
- 826 Coal Creek Cir, Louisville, CO, 80027-9710, United States
Description
Medtronic Curved Suction 90 9733451 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
Lot numbers 17030728, 17022027, 17021426, 17013125, 17012724, 17011303, 16122767, 16122066, 16112965, and 16112164
Worldwide Distribution - US (nationwide) and Internationally to Canada.
75 instruments