FDA Enforcement
Class II
Terminated
First Check 7 Panel Drug Cup: (AMP/COC/MDMA/MET/OPI/PCP/THC); AMT Item Number 990100B
Recall: Z-3050-2020
·
Reported October 7, 2020
Enforcement
- Recall Number
- Z-3050-2020
- Event ID
- 86335
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ameditech Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 7, 2020
- Initiation Date
- August 18, 2020
- Classification Date
- September 30, 2020
- Termination Date
- September 30, 2024
- Address
- 9940 Mesa Rim Rd, N/A, San Diego, CA, 92121-2910, United States
Description
First Check 7 Panel Drug Cup: (AMP/COC/MDMA/MET/OPI/PCP/THC); AMT Item Number 990100B
Reason
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Code Info
ALL Lots Within Expiry
Distribution
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Quantity
99,900 units