FDA Enforcement Class II Terminated

ORA System with VerifEye, Catalog Number 8065998300 For use during intraocular lens surgery

Recall: Z-3050-2017 · Reported August 30, 2017

Enforcement

Recall Number
Z-3050-2017
Event ID
77823
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Alcon Research, Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
August 30, 2017
Initiation Date
June 30, 2017
Classification Date
August 21, 2017
Termination Date
August 27, 2019
Address
6201 South Fwy, Fort Worth, TX, 76134-2099, United States

Description

ORA System with VerifEye, Catalog Number 8065998300 For use during intraocular lens surgery

Reason

Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.

Code Info

Serial Numbers: C0002,C2006,C2020,C2026,C2054,C3009R,C3010R,C3024,C3036,C3066,C3067,C3085R,C3090,C3093,C3108,C3136,C3153,C4001,C4003,C4005,C4006,C4007,C4009,C4010,C4011,C4012,C4013,C4014,C4015,C4016,C4017,C4019,C4020,C4021,C4022,C4023,C4024,C4025,C4026,C4027,C4028,C4029,C4030,C4031,C4035,C4036,C4037,C4038,C4039,C4040,C4043,C4045,C4046R,C4047R,C4049R,C4050R,C4052R,C4053R,C4054R,C4055R,C4056R,C4057R,C4058R,C4059R,C4060R,C4061R,C4062R,C4065R,C4066R,C4067R,C4068R,C4069R,C4070R,C4071R,C4072R,C4073R,C4074R,C4075R,C4076R,C4077R,C4078R,C4079R,C4080R,C4083R,C4085R,C4086R,C4087RC4088R,C4089R,C4090R,C4091R,C4092R,C4093R,C4094R,C4095R,C4096R,C4097R,C4098R,C4099R,C4100R,C4101R,C4102R,C4103R,C4104R,C4105R,C4106R,C4107R,C4108R,C4109R,C4110R,C4111R,C4112R,C4113R,C4115R,C4117R,C4118R,C4119R,C4120R,C4121R,C4122R,C4123R,C4124R,C4126R,C4127R,C4128R,C4129R,C4130R,C4132R,C4133R,C4134R,C4135R,C4136R,C4138R,C4140,C4142,C4143,C4144,C4145,C4146,C4147,C4148,C4149,C4150,C4151,C4152R,C4153,C4154,C4155,C4156,C4157R,C4158,C4159,C4160R,C4161R,C4162R,C4163R,C4164R,C4165R,C4166R,C4167R,C4168R,C4169R,C4170R,C4173R,C4174R,C4175,C4177R,C4178R,C4179R,C4180R,C4181R,C4182R,C4183R,C4184R,C4185R,C4186R,C4187R,C4188R,C4189R,C4190R,C4191R,C4192R,C4193R,C4195R,C4196R,C4197R,C4198,C4199R,C4200R,C4201,C4202R,C4203R,C4204R,C4205R,C4206R,C4207R,C4208R,C4209R,C4210R,C4211R,C4212R,C4213R,C4214R,C4215R,C4216R,C4217R,C4218R,C4219R,C4220R,C4221R,C4222R,C4223R,C4224R,C4225R,C4226R,C4227R,C4228R,C4229R,C4230R,C4231R,C4232R,C4233R,C4234R,C4235R,C4238R,C4239R,C4240R,C4241R,C4242R,C4243R,C4244R,C4245R,C4246R,C4247R,C4248R,C4249R,C4251R,C4252R,C4253R,C4254R,C4255R,C4256R,C4257R,C4258R,C4259R,C4260R,C4261R,C4262R,C4264R,C4266R,C4267R,C4268R,C4269R,C4271R,C4272R,C4273R,C4274R,C4275R,C4276R,C4278R,C4279R,C4280R,C4283R,C4284R,C4285R,C4286R,C4287R,C4288R,C4289R,C4290R,C4291R,C4292R,C4293R,C4294R,C4295R,C4296R,C4297R,C4298R,C4299R,C4300R,C4301R,C4302R,C4303R,C4304R,C4305R,C4306R,C4307R,C4308R,C4309R,C4310R,C4311R,C4312R,C4315R,C4316R,C4317R,C4318R,C4319R,C4320R,C4321R,C4323R,C4324R,C4326R,C4327R,C4328R,C4331R,C4332R,C4334R,C4337R,C4339R,C4340R,C4341R,C4342R,C4343R,C4344R,C4345R,C4346R,C4347R,C4348R,C4349R,C4352R,C4354R,C4355R,C4356R,C4357R,C4358R,C4359R,C4360R,C4361R,C4363R,C4364R,C4365R,C4366R,C4367R,C4368R,C4371R,C4372R,C4373R,C4374R,C4376R,C4377R,C4378R,C4379R,C4380R,C4381R,C4382R,C4383R,C4384R,C4385R,C4387R,C4388R,C4389R,C4394,C4396,C4397,C4399,C4406R,C4407R,C4408R,C4415,C4418,C4420,C4421,C4422,C4423,C4424,C4425,C4426

Distribution

Worldwide distribution. US Nationwide and Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, France, India, Japan, Mexico, Netherlands, Panama, Portugal, Romania, Spain, Thailand, UAE, and United Kingdom

Quantity

367 units