ORA System with VerifEye+ Cart, Catalog Number 8065998307 For use during intraocular lens surgery
Enforcement
- Recall Number
- Z-3049-2017
- Event ID
- 77823
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Alcon Research, Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Telephone
- Report Date
- August 30, 2017
- Initiation Date
- June 30, 2017
- Classification Date
- August 21, 2017
- Termination Date
- August 27, 2019
- Address
- 6201 South Fwy, Fort Worth, TX, 76134-2099, United States
Description
ORA System with VerifEye+ Cart, Catalog Number 8065998307 For use during intraocular lens surgery
Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.
Serial numbers: C5013,C5014,C5015,C5016,C5017,C5018,C5019,C5020,C5021,C5023,C5024,C5025,C5026,C5027,C5028,C5029,C5030,C5031,C5032,C5033,C5034,C5035,C5036,C5038,C5039,C5040,C5041,C5042,C5043,C5044, C5045,C5046,C5047,C5048,C5049,C5050,C5051,C5053,C5054,C5055,C5059,C5060,C5061,C5062,C5064, C5065,C5066,C5069,C5071,C5075,C5076,C5077,C5078,C5079,C5080,C5081,C5082,C5083,C5090,C5093, C5094,C5097,C5098,C5101,C5102,C5105,C5106,C5107,C5108,C5109,C5110,C5111,C5112,C5113,C5116, C5117,C5119,C5120,C5121,C5122,C5125,C5126,C5127,C5128,C5129,C5130,C5131,C5132,C5133,C5134, C5136,C5137,C5138,C5139,C5140,C5141,C5142,C5143,C5144,C5146,C5147,C5148,C5149,C5150,C5151, C5152,C5153,C5154,C5155,C5156,C5157,C5158,C5159,C5160,C5161,C5163,C5164,C5165,C5166,C5167, C5168,C5169,C5170,C5171,C5173,C5174,C5175,C5176,C5177,C5178,C5179,C5180,C5181,C5182,C5183, C5184,C5185,C5186,C5187,C5188,C5189,C5190,C5191,C5192,C5193,C5194,C5195,C5196,C5197,C5198, C5199,C5200,C5201,C5202,C5203,C5204,C5205,C5207,C5214,C5215,C5216,C5217,C5218,C5219,C5220, C5221,C5222,C5223,C5224,C5225,C5226,C5227,C5228,C5229,C5230,C5231,C5232,C5233,C5238,C5239, C5240,C5241,C5242,C5243,C5244,C5245,C5246,C5248,C5249,C5250,C5251,C5252,C5253,C5255,C5256, C5257,C5258,C5259,C5260,C5261,C5262,C5263,C5264,C5265,C5266,C5267,C5268,C5269,C5270,C5271, C5272,C5273,C5274,C5276,C5277,C5278,C5279,C5280,C5281,C5282,C5283,C5284,C5285,C5286,C5287, C5288,C5289,C5290,C5291,C5292,C5293,C5294,C5295,C5296,C5297,C5298,C5299,C5300,C5301,C5302, C5303,C5304,C5305,C5306,C5307,C5308,C5309,C5310,C5311,C5312,C5313,C5314,C5315,C5316,C5317, C5318,C5319,C5320,C5321,C5322,C5323,C5326,C5328,C5329,C5330,C5334,C5335,C5336,C5339,C5340, C5341,C5343,C5348,C5349,C5350,C5352,C5353,C5354,C5355,C5356,C5357,C5358,C5359,C5360,C5361, C5362,C5363,C5364,C5365,C5366,C5368,C5369,C5370,C5371,C5372,C5373,C5374,C5375,C5376,C5377, C5378,C5379,C5380,C5381,C5382,C5383,C5384,C5385,C5386,C5387,C5388,C5389,C5390,C5391,C5392, C5393,C5394,C5395,C5396,C5397,C5398,C5399,C5400,C5401,C5402,C5403,C5404,C5405,C5406,C5407, C5408,C5409,C5410,C5411,C5412,C5413,C5414,C5415,C5416,C5417,C5418,C5434,C5435,C5436,C5437, C5438,C5439,C5440,C5441,C5442,C5443,C5444,C5446,C5448,C5449,C5451,C5452,C5453,C5454,C5456, C5458,C5462,C5463,C5464,C5465,C5466,C5467,C5469,C5471,C5473,C5475,C5476,C5477,C5478,C5479, C5484,C5485,C5486,C5487,C5488,C5489,C5490,C5493,C5494,C5495,C5496,C5497,C5499,C5500,C5501, C5502,C5503,C5507,C5508,C5509,C5510,C5511,C5512,C5515,C5516,C5517,C5518,C5519,C5520,C5521, C5522,C5524,C5525,C5526,C5527,C5528,C5529,C5530,C5532,C5533,C5534,C5535,C5536,C5537,C5538, C5539,C5544,C5545,C5546,C5547,C5549,C5550,C5553,C5555
Worldwide distribution. US Nationwide and Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, France, India, Japan, Mexico, Netherlands, Panama, Portugal, Romania, Spain, Thailand, UAE, and United Kingdom
429 units