Boston Scientific Wallstent Endoprosthesis Tracheobronchial Self-Expanding Stent, 24MM X 70MM | 75CM L, resterilized.
Recall: Z-3037-2018
·
Reported September 19, 2018
Enforcement
- Recall Number
- Z-3037-2018
- Event ID
- 80317
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- SPS Sterilization, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 19, 2018
- Initiation Date
- March 7, 2018
- Classification Date
- September 13, 2018
- Termination Date
- November 27, 2018
- Address
- 1110 N Gerald St, N/A, Nixa, MO, 65714-7445, United States
Description
Boston Scientific Wallstent Endoprosthesis Tracheobronchial Self-Expanding Stent, 24MM X 70MM | 75CM L, resterilized.
Reason
The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
Code Info
Code Number 40512-R; Lot Number 17984215; Serial Number 21408032017-14, 21408032017-15
Distribution
US distribution to AZ, CA, FL, OH, PA, TX, VA, WV
Quantity
2 units total