Medline Convenience kits labeled as: 1) CENTRAL LINE CDS, Pack Number CDS930092U ; 2) SMALL WOUND TRAY, Pack Number DYNDA1822; 3) RADIOLOGY UTILITY TRAY, Pack Number DYNDA2865; 4) BLOOD CULTURE SET , Pack Number DYNDH1495A ; 5) LACERATION TRAY , Pack Number DYNDL1050; 6) LACERATION TRAY , Pack Number DYNDL1062B ; 7) LACERATION TRAY , Pack Number DYNDL1073; 8) LACERATION TRAY , Pack Number DYNDL1076; 9) LACERATION TRAY , Pack Number DYNDL1090A ; 10) ER LACERATION TRAY, Pack Number DYNDL1096; 11) LACERATION TRAY , Pack Number DYNDL1151C ; 12) LACERATION TRAY , Pack Number DYNDL1152A ; 13) LACERATION TRAY , Pack Number DYNDL1152AH; 14) LACERATION TRAY , Pack Number DYNDL1159B ; 15) LACERATION TRAY , Pack Number DYNDL1163; 16) LACERATION TRAY , Pack Number DYNDL1163H ; 17) LACERATION TRAY , Pack Number DYNDL1171; 18) CUSTOM LACERATION TRAY, Pack Number DYNDL1177C ; 19) E. R. LACERATION TRAY , Pack Number DYNDL1187A ; 20) LACERATION TRAY , Pack Number DYNDL1208B ; 21) LACERATION TRAY, Pack Number DYNDL1223A; 22) LACERATION TRAY , Pack Number DYNDL1250C ; 23) LACERATION TRAY , Pack Number DYNDL1257A ; 24) LACERATION TRAY , Pack Number DYNDL1263A ; 25) TRAY,LACERATION , Pack Number DYNDL1271B ; 26) LACERATION TRAY , Pack Number DYNDL1284; 27) LACERATION TRAY , Pack Number DYNDL1285; 28) LACERATION TRAY , Pack Number DYNDL1366A ; 29) LACERATION TRAY , Pack Number DYNDL1368; 30) TRAY,FACIAL PLASTIC LACERATION, Pack Number DYNDL1377A ; 31) TRAY,FACIAL PLASTIC LACERATION, Pack Number DYNDL1377AH; 32) LACERATION TRAY , Pack Number DYNDL1399; 33) LACERATION TRAY , Pack Number DYNDL1403; 34) LACERATION TRAY , Pack Number DYNDL1406A ; 35) TRAY,LACERATION , Pack Number DYNDL1455; 36) LACERATION TRAY , Pack Number DYNDL1457; 37) TRAY LACERATION , Pack Number DYNDL1465; 38) LACERATION TRAY , Pack Number DYNDL1496; 39) LACERATION TRAY , Pack Number DYNDL1496H ; 40) TRAY,LACERATION , Pack Number DYNDL1510; 41) TRAY,LACERATION, Pack Number DYNDL1542; 42) TRAY,LACERATION , Pack Number DYNDL1542; 43) TRAY ER LACERATION, Pack Number DYNDL1569A ; 44) TRAY SM ER LACERATION , Pack Number DYNDL1573; 45) LACERATION TRAY , Pack Number DYNDL1575A ; 46) LACERATION TRAY , Pack Number DYNDL1628A ; 47) LACERATION TRAY , Pack Number DYNDL1634; 48) LACERATION PACK , Pack Number DYNDL1637A ; 49) LACERATION TRAY , Pack Number DYNDL1678; 50) LACERATION TRAY , Pack Number DYNDL1679B ; 51) LACERATION TRAY , Pack Number DYNDL1690; 52) ER LACERATION TRAY, Pack Number DYNDL1693; 53) LACERATION TRAY , Pack Number DYNDL1705; 54) LACERATION TRAY, Pack Number DYNDL1705H ; 55) SUTURE LACERATION KIT , Pack Number DYNDL1707A ; 56) LACERATION TRAY , Pack Number DYNDL1714; 57) LACERATION KIT, Pack Number DYNDL1717; 58) NMG LACERATION TRAY , Pack Number DYNDL1722; 59) LACERATION TRAY , Pack Number DYNDL1728; 60) LACERATION TRAY , Pack Number DYNDL1729; 61) LACERATION TRAY , Pack Number DYNDL1739; 62) LACERATION TRAY , Pack Number DYNDL1740; 63) LACERATION TRAY , Pack Number DYNDL1757; 64) LACERATION TRAY , Pack Number DYNDL1758; 65) LACERATION TRAY, Pack Number DYNDL1758H ; 66) LACERATION TRAY HIGH GRADE, Pack Number DYNDL1762; 67) LACERATION TRAY , Pack Number DYNDL1768; 68) LACERATION TRAY , Pack Number DYNDL1770B ; 69) LACERATION TRAY , Pack Number DYNDL1772; 70) LACERATION TRAY , Pack Number DYNDL1774A ; 71) LACERATION TRAY W/SAFETY NDLS , Pack Number DYNDL1783; 72) LACERATION TRAY , Pack Number DYNDL1790; 73) LACERATION TRAY , Pack Number DYNDL1838; 74) LACERATION TRAY , Pack Number DYNDL1870; 75) LACERATION TRAY , Pack Number DYNDL1870H ; 76) LACERATION TRAY , Pack Number DYNDL1872; 77) LACERATION TRAY W/ UNDERPAD , Pack Number DYNDL1878; 78) LACERATION TRAY , Pack Number DYNDL1887; 79) LACERATION TRAY , Pack Number DYNDL1888; 80) LACERATION TRAY , Pack Number DYNDL1892; 81) LACERATION TRAY , Pack Number DYNDL1920; 82) LACERATION TRAY W/ ADSON FCP, Pack Number DYNDL1923A ; 83) LACERATION TRAY
Enforcement
- Recall Number
- Z-3023-2024
- Event ID
- 94583
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- MEDLINE INDUSTRIES, LP - Northfield
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- September 18, 2024
- Initiation Date
- April 8, 2024
- Classification Date
- September 6, 2024
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline Convenience kits labeled as: 1) CENTRAL LINE CDS, Pack Number CDS930092U ; 2) SMALL WOUND TRAY, Pack Number DYNDA1822; 3) RADIOLOGY UTILITY TRAY, Pack Number DYNDA2865; 4) BLOOD CULTURE SET , Pack Number DYNDH1495A ; 5) LACERATION TRAY , Pack Number DYNDL1050; 6) LACERATION TRAY , Pack Number DYNDL1062B ; 7) LACERATION TRAY , Pack Number DYNDL1073; 8) LACERATION TRAY , Pack Number DYNDL1076; 9) LACERATION TRAY , Pack Number DYNDL1090A ; 10) ER LACERATION TRAY, Pack Number DYNDL1096; 11) LACERATION TRAY , Pack Number DYNDL1151C ; 12) LACERATION TRAY , Pack Number DYNDL1152A ; 13) LACERATION TRAY , Pack Number DYNDL1152AH; 14) LACERATION TRAY , Pack Number DYNDL1159B ; 15) LACERATION TRAY , Pack Number DYNDL1163; 16) LACERATION TRAY , Pack Number DYNDL1163H ; 17) LACERATION TRAY , Pack Number DYNDL1171; 18) CUSTOM LACERATION TRAY, Pack Number DYNDL1177C ; 19) E. R. LACERATION TRAY , Pack Number DYNDL1187A ; 20) LACERATION TRAY , Pack Number DYNDL1208B ; 21) LACERATION TRAY, Pack Number DYNDL1223A; 22) LACERATION TRAY , Pack Number DYNDL1250C ; 23) LACERATION TRAY , Pack Number DYNDL1257A ; 24) LACERATION TRAY , Pack Number DYNDL1263A ; 25) TRAY,LACERATION , Pack Number DYNDL1271B ; 26) LACERATION TRAY , Pack Number DYNDL1284; 27) LACERATION TRAY , Pack Number DYNDL1285; 28) LACERATION TRAY , Pack Number DYNDL1366A ; 29) LACERATION TRAY , Pack Number DYNDL1368; 30) TRAY,FACIAL PLASTIC LACERATION, Pack Number DYNDL1377A ; 31) TRAY,FACIAL PLASTIC LACERATION, Pack Number DYNDL1377AH; 32) LACERATION TRAY , Pack Number DYNDL1399; 33) LACERATION TRAY , Pack Number DYNDL1403; 34) LACERATION TRAY , Pack Number DYNDL1406A ; 35) TRAY,LACERATION , Pack Number DYNDL1455; 36) LACERATION TRAY , Pack Number DYNDL1457; 37) TRAY LACERATION , Pack Number DYNDL1465; 38) LACERATION TRAY , Pack Number DYNDL1496; 39) LACERATION TRAY , Pack Number DYNDL1496H ; 40) TRAY,LACERATION , Pack Number DYNDL1510; 41) TRAY,LACERATION, Pack Number DYNDL1542; 42) TRAY,LACERATION , Pack Number DYNDL1542; 43) TRAY ER LACERATION, Pack Number DYNDL1569A ; 44) TRAY SM ER LACERATION , Pack Number DYNDL1573; 45) LACERATION TRAY , Pack Number DYNDL1575A ; 46) LACERATION TRAY , Pack Number DYNDL1628A ; 47) LACERATION TRAY , Pack Number DYNDL1634; 48) LACERATION PACK , Pack Number DYNDL1637A ; 49) LACERATION TRAY , Pack Number DYNDL1678; 50) LACERATION TRAY , Pack Number DYNDL1679B ; 51) LACERATION TRAY , Pack Number DYNDL1690; 52) ER LACERATION TRAY, Pack Number DYNDL1693; 53) LACERATION TRAY , Pack Number DYNDL1705; 54) LACERATION TRAY, Pack Number DYNDL1705H ; 55) SUTURE LACERATION KIT , Pack Number DYNDL1707A ; 56) LACERATION TRAY , Pack Number DYNDL1714; 57) LACERATION KIT, Pack Number DYNDL1717; 58) NMG LACERATION TRAY , Pack Number DYNDL1722; 59) LACERATION TRAY , Pack Number DYNDL1728; 60) LACERATION TRAY , Pack Number DYNDL1729; 61) LACERATION TRAY , Pack Number DYNDL1739; 62) LACERATION TRAY , Pack Number DYNDL1740; 63) LACERATION TRAY , Pack Number DYNDL1757; 64) LACERATION TRAY , Pack Number DYNDL1758; 65) LACERATION TRAY, Pack Number DYNDL1758H ; 66) LACERATION TRAY HIGH GRADE, Pack Number DYNDL1762; 67) LACERATION TRAY , Pack Number DYNDL1768; 68) LACERATION TRAY , Pack Number DYNDL1770B ; 69) LACERATION TRAY , Pack Number DYNDL1772; 70) LACERATION TRAY , Pack Number DYNDL1774A ; 71) LACERATION TRAY W/SAFETY NDLS , Pack Number DYNDL1783; 72) LACERATION TRAY , Pack Number DYNDL1790; 73) LACERATION TRAY , Pack Number DYNDL1838; 74) LACERATION TRAY , Pack Number DYNDL1870; 75) LACERATION TRAY , Pack Number DYNDL1870H ; 76) LACERATION TRAY , Pack Number DYNDL1872; 77) LACERATION TRAY W/ UNDERPAD , Pack Number DYNDL1878; 78) LACERATION TRAY , Pack Number DYNDL1887; 79) LACERATION TRAY , Pack Number DYNDL1888; 80) LACERATION TRAY , Pack Number DYNDL1892; 81) LACERATION TRAY , Pack Number DYNDL1920; 82) LACERATION TRAY W/ ADSON FCP, Pack Number DYNDL1923A ; 83) LACERATION TRAY
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
CDS930092U , Lot Number 19ABG536 ; CDS930092U , Lot Number 19BBK958 ; CDS930092U , Lot Number 19CBW838 ; CDS930092U , Lot Number 19EBW656 ; CDS930092U , Lot Number 19GBQ822 ; CDS930092U , Lot Number 19IBD005 ; CDS930092U , Lot Number 19VBC035 ; CDS930092U , Lot Number 20ABF408 ; CDS930092U , Lot Number 20BBN475 ; CDS930092U , Lot Number 20CBW941 ; CDS930092U , Lot Number 20EBD309 ; CDS930092U , Lot Number 20FBC457 ; CDS930092U , Lot Number 20GBQ330 ; CDS930092U , Lot Number 20HBQ731 ; CDS930092U , Lot Number 20IBM524 ; CDS930092U , Lot Number 20LBC008 ; DYNDA1822, Lot Number 21JBI823 ; DYNDA2865, Lot Number 22BBT129; DYNDH1495A , Lot Number 21JBQ007 ; DYNDL1050, Lot Number 21JBR511 ; DYNDL1062B , Lot Number 22JMC782 ; DYNDL1062B , Lot Number 23CMD085 ; DYNDL1062B , Lot Number 23FMB278 ; DYNDL1062B , Lot Number 23HMD252 ; DYNDL1062B , Lot Number 23IMB553 ; DYNDL1062B , Lot Number 23KMG005 ; DYNDL1073, Lot Number 21JBJ599 ; DYNDL1076, Lot Number 21JBT719 ; DYNDL1090A , Lot Number 21JBF345 ; DYNDL1096, Lot Number 19CBD823 ; DYNDL1096, Lot Number 22HBZ149 ; DYNDL1096, Lot Number 22IBB980 ; DYNDL1096, Lot Number 20DBU111 ; DYNDL1096, Lot Number 20FBP250 ; DYNDL1096, Lot Number 20JBD367 ; DYNDL1096, Lot Number 20WBC515 ; DYNDL1096, Lot Number 21DBA107 ; DYNDL1151C , Lot Number 21JBF370 ; DYNDL1152A , Lot Number 21JBP002 ; DYNDL1152AH, Lot Number 21JBP002 ; DYNDL1159B , Lot Number 21IBL692 ; DYNDL1159B , Lot Number 21KBS039 ; DYNDL1159B , Lot Number 21KBY562 ; DYNDL1159B , Lot Number 23ABJ966 ; DYNDL1159B , Lot Number 20FBI861 ; DYNDL1159B , Lot Number 20FBU760 ; DYNDL1159B , Lot Number 20HBL711 ; DYNDL1159B , Lot Number 20JBN968 ; DYNDL1159B , Lot Number 21ABH267 ; DYNDL1159B , Lot Number 21ABP087 ; DYNDL1159B , Lot Number 21DBD507 ; DYNDL1159B , Lot Number 21DBX946 ; DYNDL1163, Lot Number 21JBL241 ; DYNDL1163H , Lot Number 21JBL241 ; DYNDL1171, Lot Number 21JBG957 ; DYNDL1171, Lot Number 21JBR534 ; DYNDL1177C , Lot Number 23KBF572 ; DYNDL1187A , Lot Number 21HBG848 ; DYNDL1187A , Lot Number 21IBU306 ; DYNDL1187A , Lot Number 21JBJ576 ; DYNDL1187A , Lot Number 21KBL944 ; DYNDL1187A , Lot Number 20CBN456 ; DYNDL1187A , Lot Number 20EBM129 ; DYNDL1187A , Lot Number 20FBM667 ; DYNDL1187A , Lot Number 20GBD239 ; DYNDL1187A , Lot Number 20HBE309 ; DYNDL1187A , Lot Number 20HBS106 ; DYNDL1187A , Lot Number 20IBH861 ; DYNDL1187A , Lot Number 20IBT274 ; DYNDL1187A , Lot Number 20XBA306 ; DYNDL1187A , Lot Number 21CBH169 ; DYNDL1208B , Lot Number 21JBJ609 ; DYNDL1223A, Lot Number 24BMJ480; DYNDL1223A, Lot Number 24BMB582; DYNDL1223A, Lot Number 23JMH461; DYNDL1223A, Lot Number 23JMG194; DYNDL1223A, Lot Number 23JMC681; DYNDL1223A, Lot Number 23HMD974; DYNDL1223A, Lot Number 23GMG399; DYNDL1223A, Lot Number 23GME992; DYNDL1223A, Lot Number 23EMH614; DYNDL1223A, Lot Number 23EMA620; DYNDL1223A, Lot Number 23BMG057; DYNDL1223A, Lot Number 23BME927; DYNDL1250C , Lot Number 21GBG795 ; DYNDL1250C , Lot Number 21HBT210 ; DYNDL1250C , Lot Number 21KBV270 ; DYNDL1250C , Lot Number 22BBG382 ; DYNDL1250C , Lot Number 20DBE601 ; DYNDL1250C , Lot Number 20DBE602 ; DYNDL1250C , Lot Number 20HBP313 ; DYNDL1250C , Lot Number 20IBB936 ; DYNDL1250C , Lot Number 20IBB994 ; DYNDL1250C , Lot Number 20KBB217 ; DYNDL1250C , Lot Number 21EBQ188 ; DYNDL1250C , Lot Number 22FMA931 ; DYNDL1250C , Lot Number 22FMA932 ; DYNDL1250C , Lot Number 22FMA933 ; DYNDL1250C , Lot Number 22LMD484 ; DYNDL1250C , Lot Number 23AMB228 ; DYNDL1250C , Lot Number 23AMG102 ; DYNDL1250C , Lot Number 23BMG142 ; DYNDL1257A , Lot Number 23JBH591 ; DYNDL1263A , Lot Number 21JBF328 ; DYNDL1271B , Lot Number 21JBL279 ; DYNDL1284, Lot Number 21JBJ564 ; DYNDL1285, Lot Number 21JBI856 ; DYNDL1366A , Lot Number 21JBG953 ; DYNDL1368, Lot Number 21JBR435 ; DYNDL1377A , Lot Number 21JBL265 ; DYNDL1377AH, Lot Number 21JBL265 ; DYNDL1399, Lot Number 21JBG839 ; DYNDL1403, Lot Number 23JBH593 ; DYNDL1403, Lot Number 23KBT805 ; DYNDL1406A , Lot Number 21JBI228 ; DYNDL1455, Lot Number 21JBI239 ; DYNDL1457, Lot Number 21JBN957 ; DYNDL1465, Lot Number 21JBJ578 ; DYNDL1496, Lot Number 21JBQ625 ; DYNDL1496H , Lot Number 21JBQ625 ; DYNDL1510, Lot Number 23LBJ828 ; DYNDL1542, Lot Number 22DBJ898; DYNDL1542, Lot Number 22GBU218 ; DYNDL1542, Lot Number 22JBE733 ; DYNDL1569A , Lot Number 21GBR162 ; DYNDL1569A , Lot Number 21JBF670 ; DYNDL1569A , Lot Number 21KBC385 ; DYNDL1569A , Lot Number 22CBB437 ; DYNDL1569A , Lot Number 20DBU152 ; DYNDL1569A , Lot Number 20FBT009 ; DYNDL1569A , Lot Number 20IBS135 ; DYNDL1569A , Lot Number 20IBT238 ; DYNDL1569A , Lot Number 20KBB441 ; DYNDL1569A , Lot Number 21ABW726 ; DYNDL1569A , Lot Number 21DBG365 ; DYNDL1569A , Lot Number 22EMC376 ; DYNDL1569A , Lot Number 22HMH686 ; DYNDL1569A , Lot Number 23AMD201 ; DYNDL1569A , Lot Number 23DMA529 ; DYNDL1569A , Lot Number 23DMH503 ; DYNDL1569A , Lot Number 23EMJ362 ; DYNDL1569A , Lot Number 23HMC211 ; DYNDL1569A , Lot Number 23IMB882 ; DYNDL1569A , Lot Number 23JMI383 ; DYNDL1569A , Lot Number 23LMC888 ; DYNDL1573, Lot Number 21JBI822 ; DYNDL1575A , Lot Number 21JBR944 ; DYNDL1628A , Lot Number 21JBI865 ; DYNDL1634, Lot Number 21JBI874 ; DYNDL1637A , Lot Number 22FMB776 ; DYNDL1637A , Lot Number 22HMA660 ; DYNDL1637A , Lot Number 22IMI325 ; DYNDL1678, Lot Number 21JBJ567 ; DYNDL1679B , Lot Number 21JBQ019 ; DYNDL1690, Lot Number 21JBT704 ; DYNDL1693, Lot Number 21JBT988 ; DYNDL1705, Lot Number 21JBJ536 ; DYNDL1705H , Lot Number 21JBJ536 ; DYNDL1707A , Lot Number 21JBL276 ; DYNDL1714, Lot Number 23JBH205 ; DYNDL1717, Lot Number 21JBG848 ; DYNDL1722, Lot Number 19BBC384 ; DYNDL1722, Lot Number 19HBO045 ; DYNDL1722, Lot Number 19HBR781 ; DYNDL1722, Lot Number 19KBR480 ; DYNDL1722, Lot Number 20DBT205 ; DYNDL1722, Lot Number 20FBO668 ; DYNDL1722, Lot Number 20HBL706 ; DYNDL1728, Lot Number 21JBG852 ; DYNDL1729, Lot Number 21JBL302 ; DYNDL1729, Lot Number 21JBW818 ; DYNDL1729, Lot Number 21VBC073 ; DYNDL1739, Lot Number 21HBK959 ; DYNDL1739, Lot Number 21JBJ533 ; DYNDL1739, Lot Number 21JBR454 ; DYNDL1739, Lot Number 21KBN675 ; DYNDL1739, Lot Number 22BBD497 ; DYNDL1739, Lot Number 19ABI068 ; DYNDL1739, Lot Number 22CBZ797 ; DYNDL1739, Lot Number 22DBT105 ; DYNDL1739, Lot Number 22EBG505 ; DYNDL1739, Lot Number 22FBS619 ; DYNDL1739, Lot Number 19SBH178 ; DYNDL1739, Lot Number 22JBE671 ; DYNDL1739, Lot Number 22JBQ770 ; DYNDL1739, Lot Number 22LBI007 ; DYNDL1739, Lot Number 19JBJ585 ; DYNDL1739, Lot Number 23ABT221 ; DYNDL1739, Lot Number 19KBT814 ; DYNDL1739, Lot Number 23CBB535 ; DYNDL1739, Lot Number 23DBF370 ; DYNDL1739, Lot Number 20BBA579 ; DYNDL1739, Lot Number 23GBB342 ; DYNDL1739, Lot Number 23HBH654 ; DYNDL1739, Lot Number 20EBD505 ; DYNDL1739, Lot Number 23JBK002 ; DYNDL1739, Lot Number 23KBE377 ; DYNDL1739, Lot Number 20GBL931 ; DYNDL1739, Lot Number 20HBC771 ; DYNDL1739, Lot Number 24ABX279 ; DYNDL1739, Lot Number 20KBT938 ; DYNDL1739, Lot Number 21BBK458 ; DYNDL1739, Lot Number 21DBX906 ; DYNDL1740, Lot Number 21JBI809 ; DYNDL1757, Lot Number 21IBR811 ; DYNDL1757, Lot Number 21JBL335 ; DYNDL1757, Lot Number 21KBV205 ; DYNDL1757, Lot Number 20DBU017 ; DYNDL1757, Lot Number 20FBK159 ; DYNDL1757, Lot Number 20HBC583 ; DYNDL1757, Lot Number 20JBD339 ; DYNDL1757, Lot Number 20JBU817 ; DYNDL1757, Lot Number 21ABI431 ; DYNDL1757, Lot Number 21BBH134 ; DYNDL1757, Lot Number 21DBF085 ; DYNDL1758, Lot Number 21JBG885 ; DYNDL1758H , Lot Number 21JBG885 ; DYNDL1762, Lot Number 21JBN292 ; DYNDL1768, Lot Number 21JBF329 ; DYNDL1770B , Lot Number 21JBG905 ; DYNDL1772, Lot Number 21GBH854 ; DYNDL1772, Lot Number 21IBA805 ; DYNDL1772, Lot Number 21JBF371 ; DYNDL1772, Lot Number 21JBY657 ; DYNDL1772, Lot Number 22ABM399 ; DYNDL1772, Lot Number 22EBB075 ; DYNDL1772, Lot Number 22FBO977 ; DYNDL1772, Lot Number 22GBI644 ; DYNDL1772, Lot Number 22HBP449 ; DYNDL1772, Lot Number 22JBX177 ; DYNDL1772, Lot Number 22KBW181 ; DYNDL1772, Lot Number 23FBT579 ; DYNDL1772, Lot Number 23HBG180 ; DYNDL1772, Lot Number 23HBZ363 ; DYNDL1772, Lot Number 23IBU678 ; DYNDL1772, Lot Number 20GBK293 ; DYNDL1772, Lot Number 20HBQ617 ; DYNDL1772, Lot Number 20KBP671 ; DYNDL1772, Lot Number 21BBF576 ; DYNDL1774A , Lot Number 19BBE494 ; DYNDL1783, Lot Number 21JBT687 ; DYNDL1790, Lot Number 21JBJ612 ; DYNDL1838, Lot Number 21JBI234 ; DYNDL1870, Lot Number 21JBT738 ; DYNDL1870H , Lot Number 21JBT738 ; DYNDL1872, Lot Number 21JBG868 ; DYNDL1878, Lot Number 21JBK569 ; DYNDL1878, Lot Number 21LBI313 ; DYNDL1878, Lot Number 19ABY817 ; DYNDL1878, Lot Number 22CBX097 ; DYNDL1878, Lot Number 22FBK820 ; DYNDL1878, Lot Number 19SBB932 ; DYNDL1878, Lot Number 22LBP014 ; DYNDL1878, Lot Number 23EBB536 ; DYNDL1878, Lot Number 20DBA409 ; DYNDL1878, Lot Number 23HBQ954 ; DYNDL1878, Lot Number 20FBK030 ; DYNDL1878, Lot Number 21BBH988 ; DYNDL1887, Lot Number 21GBM032 ; DYNDL1887, Lot Number 21IBS438 ; DYNDL1887, Lot Number 21JBR579 ; DYNDL1887, Lot Number 21LBB062 ; DYNDL1887, Lot Number 19ABV364 ; DYNDL1887, Lot Number 19CBH202 ; DYNDL1887, Lot Number 22EBB553 ; DYNDL1887, Lot Number 22FBT161 ; DYNDL1887, Lot Number 22HBO808 ; DYNDL1887, Lot Number 22IBS933 ; DYNDL1887, Lot Number 22KBO911 ; DYNDL1887, Lot Number 22LBL605 ; DYNDL1887, Lot Number 23ABF985 ; DYNDL1887, Lot Number 23BBP210 ; DYNDL1887, Lot Number 23CBT493 ; DYNDL1887, Lot Number 23EBV964 ; DYNDL1887, Lot Number 23FBK355 ; DYNDL1887, Lot Number 20BBW392 ; DYNDL1887, Lot Number 23GBO224 ; DYNDL1887, Lot Number 23JBK030 ; DYNDL1887, Lot Number 23KBK477 ; DYNDL1887, Lot Number 20HBK938 ; DYNDL1887, Lot Number 20HBX643 ; DYNDL1887, Lot Number 20JBZ511 ; DYNDL1887, Lot Number 21ABF626 ; DYNDL1887, Lot Number 21OBA305 ; DYNDL1888, Lot Number 21JBT707 ; DYNDL1892, Lot Number 21JBL268 ; DYNDL1920, Lot Number 21JBS611 ; DYNDL1920, Lot Number 21JBT598 ; DYNDL1923A , Lot Number 21JBT735 ; DYNDN1004, Lot Number 21JBR580 ; DYNDN1004, Lot Number 23JBM316 ; DYNJ0190604J , Lot Number 19AKC389 ; DYNJ0190604J , Lot Number 19CKA526 ; DYNJ03000, Lot Number 21JBA523 ; DYNJ03000, Lot Number 21JBA524 ; DYNJ03000H , Lot Number 21JBA523 ; DYNJ03000H , Lot Number 21JBA524 ; DYNJ03020, Lot Number 22DBT637; DYNJ03020H, Lot Number 22DBT637; DYNJ03021, Lot Number 21JBR937 ; DYNJ03159, Lot Number 23JBQ961 ; DYNJ03159H , Lot Number 23JBQ961 ; DYNJ07554, Lot Number 21JBR543 ; DYNJ07716E , Lot Number 21JBG815 ; DYNJ0832736Q , Lot Number 19IMB063 ; DYNJ0832736Q , Lot Number 19LMA545 ; DYNJ0832736Q , Lot Number 19LMB627 ; DYNJ0832736Q , Lot Number 20BMC599 ; DYNJ0832736Q , Lot Number 20FMA518 ; DYNJ0832736Q , Lot Number 20HMB764 ; DYNJ0832736Q , Lot Number 21AMA917 ; DYNJ0832736Q , Lot Number 21BME548 ; DYNJ0832736Q , Lot Number 21DME188 ; DYNJ0832736Q , Lot Number 21GMF418 ; DYNJ0832736Q , Lot Number 21IME597 ; DYNJ0832736Q , Lot Number 21KMH791 ; DYNJ0832736Q , Lot Number 22BMC092 ; DYNJ0832736Q , Lot Number 22CME209 ; DYNJ0832736Q , Lot Number 22EMB465 ; DYNJ0832736Q , Lot Number 22GMB121 ; DYNJ0832736Q , Lot Number 22GMB122 ; DYNJ0832736Q , Lot Number 22HME629 ; DYNJ0832736Q , Lot Number 22IMB667 ; DYNJ0832736Q , Lot Number 22KMA911 ; DYNJ0832736Q , Lot Number 22LMI809 ; DYNJ0832736Q , Lot Number 23CME844 ; DYNJ0832736Q , Lot Number 23EMB896 ; DYNJ0832736Q , Lot Number 23FMA856 ; DYNJ0832736Q , Lot Number 23GME182 ; DYNJ0832736Q , Lot Number 23HMD411 ; DYNJ0832736Q , Lot Number 23IMA770 ; DYNJ0832736Q , Lot Number 23JMD174 ; DYNJ0832736Q , Lot Number 23KMJ568 ; DYNJ0832736Q , Lot Number 19AKC772 ; DYNJ0832736Q , Lot Number 19BKD637 ; DYNJ0832736Q , Lot Number 19CKD804 ; DYNJ0832736Q , Lot Number 19EKA291 ; DYNJ0832736Q , Lot Number 19FKB157 ; DYNJ0832736Q , Lot Number 19GKC580 ; DYNJ17328B , Lot Number 21LBA554 ; DYNJ17328B , Lot Number 22ABS544 ; DYNJ17328B , Lot Number 19BBI032 ; DYNJ17328B , Lot Number 22OBF248 ; DYNJ17328B , Lot Number 19EBK668 ; DYNJ17328B , Lot Number 22GBV357 ; DYNJ17328B , Lot Number 19SBF546 ; DYNJ17328B , Lot Number 19JBR470 ; DYNJ17328B , Lot Number 23EBR157 ; DYNJ17328B , Lot Number 20BBX662 ; DYNJ17328B , Lot Number 20EBE998 ; DYNJ17328B , Lot Number 23IBB581 ; DYNJ17328B , Lot Number 23JBK939 ; DYNJ17328B , Lot Number 23JBX931 ; DYNJ17328B , Lot Number 21ABK182 ; DYNJ21270I , Lot Number 19GLA040 ; DYNJ21270I , Lot Number 19HLA192 ; DYNJ21270I , Lot Number 19HLA600 ; DYNJ21270I , Lot Number 19JLA291 ; DYNJ21270I , Lot Number 20ALA174 ; DYNJ21270I , Lot Number 20ALA863 ; DYNJ21270I , Lot Number 20BLA505 ; DYNJ21270I , Lot Number 20CLA191 ; DYNJ21270I , Lot Number 20DLA502 ; DYNJ21270I , Lot Number 20ELA649 ; DYNJ21270I , Lot Number 20FLA421 ; DYNJ21270I , Lot Number 20GLA402 ; DYNJ37902F , Lot Number 21GBA316 ; DYNJ37902F , Lot Number 19ABP912 ; DYNJ37902F , Lot Number 19CBL497 ; DYNJ37902F , Lot Number 19FBF912 ; DYNJ37902F , Lot Number 19FBO736 ; DYNJ37902F , Lot Number 19GBF678 ; DYNJ37902F , Lot Number 19GBL785 ; DYNJ37902F , Lot Number 19HBZ689 ; DYNJ37902F , Lot Number 19IBP980 ; DYNJ37902F , Lot Number 19JBM709 ; DYNJ37902F , Lot Number 19KBH434 ; DYNJ37902F , Lot Number 19LBW875 ; DYNJ37902F , Lot Number 20ABU358 ; DYNJ37902F , Lot Number 20CBE029 ; DYNJ37902F , Lot Number 20EBD028 ; DYNJ37902F , Lot Number 20FBL234 ; DYNJ37902F , Lot Number 20GBK113 ; DYNJ37902F , Lot Number 20HBV064 ; DYNJ37902F , Lot Number 20KBD102 ; DYNJ37902F , Lot Number 21BBS142 ; DYNJ37902F , Lot Number 21DBI004 ; DYNJ37902F , Lot Number 21FBA411 ; DYNJ38113B , Lot Number 21GBK591 ; DYNJ38113B , Lot Number 21JBN868 ; DYNJ38113B , Lot Number 21LBW183 ; DYNJ38113B , Lot Number 22ABU107 ; DYNJ38113B , Lot Number 19ABI886 ; DYNJ38113B , Lot Number 22EBJ955 ; DYNJ38113B , Lot Number 19PBA815 ; DYNJ38113B , Lot Number 19QBD166 ; DYNJ38113B , Lot Number 19FBA380 ; DYNJ38113B , Lot Number 19HBC971 ; DYNJ38113B , Lot Number 19IBS683 ; DYNJ38113B , Lot Number 19JBV593 ; DYNJ38113B , Lot Number 20ABF050 ; DYNJ38113B , Lot Number 20ABM640 ; DYNJ38113B , Lot Number 20CBH987 ; DYNJ38113B , Lot Number 21ABW407 ; DYNJ38113B , Lot Number 21CBO822 ; DYNJ38885B , Lot Number 21ABA217 ; DYNJ38885B , Lot Number 21ABF605 ; DYNJ38885B , Lot Number 21ABR581 ; DYNJ38885B , Lot Number 22JMI236 ; DYNJ38885B , Lot Number 23BMA797 ; DYNJ38885B , Lot Number 23BMA980 ; DYNJ38885B , Lot Number 23HME739 ; DYNJ38885B , Lot Number 23IMA814 ; DYNJ38885B , Lot Number 23LMF526 ; DYNJ42902B , Lot Number 21FBP919 ; DYNJ42902B , Lot Number 21GBR036 ; DYNJ42902B , Lot Number 21HBG909 ; DYNJ42902B , Lot Number 21IBD555 ; DYNJ42902B , Lot Number 21JBL668 ; DYNJ42902B , Lot Number 21LBB981 ; DYNJ42902B , Lot Number 19ABT683 ; DYNJ42902B , Lot Number 19BBJ890 ; DYNJ42902B , Lot Number 19CBF959 ; DYNJ42902B , Lot Number 19PBC204 ; DYNJ42902B , Lot Number 19EBK216 ; DYNJ42902B , Lot Number 19FBM954 ; DYNJ42902B , Lot Number 19GBQ811 ; DYNJ42902B , Lot Number 19HBE749 ; DYNJ42902B , Lot Number 19TBE208 ; DYNJ42902B , Lot Number 19JBI409 ; DYNJ42902B , Lot Number 19KBQ972 ; DYNJ42902B , Lot Number 19LBY759 ; DYNJ42902B , Lot Number 20ABG382 ; DYNJ42902B , Lot Number 20BBV434 ; DYNJ42902B , Lot Number 20DBO242 ; DYNJ42902B , Lot Number 20DBU991 ; DYNJ42902B , Lot Number 20GBI486 ; DYNJ42902B , Lot Number 20HBD206 ; DYNJ42902B , Lot Number 20WBH469 ; DYNJ42902B , Lot Number 20LBA901 ; DYNJ42902B , Lot Number 20XBG184 ; DYNJ42902B , Lot Number 21BBE289 ; DYNJ42902B , Lot Number 21CBG066 ; DYNJ42902B , Lot Number 21DBP933 ; DYNJ47717A , Lot Number 20JMF031 ; DYNJ47717A , Lot Number 20KME872 ; DYNJ47717A , Lot Number 20LMB931 ; DYNJ47717A , Lot Number 20LMC823 ; DYNJ47717A , Lot Number 21CMD153 ; DYNJ47717A , Lot Number 21EMC033 ; DYNJ47717A , Lot Number 21HMB521 ; DYNJ47717A , Lot Number 21HMF278 ; DYNJ47717A , Lot Number 21IME100 ; DYNJ47717A , Lot Number 21IMG896 ; DYNJ47717A , Lot Number 21JMG286 ; DYNJ47717A , Lot Number 22AME235 ; DYNJ47717A , Lot Number 22BMH056 ; DYNJ47717A , Lot Number 22EMF125 ; DYNJ47717A , Lot Number 22GMI215 ; DYNJ47717A , Lot Number 22JME782 ; DYNJ47717A , Lot Number 22LMC139 ; DYNJ47717A , Lot Number 22LMC336 ; DYNJ52141B , Lot Number 23JBS200 ; DYNJ52141B , Lot Number 23LBP043 ; DYNJ52141B , Lot Number 22BMD795 ; DYNJ52141B , Lot Number 23BMG613 ; DYNJ52141B , Lot Number 23BMH523 ; DYNJ52141B , Lot Number 23FMA347 ; DYNJ52995A , Lot Number 19CBA301 ; DYNJ52995A , Lot Number 19DBW166 ; DYNJ52995A , Lot Number 19FBE443 ; DYNJ52995A , Lot Number 19FBO779 ; DYNJ52995A , Lot Number 19SBG849 ; DYNJ55211B , Lot Number 19CBF581 ; DYNJ55211B , Lot Number 19VBH994 ; DYNJ60586, Lot Number 21FBM105 ; DYNJ60586, Lot Number 21GBV128 ; DYNJ60586, Lot Number 21HBT784 ; DYNJ60586, Lot Number 21IBO195 ; DYNJ60586, Lot Number 21JBM032 ; DYNJ60586, Lot Number 21KBX131 ; DYNJ60586, Lot Number 21LBG828 ; DYNJ60586, Lot Number 21LBU623 ; DYNJ60586, Lot Number 19ABG569 ; DYNJ60586, Lot Number 19ABR760 ; DYNJ60586, Lot Number 19BBQ710 ; DYNJ60586, Lot Number 19CBU664 ; DYNJ60586, Lot Number 19EBM595 ; DYNJ60586, Lot Number 19FBJ554 ; DYNJ60586, Lot Number 22HBA764 ; DYNJ60586, Lot Number 22IBA824 ; DYNJ60586, Lot Number 22IBO218 ; DYNJ60586, Lot Number 22JBQ331 ; DYNJ60586, Lot Number 19IBQ197 ; DYNJ60586, Lot Number 22LBT247 ; DYNJ60586, Lot Number 19VBF990 ; DYNJ60586, Lot Number 23ABR342 ; DYNJ60586, Lot Number 19KBT213 ; DYNJ60586, Lot Number 23CBG271 ; DYNJ60586, Lot Number 20ABD812 ; DYNJ60586, Lot Number 23EBR349 ; DYNJ60586, Lot Number 20CBH185 ; DYNJ60586, Lot Number 20DBC510 ; DYNJ60586, Lot Number 23HBR350 ; DYNJ60586, Lot Number 23IBF075 ; DYNJ60586, Lot Number 23JBO369 ; DYNJ60586, Lot Number 20GBS185 ; DYNJ60586, Lot Number 23KBQ986 ; DYNJ60586, Lot Number 20IBD558 ; DYNJ60586, Lot Number 20JBD786 ; DYNJ60586, Lot Number 20JBK289 ; DYNJ60586, Lot Number 21ABW403 ; DYNJ60586, Lot Number 21ABX243 ; DYNJ60586, Lot Number 21CBU925 ; DYNJ67990, Lot Number 21IME909 ; DYNJ67990, Lot Number 21KMC498 ; DYNJ67990, Lot Number 21KMI066 ; DYNJ67990, Lot Number 21LMD561 ; DYNJ67990, Lot Number 22AMC023 ; DYNJ67990, Lot Number 22AME196 ; DYNJ67990, Lot Number 22DMH679 ; DYNJ67990, Lot Number 22EMG464 ; DYNJ67990, Lot Number 22FMF131 ; DYNJ67990, Lot Number 22HMB880 ; DYNJ68223, Lot Number 21GBA785 ; DYNJ68223, Lot Number 21GBQ366 ; DYNJ68223, Lot Number 21HBB432 ; DYNJ68223, Lot Number 21IBW511 ; DYNJ68223, Lot Number 21BBB437 ; DYNJ68223, Lot Number 21DBE487 ; DYNJ68223, Lot Number 21EBE127 ; DYNJ68808, Lot Number 21GBN159 ; DYNJ68808, Lot Number 21IBU081 ; DYNJ68808, Lot Number 21LBT965 ; DYNJ68808, Lot Number 22NBG365 ; DYNJ68808, Lot Number 21DBC976 ; DYNJ80201, Lot Number 22ABY939 ; DYNJ80201A , Lot Number 22HBN480 ; DYNJ80201A , Lot Number 22IBE619 ; DYNJ80201A , Lot Number 22JBK565 ; DYNJ80201A , Lot Number 23BBM778 ; DYNJ80201A , Lot Number 23CBH775 ; DYNJ80201A , Lot Number 23DBJ566 ; DYNJ80201A , Lot Number 23EBK372 ; DYNJ80201A , Lot Number 23GBL847 ; DYNJ80201A , Lot Number 23IBD093 ; DYNJ80201A , Lot Number 23IBU061 ; DYNJ80201A , Lot Number 23JBS712 ; DYNJ80201A , Lot Number 23LBO062 ; DYNJ80201A , Lot Number 24ABU885 ; DYNJ83054A , Lot Number 22LME582 ; DYNJ83054A , Lot Number 23EMH715 ; DYNJ83054A , Lot Number 23KMA828 ; P392929, Lot Number 23JBH231 ; P948416, Lot Number 21GBF901 ; P948416, Lot Number 21IBI785 ; P948416, Lot Number 21LBF893 ; P948416, Lot Number 20BBZ489 ; P948416, Lot Number 20EBD522 ; P948416, Lot Number 20GBS490 ; P948416, Lot Number 20IBB944 ; P948416, Lot Number 20IBP670 ; P948416, Lot Number 21ABK378 ; P948416, Lot Number 22DMA834 ; P948416, Lot Number 22FMA125 ; P948416, Lot Number 22JME305 ; P948416, Lot Number 23DME672 ; P948416, Lot Number 23EMJ769 ; P948416, Lot Number 23HME168 ; P948416, Lot Number 23IMB508 ; SD-4060, Lot Number 21JBT597 ; SD-4060, Lot Number 23JBH120 ; SD-4060H , Lot Number 21JBT597 ; SD-4060H , Lot Number 23JBH120 ; STL121L, Lot Number 19BBR062 ; STL121L, Lot Number 19EBT327 ; STL121L, Lot Number 19HBN562 ; STL121L, Lot Number 19HBZ519 ; STL121L, Lot Number 19KBC986
Worldwide distribution.