FDA Enforcement Class II Ongoing

RSP Standard HUMERAL SOCKET INSERT, 44MM, HXe-Plus, Part No. 509-00-044

Recall: Z-3021-2020 · Reported October 7, 2020

Enforcement

Recall Number
Z-3021-2020
Event ID
86317
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Encore Medical, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 7, 2020
Initiation Date
August 7, 2020
Classification Date
September 25, 2020
Address
9800 Metric Blvd, N/A, Austin, TX, 78758-5445, United States

Description

RSP Standard HUMERAL SOCKET INSERT, 44MM, HXe-Plus, Part No. 509-00-044

Reason

Product may be mislabeled with the size etched onto products differing from the size information shown on patient sticker labels. Risks associated with this issue may include the Surgeon implanting the incorrect size; significant physical impairment, discomfort and/or disruption of treatment of underlying condition; and the patient experiencing dislocation resulting in revision surgery.

Code Info

Part 509-00-044, Lot 380P1009A , UDI: 00888912144568

Distribution

Distributed in the U.S. to: VA, MN, IN, RI, PA, NY, WA, KS, OK, FL, GA and CA

Quantity

15 units