FDA Enforcement
Class I
Terminated
ISH Protease 3, Catalog Number 05273331001, model 780-4149 Immunohistochemistry (IHC) for in vitro diagnostic use.
Recall: Z-3005-2018
·
Reported September 26, 2018
Enforcement
- Recall Number
- Z-3005-2018
- Event ID
- 80807
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ventana Medical Systems Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 26, 2018
- Initiation Date
- August 2, 2018
- Classification Date
- September 14, 2018
- Termination Date
- November 22, 2021
- Address
- 1910 E Innovation Park Dr, N/A, Oro Valley, AZ, 85755-1962, United States
Description
ISH Protease 3, Catalog Number 05273331001, model 780-4149 Immunohistochemistry (IHC) for in vitro diagnostic use.
Reason
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
Code Info
Lot Numbers: Y13927, Y18872, Y22569, Y25883, UDI: 4015630971886
Distribution
worldwide
Quantity
2231 units