FDA Enforcement Class I Terminated

OptiView Amplification Kit (250 test), Catalog Number 06718663001, model 860-099 Immunohistochemistry (IHC) for in vitro diagnostic use.

Recall: Z-3000-2018 · Reported September 26, 2018

Enforcement

Recall Number
Z-3000-2018
Event ID
80807
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Ventana Medical Systems Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 26, 2018
Initiation Date
August 2, 2018
Classification Date
September 14, 2018
Termination Date
November 22, 2021
Address
1910 E Innovation Park Dr, N/A, Oro Valley, AZ, 85755-1962, United States

Description

OptiView Amplification Kit (250 test), Catalog Number 06718663001, model 860-099 Immunohistochemistry (IHC) for in vitro diagnostic use.

Reason

Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.

Code Info

Lot Number: Y26282, UDI: 4015630984749

Distribution

worldwide

Quantity

461 units