FDA Enforcement Class II Ongoing

Speed Stitch Needle Cassette

Recall: Z-2999-2020 · Reported September 30, 2020

Enforcement

Recall Number
Z-2999-2020
Event ID
86290
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
ArthroCare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 30, 2020
Initiation Date
August 17, 2020
Classification Date
September 21, 2020
Address
7000 W William Cannon Dr Bldg 1, N/A, Austin, TX, 78735-8509, United States

Description

Speed Stitch Needle Cassette

Reason

Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.

Code Info

Model: OM-8850 Lot Numbers: 2018434, 2018934, 2019558, 2019559, 2019929, 2019930, 2020931, 2022111, 2022641, 2030008, 2032212, 2033815, 2037604, 2036941, 2038701, 2039699, 2040500, 2042295, 2044957, 2047483, 2048575, 2051239 & 2032526

Distribution

Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT,DB,FL,GA, HI, IA, ID, IL, IN, KS, KY, MA,MD,ME,MI,MN,MO,MS,MT, NC,ND, NE, NJ, NSW, NV,NY, OH,OK, OR, PA,SC, TN,TX, UT, VA, WA, WI, WV. International Distribution: AE, AU, CH, DE, FI, FR, GB, HK, IE, PA, PR, PT, RU, SG, and ZA

Quantity

6,366 needles