FDA Enforcement Class II Ongoing

Medline Convenience kits, labeled as: 1) GAS SAM LINE .06ID 10FT MM FLT, Pack Number 35772 ; 2) ANES CIRC 90 EXP PEDS 2BV-LF, Pack Number 81842 ; 3) AIRWAY EXAM KIT, Pack Number DYKE1796 ; 4) INTUBATION KIT , Pack Number DYKSCRIPPSIN4; 5) INTUBATION KIT , Pack Number DYNDA1598C; 6) INTUBATION KIT , Pack Number DYNDA1598CH ; 7) ANESTHESIA KIT , Pack Number DYNDA2958; 8) ANESTHESIA PREP CUSTOM TRAY, Pack Number DYNDH1758; 9) ADULT LUMBAR PUNCTURE TRAY , Pack Number DYNDH1889; 10) PEDIATRIC CARDIAC PACK-LF, Pack Number DYNJ0537013Y; 11) PEDIATRIC MINOR PK, Pack Number DYNJ49035B; 12) WMC PEDIATRIC ADD ON PACK-LF, Pack Number DYNJ50644; 13) ANESTHESIA CVL KIT-LF, Pack Number DYNJ52283B; 14) ANESTHESIA CVL KIT-LF, Pack Number DYNJ52283C; 15) AIRWAY PACK , Pack Number DYNJ60239B; 16) ANESTHESIA PAIN PK LEAGUE, Pack Number DYNJ67925; 17) H-CABG INTRA OP ANESTHESIA , Pack Number DYNJ900603I ; 18) H-CABG INTRA OP ANESTHESIA , Pack Number DYNJ900603J ; 19) CV ANESTHESIA - ROOM SET UP, Pack Number DYNJ905503A ; 20) CV ANESTHESIA - ROOM SET UP, Pack Number DYNJ905503B ; 21) CV ANESTHESIA - ROOM SET UP, Pack Number DYNJ905503C ; 22) CV ANESTHESIA - ROOM SET UP, Pack Number DYNJ905503D ; 23) CV ANESTHESIA - ROOM SET UP, Pack Number DYNJ905503F ; 24) #######, Pack Number DYNJ905503F ; 25) MZ MDL AIRWAY, Pack Number DYNJ910129; 26) ANESTHESIA CIRCUIT-LF, Pack Number DYNJAA10379 ; 27) ANESTHESIA CIRCUIT-LF, Pack Number DYNJAA6572; 28) ANESTHESIA EVALUATION SAMPLE, Pack Number DYNJAASAMPLE; 29) ANESTHESIA EVALUATION SAMPLE, Pack Number DYNJAASAMPLE; 30) REGIONAL ANESTHESIA PREP TRAY , Pack Number DYNJRA0376A ; 31) ANESTHESIA KIT , Pack Number DYNJRA0424A ; 32) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA0767; 33) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA0767A ; 34) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA0767B ; 35) ANESTHESIA SUPPORT TRAY , Pack Number DYNJRA0779A ; 36) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA0887A ; 37) ANESTHESIA TRAY, Pack Number DYNJRA0909; 38) SPINAL ANESTHESIA TRAY/WHITE, Pack Number DYNJRA0955A ; 39) #####, Pack Number DYNJRA0955A ; 40) ANESTHESIA SUPPORT TRAY/B, Pack Number DYNJRA0968; 41) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA1119; 42) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA1119A ; 43) ANESTHESIA TRAY, Pack Number DYNJRA1355; 44) ANESTHESIA TRAY, Pack Number DYNJRA1355A ; 45) SPINAL ANESTHESIA TRAY, Pack Number DYNJRA1577A ; 46) ANESTHESIA SUPPORT TRAY/CHLORA, Pack Number DYNJRA1659; 47) ANESTHESIA SUPPORT TRAY/CHLORA, Pack Number DYNJRA1659A ; 48) SPINAL ANESTHESIA TRAY, Pack Number DYNJRA1757; 49) ANESTHESIA PREP KIT, Pack Number DYNJRA1899; 50) ANESTHESIA TRAY, Pack Number DYNJRA1934; 51) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA2087; 52) SPINAL ANESTHESIA TRAY, Pack Number DYNJRA2151; 53) PRE OP ANES KIT, Pack Number DYNJRA2228; 54) PRE OP ANES KIT, Pack Number DYNJRA2228H; 55) CUSTOM ANESTHESIA KIT-LF, Pack Number GSP478I; 56) REGIONAL TRAY, Pack Number PAIN0385B; 57) EPIDURAL ANESTHESIA KIT , Pack Number PAIN1982 ; 58) CIRC PACK-LF, Pack Number PHS41720I; 59) CIRC PACK-LF, Pack Number PHS41720J; 60) ANES CIRC 90 EXP, Pack Number VM96DLEX2L

Recall: Z-2983-2024 · Reported September 18, 2024

Enforcement

Recall Number
Z-2983-2024
Event ID
94583
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
MEDLINE INDUSTRIES, LP - Northfield
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 18, 2024
Initiation Date
April 8, 2024
Classification Date
September 6, 2024
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline Convenience kits, labeled as: 1) GAS SAM LINE .06ID 10FT MM FLT, Pack Number 35772 ; 2) ANES CIRC 90 EXP PEDS 2BV-LF, Pack Number 81842 ; 3) AIRWAY EXAM KIT, Pack Number DYKE1796 ; 4) INTUBATION KIT , Pack Number DYKSCRIPPSIN4; 5) INTUBATION KIT , Pack Number DYNDA1598C; 6) INTUBATION KIT , Pack Number DYNDA1598CH ; 7) ANESTHESIA KIT , Pack Number DYNDA2958; 8) ANESTHESIA PREP CUSTOM TRAY, Pack Number DYNDH1758; 9) ADULT LUMBAR PUNCTURE TRAY , Pack Number DYNDH1889; 10) PEDIATRIC CARDIAC PACK-LF, Pack Number DYNJ0537013Y; 11) PEDIATRIC MINOR PK, Pack Number DYNJ49035B; 12) WMC PEDIATRIC ADD ON PACK-LF, Pack Number DYNJ50644; 13) ANESTHESIA CVL KIT-LF, Pack Number DYNJ52283B; 14) ANESTHESIA CVL KIT-LF, Pack Number DYNJ52283C; 15) AIRWAY PACK , Pack Number DYNJ60239B; 16) ANESTHESIA PAIN PK LEAGUE, Pack Number DYNJ67925; 17) H-CABG INTRA OP ANESTHESIA , Pack Number DYNJ900603I ; 18) H-CABG INTRA OP ANESTHESIA , Pack Number DYNJ900603J ; 19) CV ANESTHESIA - ROOM SET UP, Pack Number DYNJ905503A ; 20) CV ANESTHESIA - ROOM SET UP, Pack Number DYNJ905503B ; 21) CV ANESTHESIA - ROOM SET UP, Pack Number DYNJ905503C ; 22) CV ANESTHESIA - ROOM SET UP, Pack Number DYNJ905503D ; 23) CV ANESTHESIA - ROOM SET UP, Pack Number DYNJ905503F ; 24) #######, Pack Number DYNJ905503F ; 25) MZ MDL AIRWAY, Pack Number DYNJ910129; 26) ANESTHESIA CIRCUIT-LF, Pack Number DYNJAA10379 ; 27) ANESTHESIA CIRCUIT-LF, Pack Number DYNJAA6572; 28) ANESTHESIA EVALUATION SAMPLE, Pack Number DYNJAASAMPLE; 29) ANESTHESIA EVALUATION SAMPLE, Pack Number DYNJAASAMPLE; 30) REGIONAL ANESTHESIA PREP TRAY , Pack Number DYNJRA0376A ; 31) ANESTHESIA KIT , Pack Number DYNJRA0424A ; 32) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA0767; 33) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA0767A ; 34) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA0767B ; 35) ANESTHESIA SUPPORT TRAY , Pack Number DYNJRA0779A ; 36) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA0887A ; 37) ANESTHESIA TRAY, Pack Number DYNJRA0909; 38) SPINAL ANESTHESIA TRAY/WHITE, Pack Number DYNJRA0955A ; 39) #####, Pack Number DYNJRA0955A ; 40) ANESTHESIA SUPPORT TRAY/B, Pack Number DYNJRA0968; 41) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA1119; 42) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA1119A ; 43) ANESTHESIA TRAY, Pack Number DYNJRA1355; 44) ANESTHESIA TRAY, Pack Number DYNJRA1355A ; 45) SPINAL ANESTHESIA TRAY, Pack Number DYNJRA1577A ; 46) ANESTHESIA SUPPORT TRAY/CHLORA, Pack Number DYNJRA1659; 47) ANESTHESIA SUPPORT TRAY/CHLORA, Pack Number DYNJRA1659A ; 48) SPINAL ANESTHESIA TRAY, Pack Number DYNJRA1757; 49) ANESTHESIA PREP KIT, Pack Number DYNJRA1899; 50) ANESTHESIA TRAY, Pack Number DYNJRA1934; 51) ANESTHESIA SUPPORT TRAY-LF , Pack Number DYNJRA2087; 52) SPINAL ANESTHESIA TRAY, Pack Number DYNJRA2151; 53) PRE OP ANES KIT, Pack Number DYNJRA2228; 54) PRE OP ANES KIT, Pack Number DYNJRA2228H; 55) CUSTOM ANESTHESIA KIT-LF, Pack Number GSP478I; 56) REGIONAL TRAY, Pack Number PAIN0385B; 57) EPIDURAL ANESTHESIA KIT , Pack Number PAIN1982 ; 58) CIRC PACK-LF, Pack Number PHS41720I; 59) CIRC PACK-LF, Pack Number PHS41720J; 60) ANES CIRC 90 EXP, Pack Number VM96DLEX2L

Reason

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

Code Info

35772, Lot Number 19GBK969 ; 81842, Lot Number 20JBS956 ; DYKE1796 , Lot Number 21GBR139 ; DYKE1796 , Lot Number 22HBX360 ; DYKE1796 , Lot Number 22JBR572 ; DYKSCRIPPSIN4, Lot Number 19DAA326 ; DYKSCRIPPSIN4, Lot Number 19EAA345 ; DYKSCRIPPSIN4, Lot Number 19HAB271 ; DYKSCRIPPSIN4, Lot Number 19LAA719 ; DYKSCRIPPSIN4, Lot Number 20AAA102 ; DYKSCRIPPSIN4, Lot Number 20BAA611 ; DYKSCRIPPSIN4, Lot Number 20DAA064 ; DYNDA1598C , Lot Number 23KMF159 ; DYNDA1598C , Lot Number 24AMG428 ; DYNDA1598CH, Lot Number 23KMF159 ; DYNDA1598CH, Lot Number 24AMG428 ; DYNDA2958, Lot Number 22KBH750 ; DYNDH1758, Lot Number 22HBX410 ; DYNDH1758, Lot Number 22JBU749 ; DYNDH1758, Lot Number 22LBK163 ; DYNDH1889, Lot Number 23HBZ539 ; DYNDH1889, Lot Number 23KBG134 ; DYNJ0537013Y , Lot Number 21LMD783 ; DYNJ49035B , Lot Number 19BDA167 ; DYNJ49035B , Lot Number 19CDC764 ; DYNJ49035B , Lot Number 19EDD455 ; DYNJ49035B , Lot Number 19GDC256 ; DYNJ49035B , Lot Number 19IDC024 ; DYNJ49035B , Lot Number 20ADA825 ; DYNJ49035B , Lot Number 20BDC864 ; DYNJ49035B , Lot Number 20DDA660 ; DYNJ49035B , Lot Number 20GDB084 ; DYNJ49035B , Lot Number 20IDB475 ; DYNJ49035B , Lot Number 20LDA638 ; DYNJ50644, Lot Number 21KBE704 ; DYNJ50644, Lot Number 22OBH075 ; DYNJ50644, Lot Number 22HBD748 ; DYNJ50644, Lot Number 19HBZ833 ; DYNJ50644, Lot Number 19JBI231 ; DYNJ50644, Lot Number 19VBB145 ; DYNJ50644, Lot Number 20BBB196 ; DYNJ50644, Lot Number 23JBW861 ; DYNJ52283B , Lot Number 19ABX422 ; DYNJ52283B , Lot Number 19BBT964 ; DYNJ52283C , Lot Number 21IBF288 ; DYNJ52283C , Lot Number 21JBX031 ; DYNJ52283C , Lot Number 21LBB738 ; DYNJ52283C , Lot Number 22ABA824 ; DYNJ52283C , Lot Number 22OBI904 ; DYNJ52283C , Lot Number 19DBO272 ; DYNJ52283C , Lot Number 19DBU670 ; DYNJ52283C , Lot Number 19FBM983 ; DYNJ52283C , Lot Number 22HBE294 ; DYNJ52283C , Lot Number 19GBL569 ; DYNJ52283C , Lot Number 19GBW600 ; DYNJ52283C , Lot Number 19SBC949 ; DYNJ52283C , Lot Number 22JBN210 ; DYNJ52283C , Lot Number 19JBT368 ; DYNJ52283C , Lot Number 19JBV109 ; DYNJ52283C , Lot Number 19VBH492 ; DYNJ52283C , Lot Number 19KBJ678 ; DYNJ52283C , Lot Number 19KBN283 ; DYNJ52283C , Lot Number 23BBU977 ; DYNJ52283C , Lot Number 19LBK922 ; DYNJ52283C , Lot Number 19LBS004 ; DYNJ52283C , Lot Number 20ABM461 ; DYNJ52283C , Lot Number 20ABP865 ; DYNJ52283C , Lot Number 23EBL084 ; DYNJ52283C , Lot Number 20BBJ414 ; DYNJ52283C , Lot Number 20BBS280 ; DYNJ52283C , Lot Number 23FBP678 ; DYNJ52283C , Lot Number 23HBK734 ; DYNJ52283C , Lot Number 20HBH238 ; DYNJ52283C , Lot Number 20JBJ550 ; DYNJ52283C , Lot Number 20JBN445 ; DYNJ52283C , Lot Number 20KBA832 ; DYNJ52283C , Lot Number 20KBP071 ; DYNJ52283C , Lot Number 21ABN133 ; DYNJ60239B , Lot Number 23LDA663 ; DYNJ67925, Lot Number 21GBR614 ; DYNJ67925, Lot Number 21JBL671 ; DYNJ67925, Lot Number 21LBO778 ; DYNJ67925, Lot Number 21LBO779 ; DYNJ67925, Lot Number 21LBO780 ; DYNJ67925, Lot Number 21LBO781 ; DYNJ67925, Lot Number 22ABA751 ; DYNJ67925, Lot Number 22ABQ043 ; DYNJ67925, Lot Number 22BBL497 ; DYNJ67925, Lot Number 22GBR405 ; DYNJ67925, Lot Number 22HBF644 ; DYNJ67925, Lot Number 22IBQ399 ; DYNJ67925, Lot Number 22KBU314 ; DYNJ67925, Lot Number 23ABO072 ; DYNJ67925, Lot Number 23ABT720 ; DYNJ67925, Lot Number 23CBF754 ; DYNJ67925, Lot Number 23DBP494 ; DYNJ67925, Lot Number 23FBJ744 ; DYNJ67925, Lot Number 23GBQ403 ; DYNJ67925, Lot Number 23HBE106 ; DYNJ67925, Lot Number 21ABA903 ; DYNJ67925, Lot Number 21ABW854 ; DYNJ67925, Lot Number 21DBP686 ; DYNJ67925, Lot Number 21EBH092 ; DYNJ900603I, Lot Number 20DDB314 ; DYNJ900603I, Lot Number 20FDB651 ; DYNJ900603I, Lot Number 20GDA926 ; DYNJ900603I, Lot Number 20HDA245 ; DYNJ900603I, Lot Number 20JDB313 ; DYNJ900603I, Lot Number 20JDB761 ; DYNJ900603I, Lot Number 20LDB998 ; DYNJ900603I, Lot Number 21ADB434 ; DYNJ900603I, Lot Number 21BDA664 ; DYNJ900603J, Lot Number 21HDC941 ; DYNJ900603J, Lot Number 21IDB264 ; DYNJ905503A, Lot Number 19CBB091 ; DYNJ905503A, Lot Number 19DBG305 ; DYNJ905503A, Lot Number 19PBB724 ; DYNJ905503A, Lot Number 19QBB679 ; DYNJ905503B, Lot Number 19FBZ308 ; DYNJ905503B, Lot Number 19HBV183 ; DYNJ905503B, Lot Number 19IBY178 ; DYNJ905503B, Lot Number 19LBP039 ; DYNJ905503C, Lot Number 20MBA222 ; DYNJ905503C, Lot Number 20CBF593 ; DYNJ905503C, Lot Number 20EBK549 ; DYNJ905503C, Lot Number 20EBO464 ; DYNJ905503C, Lot Number 20HBU177 ; DYNJ905503D, Lot Number 21FBN751 ; DYNJ905503D, Lot Number 20LBP237 ; DYNJ905503D, Lot Number 21DBG611 ; DYNJ905503D, Lot Number 21EBI605 ; DYNJ905503F, Lot Number 21HBL060 ; DYNJ905503F, Lot Number 21IBW859 ; DYNJ905503F, Lot Number 22DBC340 ; DYNJ905503F, Lot Number 22DBS437 ; DYNJ905503F, Lot Number 22FBR267 ; DYNJ905503F, Lot Number 22GBR286 ; DYNJ905503F, Lot Number 22HBS560 ; DYNJ905503F, Lot Number 22IBA577 ; DYNJ905503F, Lot Number 22JBI883 ; DYNJ905503F, Lot Number 22KBD587 ; DYNJ905503F, Lot Number 22LBF425 ; DYNJ905503F, Lot Number 23BBN903 ; DYNJ905503F, Lot Number 23CBL450 ; DYNJ905503F, Lot Number 23CBW776 ; DYNJ905503F, Lot Number 23FBC209 ; DYNJ905503F, Lot Number 23IBU343 ; DYNJ905503F, Lot Number 23JBP494 ; DYNJ905503F, Lot Number 23KBH749 ; DYNJ905503F, Lot Number 23KBQ236 ; DYNJ905503F, Lot Number 23LBR048 ; DYNJ905503F, Lot Number 24BBH974 ; DYNJ910129 , Lot Number 24AMF749 ; DYNJAA10379, Lot Number 19HBV537 ; DYNJAA6572 , Lot Number 20FBL919 ; DYNJAASAMPLE, Lot Number 21IBA814; DYNJAASAMPLE, Lot Number 21IBA815; DYNJAASAMPLE, Lot Number 21HBE107; DYNJAASAMPLE, Lot Number 21FBU433; DYNJAASAMPLE, Lot Number 21EBM583; DYNJAASAMPLE , Lot Number 19HBP877 ; DYNJAASAMPLE , Lot Number 19KBF202 ; DYNJRA0376A, Lot Number 21JBA836 ; DYNJRA0376A, Lot Number 21KBI074 ; DYNJRA0376A, Lot Number 21KBP359 ; DYNJRA0376A, Lot Number 21KBU250 ; DYNJRA0376A, Lot Number 22ABU499 ; DYNJRA0376A, Lot Number 19ABT679 ; DYNJRA0376A, Lot Number 19BBH464 ; DYNJRA0376A, Lot Number 19CBM742 ; DYNJRA0376A, Lot Number 19DBV170 ; DYNJRA0376A, Lot Number 19EBC310 ; DYNJRA0376A, Lot Number 19EBZ313 ; DYNJRA0376A, Lot Number 19FBP671 ; DYNJRA0376A, Lot Number 22JBU142 ; DYNJRA0376A, Lot Number 19IBN472 ; DYNJRA0376A, Lot Number 19JBN388 ; DYNJRA0376A, Lot Number 23ABN343 ; DYNJRA0376A, Lot Number 19WBB195 ; DYNJRA0376A, Lot Number 23DBN153 ; DYNJRA0376A, Lot Number 20ABP822 ; DYNJRA0376A, Lot Number 20BBG986 ; DYNJRA0376A, Lot Number 23FBJ727 ; DYNJRA0376A, Lot Number 20FBS238 ; DYNJRA0376A, Lot Number 23JBT431 ; DYNJRA0376A, Lot Number 20JBB325 ; DYNJRA0376A, Lot Number 20JBS327 ; DYNJRA0376A, Lot Number 21CBC005 ; DYNJRA0376A, Lot Number 21FBA847 ; DYNJRA0376A, Lot Number 21FBF515 ; DYNJRA0424A, Lot Number 19BLA192 ; DYNJRA0424A, Lot Number 19ELA369 ; DYNJRA0424A, Lot Number 19ILA097 ; DYNJRA0424A, Lot Number 19ILA581 ; DYNJRA0424A, Lot Number 20BLA165 ; DYNJRA0424A, Lot Number 20DLA320 ; DYNJRA0424A, Lot Number 20ILA236 ; DYNJRA0424A, Lot Number 20ILA689 ; DYNJRA0424A, Lot Number 20JLA266 ; DYNJRA0767 , Lot Number 19ABB420 ; DYNJRA0767 , Lot Number 19ABB495 ; DYNJRA0767 , Lot Number 19BBB590 ; DYNJRA0767 , Lot Number 19EBV290 ; DYNJRA0767A, Lot Number 19GLA471 ; DYNJRA0767A, Lot Number 19JLA131 ; DYNJRA0767A, Lot Number 19KLA574 ; DYNJRA0767A, Lot Number 20ALA762 ; DYNJRA0767B, Lot Number 20BLA575 ; DYNJRA0767B, Lot Number 20DLA137 ; DYNJRA0767B, Lot Number 20GLA730 ; DYNJRA0767B, Lot Number 20ILA116 ; DYNJRA0767B, Lot Number 20KLA503 ; DYNJRA0767B, Lot Number 20LLA193 ; DYNJRA0767B, Lot Number 21ALA504 ; DYNJRA0767B, Lot Number 21DLA280 ; DYNJRA0767B, Lot Number 21ELA804 ; DYNJRA0767B, Lot Number 21GLA351 ; DYNJRA0779A, Lot Number 19BBE138 ; DYNJRA0887A, Lot Number 21HBJ813 ; DYNJRA0887A, Lot Number 21KBF202 ; DYNJRA0887A, Lot Number 22ABL039 ; DYNJRA0887A, Lot Number 19ABB764 ; DYNJRA0887A, Lot Number 22BBO561 ; DYNJRA0887A, Lot Number 22DBL385 ; DYNJRA0887A, Lot Number 22DBL387 ; DYNJRA0887A, Lot Number 19DBQ652 ; DYNJRA0887A, Lot Number 22EBN188 ; DYNJRA0887A, Lot Number 19EBM156 ; DYNJRA0887A, Lot Number 22FBN499 ; DYNJRA0887A, Lot Number 19FBG819 ; DYNJRA0887A, Lot Number 19GBN973 ; DYNJRA0887A, Lot Number 20ABG794 ; DYNJRA0887A, Lot Number 23EBA855 ; DYNJRA0887A, Lot Number 20CBA118 ; DYNJRA0887A, Lot Number 20CBT347 ; DYNJRA0887A, Lot Number 20CBZ291 ; DYNJRA0887A, Lot Number 23HBX546 ; DYNJRA0887A, Lot Number 20EBP521 ; DYNJRA0887A, Lot Number 23IBM658 ; DYNJRA0887A, Lot Number 20GBA834 ; DYNJRA0887A, Lot Number 23JBY667 ; DYNJRA0887A, Lot Number 23LBA050 ; DYNJRA0887A, Lot Number 20KBS435 ; DYNJRA0887A, Lot Number 20WBA633 ; DYNJRA0887A, Lot Number 20LBU607 ; DYNJRA0909 , Lot Number 21IBM432 ; DYNJRA0909 , Lot Number 21JBB919 ; DYNJRA0909 , Lot Number 21JBJ840 ; DYNJRA0909 , Lot Number 21LBI511 ; DYNJRA0909 , Lot Number 22BBK053 ; DYNJRA0909 , Lot Number 19ABK394 ; DYNJRA0909 , Lot Number 22CBV876 ; DYNJRA0909 , Lot Number 19DBH758 ; DYNJRA0909 , Lot Number 22EBB982 ; DYNJRA0909 , Lot Number 19EBX778 ; DYNJRA0909 , Lot Number 19SBA717 ; DYNJRA0909 , Lot Number 19IBH329 ; DYNJRA0909 , Lot Number 22KBK484 ; DYNJRA0909 , Lot Number 22LBH510 ; DYNJRA0909 , Lot Number 19KBD605 ; DYNJRA0909 , Lot Number 23BBS655 ; DYNJRA0909 , Lot Number 19KBX403 ; DYNJRA0909 , Lot Number 23DBC885 ; DYNJRA0909 , Lot Number 20BBQ111 ; DYNJRA0909 , Lot Number 20BBR726 ; DYNJRA0909 , Lot Number 23FBK613 ; DYNJRA0909 , Lot Number 23GBT449 ; DYNJRA0909 , Lot Number 23IBL594 ; DYNJRA0909 , Lot Number 20FBN090 ; DYNJRA0909 , Lot Number 23KBC242 ; DYNJRA0909 , Lot Number 20HBN921 ; DYNJRA0909 , Lot Number 23LBU798 ; DYNJRA0909 , Lot Number 20JBS343 ; DYNJRA0909 , Lot Number 20LBP468 ; DYNJRA0909 , Lot Number 21OBA552 ; DYNJRA0909 , Lot Number 21FBD258 ; DYNJRA0955A, Lot Number 19ADB360 ; DYNJRA0955A, Lot Number 19EDB618 ; DYNJRA0955A, Lot Number 19HDA472 ; DYNJRA0955A, Lot Number 20ADC459 ; DYNJRA0955A, Lot Number 20CDB147 ; DYNJRA0955A, Lot Number 20FDC321 ; DYNJRA0955A, Lot Number 20HDC485 ; DYNJRA0955A, Lot Number 20JDA615 ; DYNJRA0955A, Lot Number 20LDA004 ; DYNJRA0955A, Lot Number 20LDB451 ; DYNJRA0955A, Lot Number 21ADA927 ; DYNJRA0955A, Lot Number 21EDB811 ; DYNJRA0955A, Lot Number 21IDB370 ; DYNJRA0955A, Lot Number 21JDA420 ; DYNJRA0955A, Lot Number 21KDB440 ; DYNJRA0955A, Lot Number 21LDA933 ; DYNJRA0955A, Lot Number 22BDB266 ; DYNJRA0955A, Lot Number 22EDB897 ; DYNJRA0955A, Lot Number 22JDA402 ; DYNJRA0955A, Lot Number 22LDB425 ; DYNJRA0955A, Lot Number 23BDA481 ; DYNJRA0955A, Lot Number 23EDA343 ; DYNJRA0955A, Lot Number 23IDB489 ; DYNJRA0955A, Lot Number 23LDA640 ; DYNJRA0955A, Lot Number 24ADC151 ; DYNJRA0955A, Lot Number 24CDA833 ; DYNJRA0968 , Lot Number 21HBF288 ; DYNJRA0968 , Lot Number 21JBT830 ; DYNJRA0968 , Lot Number 21LBY093 ; DYNJRA0968 , Lot Number 19BBE830 ; DYNJRA0968 , Lot Number 22OBI659 ; DYNJRA0968 , Lot Number 19DBR761 ; DYNJRA0968 , Lot Number 22EBT662 ; DYNJRA0968 , Lot Number 19EBR863 ; DYNJRA0968 , Lot Number 19FBV435 ; DYNJRA0968 , Lot Number 22JBL305 ; DYNJRA0968 , Lot Number 19IBU668 ; DYNJRA0968 , Lot Number 19KBO051 ; DYNJRA0968 , Lot Number 20BBQ119 ; DYNJRA0968 , Lot Number 20EBA140 ; DYNJRA0968 , Lot Number 20HBV152 ; DYNJRA1119 , Lot Number 19ABB888 ; DYNJRA1119 , Lot Number 19ABF794 ; DYNJRA1119 , Lot Number 19DBS320 ; DYNJRA1119 , Lot Number 19FBU183 ; DYNJRA1119 , Lot Number 19JBJ900 ; DYNJRA1119 , Lot Number 19LBI573 ; DYNJRA1119 , Lot Number 20CBO496 ; DYNJRA1119 , Lot Number 20EBA135 ; DYNJRA1119 , Lot Number 20FBC818 ; DYNJRA1119 , Lot Number 20KBC430 ; DYNJRA1119 , Lot Number 20KBS507 ; DYNJRA1119 , Lot Number 20WBA649 ; DYNJRA1119 , Lot Number 20XBA718 ; DYNJRA1119A, Lot Number 21LBM005 ; DYNJRA1119A, Lot Number 22OBE977 ; DYNJRA1119A, Lot Number 22OBG228 ; DYNJRA1119A, Lot Number 22OBH490 ; DYNJRA1119A, Lot Number 22DBL388 ; DYNJRA1119A, Lot Number 22DBL389 ; DYNJRA1119A, Lot Number 22DBL390 ; DYNJRA1119A, Lot Number 22HBM810 ; DYNJRA1119A, Lot Number 22IBT110 ; DYNJRA1355 , Lot Number 19DDB243 ; DYNJRA1355 , Lot Number 20EDB062 ; DYNJRA1355 , Lot Number 20KDA418 ; DYNJRA1355 , Lot Number 21HDA638 ; DYNJRA1355 , Lot Number 21KDC271 ; DYNJRA1355 , Lot Number 22BDB953 ; DYNJRA1355 , Lot Number 22CDA294 ; DYNJRA1355A, Lot Number 23EBS294 ; DYNJRA1355A, Lot Number 23FBB136 ; DYNJRA1355A, Lot Number 23LBU793 ; DYNJRA1577A, Lot Number 20BLA161 ; DYNJRA1577A, Lot Number 20GLA679 ; DYNJRA1577A, Lot Number 20HLA024 ; DYNJRA1577A, Lot Number 21DLA789 ; DYNJRA1577A, Lot Number 21HLA237 ; DYNJRA1577A, Lot Number 21JLA880 ; DYNJRA1577A, Lot Number 22ALA178 ; DYNJRA1577A, Lot Number 22FLB148 ; DYNJRA1577A, Lot Number 22JLA138 ; DYNJRA1577A, Lot Number 22KLA017 ; DYNJRA1659 , Lot Number 20GBL316 ; DYNJRA1659A, Lot Number 21HBL723 ; DYNJRA1659A, Lot Number 21KBD763 ; DYNJRA1659A, Lot Number 22ABH121 ; DYNJRA1659A, Lot Number 22BBH682 ; DYNJRA1659A, Lot Number 22EBO710 ; DYNJRA1659A, Lot Number 20JBK938 ; DYNJRA1659A, Lot Number 20KBS464 ; DYNJRA1659A, Lot Number 20KBX281 ; DYNJRA1659A, Lot Number 20XBB194 ; DYNJRA1659A, Lot Number 21DBO798 ; DYNJRA1659A, Lot Number 21DBU473 ; DYNJRA1659A, Lot Number 22GMG057 ; DYNJRA1659A, Lot Number 22JMG120 ; DYNJRA1757 , Lot Number 21GBH846 ; DYNJRA1757 , Lot Number 21HBI684 ; DYNJRA1757 , Lot Number 21HBL719 ; DYNJRA1757 , Lot Number 21JBB912 ; DYNJRA1757 , Lot Number 21JBX538 ; DYNJRA1757 , Lot Number 21LBP546 ; DYNJRA1757 , Lot Number 22ABN356 ; DYNJRA1757 , Lot Number 22ABU537 ; DYNJRA1757 , Lot Number 22EBI319 ; DYNJRA1757 , Lot Number 22GBD823 ; DYNJRA1757 , Lot Number 22HBL721 ; DYNJRA1757 , Lot Number 22LBC462 ; DYNJRA1757 , Lot Number 23ABF387 ; DYNJRA1757 , Lot Number 23BBC744 ; DYNJRA1757 , Lot Number 23CBR306 ; DYNJRA1757 , Lot Number 23EBE047 ; DYNJRA1757 , Lot Number 23GBM489 ; DYNJRA1757 , Lot Number 23IBW129 ; DYNJRA1757 , Lot Number 23KBX075 ; DYNJRA1757 , Lot Number 21CBD668 ; DYNJRA1757 , Lot Number 21EBI263 ; DYNJRA1757 , Lot Number 21FBC480 ; DYNJRA1899 , Lot Number 22BBB022 ; DYNJRA1899 , Lot Number 22FBN862 ; DYNJRA1899 , Lot Number 23ABG680 ; DYNJRA1899 , Lot Number 23CBR650 ; DYNJRA1899 , Lot Number 23EBX287 ; DYNJRA1899 , Lot Number 23KBW986 ; DYNJRA1934 , Lot Number 22FBG297 ; DYNJRA1934 , Lot Number 22JBO101 ; DYNJRA1934 , Lot Number 23GBP086 ; DYNJRA1934 , Lot Number 23IBM814 ; DYNJRA1934 , Lot Number 23KBB573 ; DYNJRA1934 , Lot Number 23LBB359 ; DYNJRA2087 , Lot Number 23ABU062 ; DYNJRA2087 , Lot Number 23FBK599 ; DYNJRA2151 , Lot Number 23DBD158 ; DYNJRA2228, Lot Number 23IBH340; DYNJRA2228H, Lot Number 23IBH340; GSP478I, Lot Number 21GBH137 ; GSP478I, Lot Number 19DBC561 ; GSP478I, Lot Number 19EBJ393 ; GSP478I, Lot Number 19GBY924 ; GSP478I, Lot Number 19TBC432 ; GSP478I, Lot Number 19JBI604 ; GSP478I, Lot Number 19KBD438 ; GSP478I, Lot Number 20ABK386 ; GSP478I, Lot Number 20ABO197 ; GSP478I, Lot Number 20EBN095 ; GSP478I, Lot Number 20GBG243 ; GSP478I, Lot Number 20JBN525 ; GSP478I, Lot Number 20KBK235 ; GSP478I, Lot Number 20WBB596 ; GSP478I, Lot Number 21BBI524 ; GSP478I, Lot Number 21DBN817 ; PAIN0385B, Lot Number 19EDB505 ; PAIN1982 , Lot Number 23IBE536 ; PAIN1982 , Lot Number 23KBC691 ; PAIN1982 , Lot Number 23LBC334 ; PAIN1982 , Lot Number 23LBC459 ; PHS41720I, Lot Number 20JMC326 ; PHS41720I, Lot Number 20KMF038 ; PHS41720I, Lot Number 19KKB847 ; PHS41720I, Lot Number 19LKB657 ; PHS41720I, Lot Number 20EKB947 ; PHS41720I, Lot Number 20GKB697 ; PHS41720I, Lot Number 20IKB163 ; PHS41720J, Lot Number 21BMD711 ; PHS41720J, Lot Number 21CME416 ; PHS41720J, Lot Number 21GMB611 ; PHS41720J, Lot Number 21IMB848 ; PHS41720J, Lot Number 21LMA466 ; PHS41720J, Lot Number 22AMC812 ; PHS41720J, Lot Number 22CMH766 ; PHS41720J, Lot Number 22EMD326 ; PHS41720J, Lot Number 22HMB227 ; PHS41720J, Lot Number 22JMF258 ; PHS41720J, Lot Number 23AMB719 ; PHS41720J, Lot Number 23CMI096 ; PHS41720J, Lot Number 23EMD393 ; PHS41720J, Lot Number 23FMD285 ; PHS41720J, Lot Number 23GMI391 ; PHS41720J, Lot Number 23HMF798 ; PHS41720J, Lot Number 23JMH827 ; PHS41720J, Lot Number 24AMB622 ; VM96DLEX2L , Lot Number 22DBC917

Distribution

Worldwide distribution.