Aptio Automation systems with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Clinical Chemistry Test Systems - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system consolidates analytical instruments, such as the ARCHITECT c8000 System into a unified workstation that perform a variety of instrument specific assays such as Sodium, Potassium and Chloride. Sodium, Potassium and Chloride measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance.

Recall: Z-2975-2020 · Reported September 23, 2020

Enforcement

Recall Number
Z-2975-2020
Event ID
86263
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Inpeco S.A.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 23, 2020
Initiation Date
August 7, 2020
Classification Date
September 16, 2020
Termination Date
December 2, 2021
Address
Via San Gottardo 10, N/A, Lugano, N/A, N/A, Switzerland

Description

Aptio Automation systems with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Clinical Chemistry Test Systems - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system consolidates analytical instruments, such as the ARCHITECT c8000 System into a unified workstation that perform a variety of instrument specific assays such as Sodium, Potassium and Chloride. Sodium, Potassium and Chloride measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance.

Reason

Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results

Code Info

Serial Numbers: US: AP2.0111 AP2.0137 AP2.0190 AP2.0191 AP2.0290 AP2.0296 AP2.0321 AP2.0322 AP2.0349 AP2.0385 AP2.0398 OUS: AP2.0111 AP2.0113 AP2.0143 AP2.0170 AP2.0225 AP2.0242 AP2.0250 AP2.0259 AP2.0265 AP2.0270 AP2.0285 AP2.0295 AP2.0316 AP2.0350 AP2.0382 AP2.0388 AP2.0389 AP2.0390 AP2.0392 AP2.0448 APT.0058 APT.0065

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AK, FL, IA, LA, MA, MS, MT, NC, OH, TX, VA and the countries of AUSTRALIA, AUSTRIA, BELGIUM, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, NETHERLANDS. NORWAY SAUDI ARABIA SOUTH KOREA SWEDEN UNITED ARAB EMIRATES

Quantity

10 US and 22 OUS