FDA Enforcement
Class II
Terminated
Overhead Counterpoise Systems (OCS), OCS, 580mm CEILING MOUNT, PORTEGRA2, Catalog # OCS CEIL 58, Material Number 84433765
Recall: Z-2962-2018
·
Reported September 12, 2018
Enforcement
- Recall Number
- Z-2962-2018
- Event ID
- 80727
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Bayer Medical Care, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 12, 2018
- Initiation Date
- June 28, 2018
- Classification Date
- August 31, 2018
- Termination Date
- November 26, 2018
- Address
- 1 Bayer Dr, N/A, Indianola, PA, 15051-9702, United States
Description
Overhead Counterpoise Systems (OCS), OCS, 580mm CEILING MOUNT, PORTEGRA2, Catalog # OCS CEIL 58, Material Number 84433765
Reason
The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
Code Info
Serial Numbers: 917/21956, 516/21035, 817/21916, 817/21910
Distribution
The devices were distributed to the following US states: OH, TX, and WV.
Quantity
4