FDA Enforcement Class II Terminated

Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a support for an injector head. , SHORT CEIL MNT, STD HORZ ARM, Catalog # OCS 115AS, Material Number 59880288

Recall: Z-2961-2018 · Reported September 12, 2018

Enforcement

Recall Number
Z-2961-2018
Event ID
80727
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Bayer Medical Care, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 12, 2018
Initiation Date
June 28, 2018
Classification Date
August 31, 2018
Termination Date
November 26, 2018
Address
1 Bayer Dr, N/A, Indianola, PA, 15051-9702, United States

Description

Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a support for an injector head. , SHORT CEIL MNT, STD HORZ ARM, Catalog # OCS 115AS, Material Number 59880288

Reason

The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.

Code Info

Serial Number: 4061630901

Distribution

The devices were distributed to the following US states: OH, TX, and WV.

Quantity

1