FDA Enforcement Class II Ongoing

PrimeStore Molecular Transport Medium (PS-MTM), Model Numbers LH-1-2, LH-1-3, LH-1-4, LH-1-5, LH-1-5CC, LH-1-15CC - Product Usage: intended for stabilization, transportation, and inactivation of infectious unprocessed nasal washes, nasopharyngeal, oral/throat swabs* suspected of containing Influenza A virus RNA. PS-MTM is also intended for the stabilization, transportation, and inactivation of infectious unprocessed sputum samples suspected of containing Mycobacterium tuberculosis (MTB) DNA from human samples.

Recall: Z-2952-2020 · Reported September 23, 2020

Enforcement

Recall Number
Z-2952-2020
Event ID
86189
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
LONGHORN VACCINES AND DIAGNOSTIC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 23, 2020
Initiation Date
June 3, 2020
Classification Date
September 11, 2020
Address
2 Bethesda Metro Cente, N/A, Bethesda, MD, 20814, United States

Description

PrimeStore Molecular Transport Medium (PS-MTM), Model Numbers LH-1-2, LH-1-3, LH-1-4, LH-1-5, LH-1-5CC, LH-1-15CC - Product Usage: intended for stabilization, transportation, and inactivation of infectious unprocessed nasal washes, nasopharyngeal, oral/throat swabs* suspected of containing Influenza A virus RNA. PS-MTM is also intended for the stabilization, transportation, and inactivation of infectious unprocessed sputum samples suspected of containing Mycobacterium tuberculosis (MTB) DNA from human samples.

Reason

The tubes may have been delivered to some users without labels. Users may not be able to identify the tubes without these labels. If samples stored or transported in tubes are tested on certain systems or come into contact with bleach they can produce cyanide gas. The solution in these tubes contains guanidine thiocynate and may put laboratory personnel at risk of exposure to cyanide gas when it comes into contact with bleach or reagents containing bleach.

Code Info

Lot Numbers: 200421A 200421B E200427A E200427B E2000505B CN200423 CN200427

Distribution

Nationwide distribution including in the state of MD.

Quantity

1,936,050