Verigene Enteric Pathogens Nucleic Acid Test (EP) for detecting and identification of pathogenic enteric bacteria, viruses, and genetic virulence markers from liquid or soft stool preserved in Cary-Blair medium, collected from individuals with signs and symptoms of gastrointestinal infection.
Enforcement
- Recall Number
- Z-2952-2018
- Event ID
- 80517
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Luminex Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 5, 2018
- Initiation Date
- May 8, 2018
- Classification Date
- August 30, 2018
- Termination Date
- June 29, 2020
- Address
- 12212 Technology Blvd, N/A, Austin, TX, 78727-6101, United States
Description
Verigene Enteric Pathogens Nucleic Acid Test (EP) for detecting and identification of pathogenic enteric bacteria, viruses, and genetic virulence markers from liquid or soft stool preserved in Cary-Blair medium, collected from individuals with signs and symptoms of gastrointestinal infection.
Elevated false positive results Yersinia enterocolitica (Yersinia) from customers using the EP Nucleic Acid Test Cartridges component (20-006-023) from the Verigene EP Nucleic Acid Test Kit (20-005-023).
Catalog number 20-005-023; Cartridge Catalog number 20-006-023; Lot 032718023A; UDI 00857573006041
Domestic Distribution: Arizona, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Kansas, Michigan, Minnesota, Missouri, New Hampshire, Tennessee, Texas, Utah, West Virginia, Wisconsin.
790 Cartridge Carriers containing 3160 Cartridges