FDA Enforcement Class II Ongoing

Isoflurane E_Vap.(Safety Filling adapter): used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: Part Number 115-066758-00

Recall: Z-2951-2024 · Reported September 11, 2024

Enforcement

Recall Number
Z-2951-2024
Event ID
95099
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Mindray DS USA, Inc. dba Mindray North America
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 11, 2024
Initiation Date
July 22, 2024
Classification Date
September 4, 2024
Address
800 Macarthur Blvd, Mahwah, NJ, 07430-2001, United States

Description

Isoflurane E_Vap.(Safety Filling adapter): used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: Part Number 115-066758-00

Reason

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

Code Info

UDI-DI: 06936415975753 Serial Number: AH6B24000165 AH6B24000170 AH6B24000167 AH6B24000163 AH6B24000166 AH6B24000173 AH6B24000174 AH6B24000183

Distribution

US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada

Quantity

24 units